Skip to content

HeartMate®, Mobile Power Unit 108285Recalls & reported problems

Thoratec Corporation · Class 3

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00813024012696), received between Aug 9, 2021 and Jul 14, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Electrical Power Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Electrical Power Problem1
  • Failure to Power Up1
  • Material Twisted/Bent1
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 14, 2025Failure to Power Up; Failure to Power Up

    IT WAS REPORTED THAT MOBILE POWER UNIT (MPU) WAS NOT WORKING, TURNED OFF. THE MPU WAS EXCHANGED. THE LOG FILES DID NOT CAPTURE ANY TYPE OF VOLTAGE EVENTS AND ALSO NOTED THAT THE PATIENT WAS ONLY USING BATTERY POWER IN THIS DATA CAPTURE AND NOT THE MPU.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22500721 ↗
  • MalfunctionSep 26, 2023Electrical Power Problem; Electrical Power Problem

    IT WAS REPORTED THAT LOG FILES WERE SUBMITTED FOR REVIEW. THE LOG FILES DISPLAYED ONE TRANSIENT LOW VOLTAGE ALARM ON (B)(6) 2023 AT 3:31 PM WHILE TETHERED TO THE MOBILE POWER UNIT (MPU). THIS ALARM SEEMED TO BE CAUSED BY THE POWER TO THE MPU BEING BRIEFLY INTERRUPTED. THIS MIGHT HAVE BEEN DUE TO THE POWER CORD COMING LOOSE FROM THE WALL OUTLET, THE HOUSE LOSING POWER, OR POSSIBLE USER ERROR. THERE WERE NO OTHER UNUSUAL EVENTS SEEN WITHIN THE LOG FILE. THE MECHANICAL CIRCULATORY SYSTEM (MCS) EQUIPMENT OPERATED AS EXPECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17818831 ↗
  • MalfunctionAug 9, 2021Material Twisted/Bent; No Apparent Adverse Event; Material Twisted/Bent; No Apparent Adverse Event

    IT WAS REPORTED THAT THE CONNECTOR PINS OF THE PATIENT CABLE WERE BENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12293364 ↗

Other Ventricular (Assisst) Bypass devices

All Ventricular (Assisst) Bypass devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page