Skip to content

Quick-Cross Support Catheter 518-065 (UDI 00813132020712)Recalls & reported problems

Spectranetics Corporation · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00813132020712), received between Sep 28, 2018 and Sep 28, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Damaged by Another Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Damaged by Another Device1
  • Material Perforation1
  • Material Puncture/Hole1
  • Mechanical Problem1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2018Mechanical Problem; Material Puncture/Hole; Material Perforation; Device Damaged by Another Device; Mechanical Problem; Material Puncture/Hole; Material Perforation; Device Damaged by Another Device

    DURING A PERIPHERAL VASCULAR SURGERY UTILIZING FEMORAL ACCESS, THE SURGEON STATED WHEN USING THE QUICK-CROSS SUPPORT CATHETER 518-065, THEY COULD NOT GO AS LOW IN THE LEG AS WISHED, PAST THE PROXIMAL ANTERIOR TIBIAL ARTERY. ALSO, WHEN TAKING OUT THE QUICK-CROSS THEY NOTICED A HOLE IN THE CATHETER ABOVE THE TIP. NO DEVICE REMNANTS WERE DETECTED TO BE LEFT IN THE PATIENT. THE CASE WAS SUCCESSFULLY COMPLETED WITH A BALLOON AND GUIDE-WIRE. NO IMPACT ON PATIENT OUTCOME DUE TO THIS EVENT. THIS IS BEING REPORTED DUE TO THE POTENTIAL FOR SERIOUS INJURY WITH RECURRENCE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7920057 ↗

Other Catheter, Percutaneous devices

All Catheter, Percutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page