THORATEC® HEARTMATE II®, LVAS IMPLANT KIT (WITH SEALED GRAFTS) 104911Recalls & reported problems
Thoratec Corporation · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Thoratec HeartMate II is a heart pump implanted to assist the heart's pumping. Many of its 1,090 reports describe the pump stopping or slowing, mechanical problems and breaks, and the reports include 449 deaths and 462 injuries. Reports fell to 6 in the last 12 months from 22 in the previous 12 months. One open recall, from February 2024, concerns outflow graft deformation that can block blood flow.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,092
- Deaths reported
- 451
- Injuries reported
- 463
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1,092 adverse event reports that list this exact device (DI 00813024011170), received between Jun 11, 2018 and Sep 25, 2026 — classified as 451 deaths, 463 injuries, 178 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (188 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 19, 2024Class IOpen, Classified
Observed outflow graft deformation known as Extrinsic Outflow Graft Obstruction (EOGO) associated with Left Ventricular Assist System.
Z-1300-2024 · Root cause: Material/Component Contamination · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem693
- Mechanical Problem188
- Pumping Stopped157
- Break108
- Infusion or Flow Problem90
- Decreased Pump Speed62
- Material Split, Cut or Torn30
- Naturally Worn22
Patient problems
- Insufficient Information312
- No Known Impact Or Consequence To Patient140
- No Clinical Signs, Symptoms or Conditions117
- Death109
- Unspecified Infection66
- No Consequences Or Impact To Patient64
- Hemolysis51
- Thrombus51
Recent adverse event reports
The 25 most recent of 1,092 reports, in the reporter's own words as published by the FDA.
InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Hemorrhage/Blood Loss/Bleeding; Thrombosis/Thrombus; Unspecified Infection
MAUDE report 50901225 ↗DeathSep 25, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Insufficient Information
MAUDE report 50901141 ↗DeathSep 21, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient passed away due to heart failure. Whether the outcome was device or therapy related was not provided by the customer.
Patient problems: Heart Failure/Congestive Heart Failure
MAUDE report 50841723 ↗DeathAug 27, 2026Disconnection; Adverse Event Without Identified Device or Use Problem
It was reported that the patient's wife found the patient down and disconnected form their power source with no alarms active at the time. The patient's wife stated that she was sleeping in a different room and never heard any alarms. She believed the patient was possibly getting up for the restroom, got confused and could not get themselves on Batteries properly. The patient passed away. Log Files were sent for review and captured the Driveline disconnect and the Pump not operating. The Driveline was disconnected on (b)(6) 2026 at 10:30. The Pump operated at he set speed for the remainder of the Log File which ended on (b)(6) 2026 at 4:18 with Low Battery Advisory and Hazard events. It was unknown if the device would be returned for evaluation.
Patient problems: Insufficient Information
MAUDE report 50484244 ↗DeathAug 17, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient passed away due to failure to thrive, progressive renal failure, and systolic and diastolic heart failure. The death was considered to be device/therapy related. The Pump was not explanted, and the device parameters were stated to be within normal limits for the patient.
Patient problems: Heart Failure/Congestive Heart Failure; Renal Failure
MAUDE report 50277091 ↗MalfunctionJan 2, 2026Mechanical Problem
IT WAS REPORTED THAT A CLAMSHELL REPAIR WAS PERFORMED WITH NO ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23950054 ↗DeathDec 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS ADMITTED ON (B)(6) 2025 TO (B)(6) 2025 WITH GASTROINTESTINAL BLEEDING (GIB). ESOPHAGOGASTRODUODENOSCOPY (EGD)/COLONOSCOPY WAS UNABLE TO FIND THE SOURCE, BUT HEMOGLOBIN CONTINUED TO TREND DOWNWARD EVEN WITH BLOOD TRANSFUSIONS. THE PATIENT PASSED AWAY AT HOME WHILE ON HOSPICE ON (B)(6) 2025 AT 14:05. THE OUTCOME WAS NOT CONSIDERED TO BE DEVICE RELATED. THE OUTCOME WAS RELATED TO GIB SECONDARY TO ANTICOAGULATION AND VENTRICULAR ASSIST DEVICE (VAD) SUPPORT. THE DEVICE PARAMETERS WERE WITHIN NORMAL LIMITS FOR THIS PATIENT.
Patient problems: Gastrointestinal Hemorrhage
MAUDE report 23759662 ↗InjuryNov 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD LOW HEMOGLOBIN AND HEMATOCRIT. IT WAS SAID THE PATIENT HAD BLOOD TRANSFUSIONS. THEY ALSO HAD AN INCISION AND DRAINAGE AS WELL. THE BLOOD TRANSFUSION IMPROVED THE LOW HEMOGLOBIN AND HEMATOCRIT. THE ONSET OF THE INFECTION WAS (B)(6) 2022 (THE ONSET OF THE INFECTION IS COVERED UNDER MANUFACTURER REPORT NUMBER: 2916596-2023-05254). THE PATIENT HAD FATIGUE, AND THE SOURCE OF THE INFECTION WAS PUMP POCKET. DRAINAGE AND METHICILLIN-RESISTANT STAPHYLOCOCCUS AUREUS (MRSA) WAS IDENTIFIED. THE INFECTION WAS BEING TREATED WITH ERTAPENEM INTRAVENOUS AND WAS NOT RESOLVED. IT WAS PLANNED TO CONTINUE TREATING THE PATIENT WITH ERTAPENEM INTRAVENOUS.
Patient problems: Drug Resistant Bacterial Infection; Fatigue; Hemorrhage/Blood Loss/Bleeding; Unspecified Infection
MAUDE report 23451212 ↗MalfunctionOct 15, 2025Disconnection; Electrical /Electronic Property Problem; Disconnection; Electrical /Electronic Property Problem
IT WAS REPORTED THAT LOG FILES WERE SUBMITTED FOR REVIEW WHICH CAPTURED DRIVELINE DISCONNECT ALARMS ON (B)(6) 2025.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23294429 ↗InjurySep 30, 2025Mechanical Problem; Infusion or Flow Problem; Mechanical Problem; Infusion or Flow Problem
IT WAS REPORTED THAT THE PATIENT HAD KNOWN OUTFLOW THROMBUS SINCE (B)(6) 2023. THE OUTFLOW THROMBUS WAS INITIALLY IDENTIFIED THROUGH A ROUTINE ECHOCARDIOGRAM (ECHO). AN ECHO AND A GATED COMPUTED TOMOGRAPHY (CT) WERE USED AT THE TIME FOR DIAGNOSTIC TESTING. THE PATIENT WAS BRIEFLY PLACED ON HEPARIN. LOG FILES WERE NOT AVAILABLE FROM THE ORIGINAL EVENT IN 2023.
Patient problems: Thrombosis/Thrombus
MAUDE report 23183783 ↗MalfunctionSep 28, 2025Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE CUSTOMER SUBMITTED IMAGES OF THE DRIVELINE OUTER LAYER DAMAGE COVERED BY RESCUE TAPE. THE DRIVELINE SEEMED EXTREMELY FRAGILE, PAPER THIN LIKE, AND WITH MULTIPLE AREAS OF SMALL TEARS THAT REQUIRED RESCUE TAPE. THE IMAGES SUBMITTED SHOWED AREAS THAT APPEARED TO HAVE EXISTING AREAS COVERED BY RESCUE TAPE. LOG FILES WERE SENT FOR REVIEW. THE LOG FILE DATA SHOWED NO INDICATIONS OF CONDUCTOR WEAR OR DAMAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23163976 ↗InjurySep 23, 2025Mechanical Problem; Infusion or Flow Problem; Adverse Event Without Identified Device or Use Problem; Mechanical Problem; Infusion or Flow Problem; Adverse Event Without Identified Device or Use Problem
FROM ECHOCARDIOGRAM (ECHO) DONE ON (B)(6) 2025, IT WAS SAID AN ULTRASOUND ENHANCING AGENT WAS USED. THE HEARTMATE II LVAD SETTINGS AT THE TIME OF THIS EXAM WERE THAT THE PUMP SPEED WAS AT 10,000 RPM, FLOW RATE WAS AT 5.5 L/MIN, PULSE INDEX WAS AT 5.3, POWER WAS AT 6.6 W. THE LEFT VENTRICULAR CHAMBER SIZE AND WALL THICKNESS WERE NORMAL. LEFT VENTRICULAR CHAMBER SIZE AND WALL THICKNESS ARE NORMAL. THE LEFT VENTRICULAR INTERNAL DIAMETER (LVID) WAS 5.6CM. THERE WAS NO LEFT VENTRICULAR THROMBUS SEEN ON THE CONTRAST ENHANCED IMAGES. THE LEFT VENTRICULAR SYSTOLIC FUNCTION WAS SEVERELY REDUCED WITH A VISUALLY ESTIMATED EJECTION FRACTION OF 10%. THERE WAS NO OBVIOUS OBSTRUCTION, SEPTAL SUCTIONING, OR THROMBUS AT THE VENTRICULAR ASSIST DEVICE (VAD) INFLOW CANNULA. THE RIGHT VENTRICLE WAS DILATED WITH GROSSLY NORMAL SYSTOLIC FUNCTION BY VISUAL ASSESSMENT IN LIMITED VIEWS. WITH THE LVAD IN PLACE, THE AORTIC VALVE OPENED MINIMALLY (ONLY SEEN IN SHORT AXIS) WITH EVERY CARDIAC CYCLE WITH TRACE REGURGITATION. THERE WAS MILD TRICUSPID VALVE REGURGITATION. THERE WAS NO PULMONARY HYPERTENSION AND THE ESTIMATED PULMONARY ARTERIAL SYSTOLIC PRESSURE WAS 23 MMHG. PERICARDIUM WAS NORMAL IN APPEARANCE WIT
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Pulmonary Valve Insufficiency/ Regurgitation; Thrombosis/Thrombus; Tricuspid Valve Insufficiency/ Regurgitation
MAUDE report 23126681 ↗DeathJul 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PASSED AWAY FROM A COMBINATION OF PSEUDOMONAS INFECTION, ACUTE KIDNEY INJURY, ENCEPHALOPATHY, AND FAILURE TO THRIVE.
Patient problems: Bacterial Infection; Encephalopathy; Kidney injury
MAUDE report 22645675 ↗MalfunctionJul 17, 2025Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT WAS IN NEED OF A NON-URGENT ¿WIRE SURGERY.¿.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22534722 ↗DeathJul 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THROUGH PATIENT OUTCOME THAT THE PATIENT PASSED AWAY.
Patient problems: Insufficient Information
MAUDE report 22522024 ↗DeathJun 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PASSED AWAY DUE TO NATURAL CAUSES. THE OUTCOME WAS NOT CONSIDERED TO BE DEVICE OR THERAPY RELATED. THE DEVICE WOULD NOT BE EXPLANTED.
Patient problems: Insufficient Information
MAUDE report 22289934 ↗DeathJun 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PASSED AWAY DUE TO END STAGE HEART FAILURE. THE PATIENT WAS ON HOSPICE. ADDITIONAL DETAILS LEADING UP TO DEATH WERE UNKNOWN. THE DEATH WAS NOT CONSIDERED TO BE DEVICE RELATED. THE DEVICE OPERATED AS EXPECTED. NO DEVICE WOULD BE RETURNED.
Patient problems: Heart Failure/Congestive Heart Failure
MAUDE report 22177072 ↗DeathJun 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PASSED AWAY. THE CAUSE OF DEATH WAS UNKNOWN. WHETHER THE OUTCOME WAS DEVICE OR THERAPY RELATED WAS NOT PROVIDED BY THE CUSTOMER. THE PUMP WOULD NOT BE EXPLANTED.
Patient problems: Insufficient Information
MAUDE report 22160201 ↗InjuryJun 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT HAD A STERNAL WOUND WITH ABSCESS. THE CULTURE WAS ESCHERICHIA COLI. THEY RETURNED TO OPERATING ROOM FOR DEBRIDEMENT. AT THE TIME, THEY HAD A WOUND VAC IN PLACE. THE PATIENT REMAINED CHRONIC ANTIBIOTICS. HEALTHCARE PLANNED TO CONTINUE TO MONITOR. THE PATIENT WAS ADMITTED FROM (B)(6) 2025.
Patient problems: Abscess; Bacterial Infection
MAUDE report 22136258 ↗InjuryMay 23, 2025Infusion or Flow Problem; Adverse Event Without Identified Device or Use Problem; Infusion or Flow Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS ADMITTED WITH REPORTS OF PROFOUND SHORTNESS OF BREATH AND EXERTIONAL DYSPNEA STATING THEY FELT LIKE HOW THEY DID BEFORE THEIR IMPLANT. ECHOCARDIOGRAM IMAGES AND COMPUTED TOMOGRAPHY ANGIOGRAPHY WERE TO FOLLOW, BUT THE PATIENT HAD A KNOWN OUTFLOW GRAFT OBSTRUCTION THAT WAS SUSPECTED TO BE WORSENING. LOG FILES WERE SENT FOR REVIEW TO DIVULGE ANY APPRECIABLE TRENDS IN POWER OR FLOW OVER TIME. THE LOG FILES COULD CAPTURE A SPEED INCREASE TO OVERCOME THE OBSTRUCTION. THE BASELINE SPEED WAS NOTED TO BE 10000. NO ALARMS WERE FOUND ON INTERROGATION. THE PATIENT HAD SUSPICIOUS PULSATILITY INDEX (PI) EVENTS IN THE HISTORY. THE LOG FILE CAPTURED ROUTINE EVENTS THROUGHOUT THE LOG FILE. THE VENTRICULAR ASSIST DEVICE (VAD) PARAMETERS LOOKED STABLE OVERALL, THERE WERE A FEW BLIPS OF HIGH POWER IN THE 9W RANGE ON (B)(6) 2025, WHICH WERE NOT SUSTAINED AND ALSO ASSOCIATED WITH PI EVENTS. THE PATIENT UNDERWENT A SURGICAL REPAIR FOR THE OUTFLOW GRAFT OBSTRUCTION ON (B)(6) 2025.
Patient problems: Dyspnea; Hemolysis; Lactate Dehydrogenase Increased; Thrombosis/Thrombus
MAUDE report 22079715 ↗DeathApr 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION WAS RECEIVED THAT THE AORTIC ANEURYSM WAS GENETIC IN ORIGIN. THERE WAS NO DEVICE INVOLVEMENT IN THE PATIENT'S DEATH AND THE DEVICE OPERATED AS EXPECTED.
Patient problems: Ruptured Aneurysm
MAUDE report 21901847 ↗InjuryApr 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD A LEFT VENTRICULAR ASSIST DEVICE (LVAD) RELATED PARA-DEVICE SOFT TISSUE INFECTION SINCE (B)(6) 2024. THE PATIENT HAD AN LVAD DRIVELINE WOUND PUS DISCHARGE LEADING TO RE-HOSPITALIZATION FOR INFECTION SURVEY AND THE PATIENT WAS STABLE. THE PATIENT HAD SYMPTOMS OF PUS DISCHARGE AT THE DRIVELINE WOUND AND THE PATIENT FELT PAIN AT THE WOUND AREA. A COMPUTED TOMOGRAPHY (CT) SCAN CONFIRMED TISSUE INFECTION AROUND THE LVAD DEVICE POCKET. THE PUS CULTURE FOUND PROTEUS MIRABILIS AND ACHROMOBACTER SPP. THE PATIENT WAS TREATED WITH ORAL MOXIFLOXACIN AND THEN DISCHARGED. ON (B)(6) 2025 THE PATIENT UNDERWENT A HEART TRANSPLANT. THIS WAS DEVICE OR THERAPY RELATED. THE HEART TRANSPLANT RESOLVED THE INFECTION AND THE DRIVELINE WOUND HEALED. THE DEVICE OPERATED AS EXPECTED. THE EVENT DATE WAS ESTIMATED AND THE EXACT DATE WAS NOT PROVIDED.
Patient problems: Bacterial Infection; Pain; Purulent Discharge; Skin Infection
MAUDE report 21755906 ↗DeathMar 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PASSED AWAY DUE TO SEPSIS AND VENTRICULAR TACHYCARDIA.
Patient problems: Sepsis; Tachycardia
MAUDE report 21518017 ↗DeathJan 23, 2025Mechanical Problem; Mechanical Problem
THE PATIENT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE (LVAD). IT WAS REPORTED THAT THE PATIENT PRESENTED WITH ELEVATED LACTATE DEHYDROGENASE (LDH) LEVELS, HEMATURIA, AND POWER ELEVATIONS. BIVALIRUDIN WAS INITIATED AND THE PATIENT'S NUMBERS NORMALIZED; HOWEVER, WHEN BIVALIRUDIN WAS DISCONTINUED, LDH BEGAN TO RISE AND HEMATURIA OCCURRED. IT WAS REPORTED THAT THE SURGICAL PLACEMENT OF THE ORIGINAL PUMP WAS DIFFICULT. THE PUMP WAS SUBSEQUENTLY EXCHANGED AND THE HOSPITAL REPORTED THAT THEY PLUGGED THE OLD SITE USING A FELT PLUG AND RE-CORED THE HEART. A THROMBUS WAS SEEN ON THE OUTFLOW STATOR BLADES. ADDITIONAL INFORMATION PROVIDED TO THE MANUFACTURER INDICATED THAT THE HEART WAS RE-CORED BECAUSE THE HOSPITAL FELT THAT THE INFLOW WAS MALPOSITIONED AND WAS POTENTIALLY COMPROMISING FLOW. A DEVICE TRACKING FORM WAS RECEIVED INDICATING THAT THE PATIENT EXPIRED ON (B)(6) 2014 APPROXIMATELY 1 MONTH AFTER THE PUMP EXCHANGE. THE CAUSE OF DEATH WAS NOTED TO BE A MIDDLE CEREBRAL ARTERY (MCA) CEREBROVASCULAR ACCIDENT (CVA) DUE TO VAD THROMBUS. IT WAS REPORTED THAT ANTICOAGULATION LIKELY CONTRIBUTED TO THE STROKE. THE SECOND PUMP WAS REPORTED TO BE FUNCTIONING AS INTENDED AND NO DEVICE IS
Patient problems: Hematuria; Hemolysis; Stroke/CVA; Thrombosis/Thrombus
MAUDE report 21217204 ↗DeathJan 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PASSED AWAY AND THE CAUSE WAS UNKNOWN. THE OUTCOME WAS NOT DEVICE OR THERAPY RELATED. AN AUTOPSY WAS NOT PERFORMED AND WOULD NOT BE PROVIDED. THE PUMP WAS NOT EXPLANTED AND WOULD NOT BE RETURNED.
Patient problems: Insufficient Information
MAUDE report 21067292 ↗
Other Ventricular (Assisst) Bypass devices
- Thoratec® HeartMate 3™ 106524USThoratec Corporation · Ventricular (Assisst) Bypass7 recalls5,096 deaths16,478 reports
- THORATEC® HEARTMATE®, LVAS IMPLANT KIT (WITH SEALED GRAFTS) 106015Thoratec Corporation · Ventricular (Assisst) Bypass2 recalls2,141 deaths8,000 reports
- Thoratec® HeartMate 3™ 106531USThoratec Corporation · Ventricular (Assisst) Bypass1 recall11 deaths4,646 reports
- HVAD Pump Implant Kit 1103Heartware Inc. · Ventricular (Assist) Bypass6 recalls424 deaths2,952 reports
- HVAD® Pump Implant Kit (Enhanced) 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls573 deaths2,807 reports
- HVAD® Pump Implant Kit 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls666 deaths2,737 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assisst) Bypass5 recalls14 deaths2,161 reports
- HEARTMATE®, MOBILE POWER UNIT, NA 107754Thoratec Corporation · Ventricular (Assisst) Bypass1 death2,098 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page