HeartMate, 14v Li-Ion Battery (SING) 102515-SINGRecalls & reported problems
Thoratec Corporation · Class 3
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00813024012733), received between Dec 31, 2018 and Jul 20, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Improper or Incorrect Procedure or Method1
- Mechanical Problem1
- Power Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Shock1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 20, 2021Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT SPILLED LIQUID ON THE BATTERY CHARGER, WHICH MADE AN ELECTRIC ZAPPING SOUND AND STARTED TO SMOKE. THE PATIENT THEN USED A SET OF BATTERIES FROM THE CHARGER AND STARTED HAVING FREQUENT LOW VOLTAGE AND POWER CABLE DISCONNECT ALARMS. THE BATTERIES AND CHARGER WERE EXCHANGED AND THE PATIENT HAD NO FURTHER ALARMS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12197355 ↗MalfunctionJan 21, 2021Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
THE PATIENT'S DEVICE HAD LOW POWER HAZARD ALARMS ON (B)(6) 2021 WHICH CONTRIBUTED TO ADDITIONAL PUMP STOPPAGES. THESE WERE DUE TOTAL DEPLETION OF BATTERY POWER. RELATED MANUFACTURER REPORT NUMBER # 2916596-2021-00046. RELATED MANUFACTURER REPORT NUMBER # 2916596-2021-00054.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11208575 ↗MalfunctionDec 31, 2018Power Problem; Power Problem
THE PATIENT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE (LVAD) ON (B)(6) 2017. IT WAS REPORTED THAT THE PATIENT¿S PRIMARY SYSTEM CONTROLLER WAS PRODUCING ALARMS OF RANDOM BEEP WHEN CONNECTED TO ONE SET OF BATTERIES. IT WAS ALSO REPORTED THAT ON AN UNSPECIFIED DATE WHEN DISCONNECTING FROM THE BATTERIES, THE PATIENT FELT ONE OF THE BATTERIES VIBRATE AND THEN RECEIVED AN EXTERNAL ELECTRICAL SHOCK TWICE. THE BATTERIES WERE EXCHANGED. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT DUE TO THE EVENT.
Patient problems: Shock
MAUDE report 8205661 ↗
Other Ventricular (Assisst) Bypass devices
- Thoratec® HeartMate 3™ 106524USThoratec Corporation · Ventricular (Assisst) Bypass7 recalls5,096 deaths16,478 reports
- THORATEC® HEARTMATE®, LVAS IMPLANT KIT (WITH SEALED GRAFTS) 106015Thoratec Corporation · Ventricular (Assisst) Bypass2 recalls2,141 deaths8,000 reports
- Thoratec® HeartMate 3™ 106531USThoratec Corporation · Ventricular (Assisst) Bypass1 recall11 deaths4,646 reports
- HVAD Pump Implant Kit 1103Heartware Inc. · Ventricular (Assist) Bypass6 recalls424 deaths2,952 reports
- HVAD® Pump Implant Kit (Enhanced) 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls573 deaths2,807 reports
- HVAD® Pump Implant Kit 1103Heartware Inc. · Ventricular (Assisst) Bypass4 recalls666 deaths2,737 reports
- HVAD® Controller 1420Heartware Inc. · Ventricular (Assisst) Bypass5 recalls14 deaths2,161 reports
- HEARTMATE®, MOBILE POWER UNIT, NA 107754Thoratec Corporation · Ventricular (Assisst) Bypass1 death2,098 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗