Skip to content

Cyclo G6 Laser Console 66245Recalls & reported problems

Iridex Corporation · Class 2

All models of Cyclo G6 Laser Console (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00813125015374), received between Feb 12, 2020 and Jul 28, 2026 — classified as 6 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information3
  • Adverse Event Without Identified Device or Use Problem2
  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • Eye Burn2
  • Visual Impairment2
  • Eye Injury1
  • Visual Disturbances1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 28, 2026Adverse Event Without Identified Device or Use Problem

    It was reported to Iridex that a patient with bilateral secondary glaucoma & bilateral nanophthalmos, underwent MicroPulse transscleral laser treatment using an IRIDEX Cyclo G6 laser console and a MicroPulse P3 (MP3) delivery device. Six months following treatment, the patient was unable to open the treated left eye. Clinical evaluation revealed phthisis bulbi, characterized by significant globe atrophy. Phthisis bulbi is a shrunken, non-functional, and disorganized end-stage ocular condition that may result from severe trauma, chronic inflammation, or infection. It is typically associated with permanent vision loss, hypotony (low intraocular pressure), and structural collapse of the globe.The patient had a pre-existing diagnosis of severe nanophthalmos, a developmental ocular condition characterized by an abnormally small eye due to compromised ocular growth. Nanophthalmos is generally congenital or present from early childhood and is recognized within the spectrum of developmental eye disorders.According to the MP3 Probe Instructions for Use (IFU), the probe tip protrudes 0.4 mm from the handpiece and is specifically designed to facilitate accurate positioning of the fiber-optic

    Patient problems: Visual Disturbances

    MAUDE report 50059856 ↗
  • InjuryOct 14, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IRIDEX BECAME AWARE OF A PATIENT EXPERIENCING CONJUNCTIVAL BLANCHING DURING TREATMENT WITH A CYCLO G6 LASER SYSTEM. THE DELIVERY DEVICE USED FOR THE TREATMENT IS UNKNOWN. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED, AND NO ADDITIONAL INFORMATION HAS BEEN PROVIDED BY THE CLINIC OR TREATING PHYSICIAN. ENGINEERING FAILURE ANALYSIS COULD NOT BE PERFORMED BECAUSE THE DEVICE WAS NOT RETURNED FOR EVALUATION. IRIDEX WILL CONTINUE TO FOLLOW-UP WITH THE DOCTOR FOR INFORMATION REGARDING PATIENT RECOVERY.

    Patient problems: Eye Burn

    MAUDE report 15608036 ↗
  • InjuryJul 15, 2022Insufficient Device Problem Information

    IRIDEX BECAME AWARE OF A PATIENT EXPERIENCING CHARRING ON THE EYE DURING TREATMENT WITH A CYCLO G6 LASER SYSTEM. THE DELIVERY DEVICE THAT WAS USED IN COMBINATION WITH THE CYCLO G6 IS UNKNOWN, AND NO RESPONSE WAS PROVIDED BY THE CUSTOMER TO IRIDEX FOLLOW-UP EMAILS. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED, AND NO ADDITIONAL INFORMATION HAS BEEN PROVIDED BY THE CLINIC OR TREATING PHYSICIAN. ENGINEERING FAILURE ANALYSIS COULD NOT BE PERFORMED BECAUSE THE DEVICE WAS NOT RETURNED FOR EVALUATION. IRIDEX WILL CONTINUE TO FOLLOW-UP WITH THE DOCTOR FOR INFORMATION REGARDING PATIENT RECOVERY.

    Patient problems: Eye Injury

    MAUDE report 15028823 ↗
  • InjuryMay 27, 2022Adverse Event Without Identified Device or Use Problem

    IRIDEX BECAME AWARE OF A COMPLAINT REPORTING A PATIENT EXPERIENCING CONJUNCTIVAL BURN DURING TREATMENT WITH THE MICROPULSE P3 PROBE. THE TREATMENT WAS PERFORMED WITHIN SPECIFICATIONS. A SPECIFIC ROOT CAUSE COULD NOT BE DETERMINED. BASED ON SIMILAR REPORTED EVENTS, THE MOST LIKELY CAUSE OF THE ISSUE IS SUSPECTED TO BE CONTAMINATION (E.G. BLOOD, TISSUE OR BETADINE) ON THE PROBE TIP WHEN THE LASER WAS ACTIVATED OR LASER ACTIVATION WHEN THE PROBE IS OVER PIGMENTED CONJUNCTIVAL TISSUE. ENGINEERING FAILURE ANALYSIS WAS NOT PERFORMED BECAUSE THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE MICROPULSE P3 IFU HAS CLEAR INSTRUCTIONS TO EXAMINE THE PROBE TIP BEFORE TREATMENT AND TO AVOID AREAS OF PERILIMBAL PIGMENTATION. PER THE IFU 'EXCESSIVE TREATMENT POWER MAY RESULT IN OCULAR SURFACE BURNS OR CILIARY BODY HEMORRHAGE. CONTAMINATION OF THE FIBER OPTIC TIP BY BLOOD OR TISSUE CHAR MAY RESULT IN OCULAR SURFACE BURNS. EXCESSIVE ENERGY MAY CAUSE EQUATORIAL BURNS. IF THE DEVICES ARE RETURNED AND THE RESULTS OF FAILURE ANALYSIS INDICATE NEW INFORMATION, THE ADVERSE EVENT REPORT WILL BE AMENDED. THE RESULTS OF POST MARKET DATA REVIEW INDICATE THAT THIS FAILURE MODE IS NOT OCCURRING WITH GREATER SEVE

    Patient problems: Eye Burn

    MAUDE report 14528471 ↗
  • InjuryFeb 12, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    IRIDEX RECEIVED A REPORT THERE WERE TWO PATIENTS THAT HAD SUFFERED FROM MYDRIASIS; ONE SUFFERED FROM CHRONIC MYDRIASIS AND THE OTHER LONG-TERM MYDRIASIS. THE PATIENTS DEVELOPED MYDRIASIS AFTER TREATMENT FROM A MICROPULSE P3 DEVICE WITH A CYCLO G6 LASER CONSOLE.

    Patient problems: Visual Impairment

    MAUDE report 9703498 ↗
  • InjuryFeb 12, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    IRIDEX RECEIVED A REPORT THERE WERE TWO PATIENTS THAT HAD SUFFERED FROM MYDRIASIS; ONE SUFFERED FROM CHRONIC MYDRIASIS AND THE OTHER LONG-TERM MYDRIASIS. THE PATIENTS DEVELOPED MYDRIASIS AFTER TREATMENT FROM A MICROPULSE P3 DEVICE WITH A CYCLO G6 LASER CONSOLE.

    Patient problems: Visual Impairment

    MAUDE report 9703497 ↗

Other Powered Laser Surgical Instrument devices

All Powered Laser Surgical Instrument devices →

Sources and updates