THORATEC® HEARTMATE II®, PUMP AND CONTROLLER ONLY KIT 105903Recalls & reported problems
Thoratec Corporation · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 2
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00813024011217), received between Jun 26, 2019 and Jan 25, 2023 — classified as 2 deaths, 1 injury.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Material Split, Cut or Torn1
- Naturally Worn1
Patient problems
- Insufficient Information2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
DeathJan 25, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE ACCOUNT REPORTED THAT THE PATIENT EXPIRED ON (B)(6) 2018. THE CAUSE OF DEATH WAS UNKNOWN AS THE PATIENT WAS FOUND DECEASED IN BED, AND AN AUTOPSY WAS NOT PERFORMED. THE PUMP REPORTEDLY FUNCTIONED AS INTENDED AND THE PATIENT'S OUTCOME WAS NOT SUSPECTED TO BE DEVICE RELATED.
Patient problems: Insufficient Information
MAUDE report 16243833 ↗DeathApr 1, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PASSED AWAY. THE CAUSE OF DEATH WAS UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 13976179 ↗InjuryJun 26, 2019Naturally Worn; Material Split, Cut or Torn; Naturally Worn; Material Split, Cut or Torn
THE PATIENT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE (LVAD) ON (B)(6) 2013. IT WAS REPORTED THE PATIENT WAS SEEN FOR A FOLLOW VISIT AND NOTED TO HAVE TWISTING OF DRIVELINE (DL) WITH MULTIPLE PORTIONS COVERED WITH RESCUE TAPE. INTERROGATED DUE TO CONCERN OF INTEGRITY OF WIRES INSIDE DRIVELINE. ON INTERROGATION OF DEVICE, PATIENT HAD NO EXTERNAL POWER ALARM ON (B)(6) THAT OCCURRED WHEN SWITCHING WHITE LEAD ON PATIENT CABLE. THE LOG FILE WAS REVIEWED. THE LOG FILE ANALYSIS SHOWED THE NO EXTERNAL POWER ON (B)(6) 2019 WHILE THE PATIENT WAS SWITCHING POWER SOURCES. THE DL DATA APPEARED NORMAL SO THE TWISTING HAD NOT DEGRADED THE CONDUCTORS AT THIS TIME. THERE WERE NO OTHER UNUSUAL EVENTS RECORDED IN THE LOG FILE. THE MCS EQUIPMENT WAS OPERATING AS INTENDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8735884 ↗
Other Ventricular (Assisst) Bypass devices
- Thoratec® HeartMate 3™ 106524USThoratec Corporation · Ventricular (Assisst) Bypass7 recalls5,096 deaths16,478 reports
- THORATEC® HEARTMATE®, LVAS IMPLANT KIT (WITH SEALED GRAFTS) 106015Thoratec Corporation · Ventricular (Assisst) Bypass2 recalls2,141 deaths8,000 reports
- Thoratec® HeartMate 3™ 106531USThoratec Corporation · Ventricular (Assisst) Bypass1 recall11 deaths4,646 reports
- HVAD Pump Implant Kit 1103Heartware Inc. · Ventricular (Assist) Bypass6 recalls424 deaths2,952 reports
- HVAD® Pump Implant Kit (Enhanced) 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls573 deaths2,807 reports
- HVAD® Pump Implant Kit 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls666 deaths2,737 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assisst) Bypass5 recalls14 deaths2,161 reports
- HEARTMATE®, MOBILE POWER UNIT, NA 107754Thoratec Corporation · Ventricular (Assisst) Bypass1 death2,098 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page