AngioSculpt OTW PTA Scoring Balloon Catheter 2092-5040Recalls & reported problems
The Spectranetics Corporation · Class 2
All models of AngioSculpt OTW PTA Scoring Balloon Catheter (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Spectranetics AngioSculpt scoring balloon catheter is used to open narrowed blood vessels, including dialysis shunts. Most of the 28 reports describe the balloon popping during inflation, often below its rated pressure, with a few mentioning the device becoming stuck or cases where blood clots, an embolus or vessel perforation were noted. There are 3 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 28
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 00813132021412), received between Jul 10, 2020 and Apr 9, 2026 — classified as 3 injuries, 25 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (26 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture26
- Adverse Event Without Identified Device or Use Problem2
- Entrapment of Device1
- Improper or Incorrect Procedure or Method1
Patient problems
- No Clinical Signs, Symptoms or Conditions18
- No Consequences Or Impact To Patient7
- Thrombosis/Thrombus2
- Embolism/Embolus1
- Perforation of Vessels1
Recent adverse event reports
The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 9, 2026Material Rupture
AN ANGIOSCULPT DEVICE WAS USED FOR A PERIPHERAL THERAPEUTIC PROCEDURE IN THE AV SHUNT. DURING THE FIRST INFLATION, THE BALLOON RUPTURED AT 12 ATM (RBP: 14 ATM). THE PROCEDURE WAS COMPLETED WITH A NEW ANGIOSCULPT. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24832739 ↗MalfunctionFeb 5, 2026Material Rupture
AN ANGIOSCULPT DEVICE WAS USED FOR A PERIPHERAL AV SHUNT PROCEDURE IN THE RADIAL ARTERY. DURING THE FIRST INFLATION, THE BALLOON RUPTURED AT 6 ATM (RBP: 14 ATM). THE PROCEDURE WAS COMPLETED WITH A NEW ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24272697 ↗MalfunctionMar 28, 2025Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED IN A PERIPHERAL PROCEDURE. DURING THE FIRST INFLATION BELOW NOMINAL PRESSURE (8 ATM), THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH THE ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21719396 ↗MalfunctionFeb 20, 2025Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED IN A PERIPHERAL PROCEDURE. DURING THE FIRST INFLATION AT 12 ATM, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NON-ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21426268 ↗MalfunctionDec 27, 2024Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED TO TREAT A CALCIFIED LESION. DURING THE FIRST INFLATION AT 12 ATM, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NON-ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21030628 ↗MalfunctionOct 22, 2024Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED TO TREAT THE SFA. DURING THE FIRST INFLATION AT NOMINAL PRESSURE (8 ATM), THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH THE ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20510919 ↗MalfunctionSep 4, 2024Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED TO TREAT A CALCIFIED PROXIMAL SFA. DURING THE THIRD INFLATION AT 14 ATM, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NEW ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20133320 ↗MalfunctionAug 7, 2024Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED IN A PERIPHERAL PROCEDURE. DURING THE FIRST INFLATION AT 12 ATM, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NON-ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19935003 ↗MalfunctionJun 27, 2024Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED TO TREAT THE MODERATELY CALCIFIED AV FISTULA. DURING THE FIRST INFLATION AT 12 ATM, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NEW ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19629478 ↗MalfunctionMay 3, 2024Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED TO TREAT A SEVERELY CALCIFIED MID RADIAL VEIN. DURING THE SECOND INFLATION AT 16 ATM, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NON-ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED BECAUSE THE BALLOON RUPTURED ABOVE RBP (14 ATM). THERE IS POTENTIAL FOR HARM IF IT WERE TO RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19243831 ↗MalfunctionMay 3, 2024Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED TO TREAT THE SFA ARTERY. DURING INFLATION BELOW NOMINAL PRESSURE, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NEW ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19243726 ↗MalfunctionJun 7, 2023Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED TO TREAT THE DISTAL SFA. DURING THE FIRST INFLATION BELOW RBP (14 ATM), THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NEW ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17083106 ↗MalfunctionMar 3, 2022Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED TO TREAT A MODERATELY TORTUOUS AND SEVERELY CALCIFIED DISTAL SFA. DURING INFLATION AT 12 ATM, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NEW BALLOON. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13660381 ↗MalfunctionFeb 15, 2022Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED TO TREAT A SEVERELY CALCIFIED PERIPHERAL LESION. DURING THE FIRST INFLATION BEFORE REACHING NOMINAL PRESSURE FOR 4 TO 5 SECONDS, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NEW ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13539031 ↗MalfunctionJan 28, 2022Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED TO TREAT THE MID SFA. DURING THE THIRD INFLATION AT 8 ATM, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NEW ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13387292 ↗MalfunctionJan 13, 2022Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED TO TREAT A SEVERELY TORTUOUS AND SEVERELY CALCIFIED MID SFA. DURING THE 3RD INFLATION AT 14 ATM FOR 10 SECONDS, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NEW ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED CONSERVATIVELY BECAUSE THE BALLOON RUPTURED AT RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13256294 ↗InjuryNov 2, 2021Entrapment of Device; Material Rupture; Entrapment of Device; Material Rupture
A STENT PLACEMENT WAS REQUIRED TO TREAT THE HEMOSTASIS THAT RESULTED WITH THE USE OF THE ANGIOSCULPT DEVICE. THE PATIENT WAS DISCHARGED AS SCHEDULED WITHOUT ISSUES. NO PORTION OF THE DEVICE REMAINED IN THE PATIENT.
Patient problems: Perforation of Vessels
MAUDE report 12740890 ↗InjuryAug 2, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE ANGIOSCULPT DEVICE WAS USED TO TREAT A PERIPHERAL LESION ON (B)(6) 2021 WITH NO ISSUES REPORTED. HOWEVER, THE FOLLOWING DAY THE PATIENT EXPERIENCED A THROMBUS OCCLUSION AND EMBOLIZATION IN THE ULNAR ARTERY AND REQUIRED INTERVENTION. THE PATIENT WAS TREATED WITH AN ULTRAVERSE CATHETER WITH NO PATIENT PROBLEM REPORTED.
Patient problems: Embolism/Embolus; Thrombosis/Thrombus
MAUDE report 12264339 ↗InjuryAug 2, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE ANGIOSCULPT DEVICE WAS USED TO TREAT A PERIPHERAL LESION ON (B)(6) 2021 WITH NO ISSUES REPORTED. HOWEVER, THE FOLLOWING DAY THE PATIENT EXPERIENCED A THROMBUS OCCLUSION IN THE ULNAR ARTERY AND REQUIRED INTERVENTION. THE PATIENT WAS TREATED WITH A NSE BALLOON.
Patient problems: Thrombosis/Thrombus
MAUDE report 12264282 ↗MalfunctionJun 3, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT DURING A PERIPHERAL PROCEDURE, THE ANGIOSCULPT DEVICE WAS USED TO TREAT THE MID SFA. DURING THE FIRST INFLATION AT 6 ATM, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NON-ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11936020 ↗MalfunctionJun 3, 2021Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED TO TREAT THE MID CEPHALIC VEIN AND DURING INFLATION, THE BALLOON RUPTURED AT 10 ATM. THE PROCEDURE WAS COMPLETED WITH A NON-ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11935543 ↗MalfunctionFeb 1, 2021Material Rupture; Material Rupture
DURING THE FIRST INFLATION AT 7 ATM, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A COMPETITOR DEVICE. NO PATIENT INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11263459 ↗MalfunctionNov 17, 2020Material Rupture; Material Rupture
DURING THE FIRST INFLATION AT 7 ATM, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH THE SUSPECT DEVICE. NO PATIENT INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10853992 ↗MalfunctionNov 12, 2020Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED TO TREAT A MODERATELY CALCIFIED AND MODERATELY TORTUOUS LESION. DURING THE FIRST INFLATION AT 14 ATM FOR 1 SECOND, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A NEW ANGIOSCULPT DEVICE. NO PATIENT INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10833511 ↗MalfunctionNov 4, 2020Material Rupture; Material Rupture
THE ANGIOSCULPT DEVICE WAS USED IN A SLIGHTLY CALCIFIED AND MODERATELY TORTUOUS DISTAL SFA. DURING THE 8TH INFLATION AT 14 ATM, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A COMPETITOR DEVICE. NO PATIENT INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10785348 ↗
Other Catheter, Percutaneous, Cutting/Scoring devices
- AngioSculpt OTW PTA Scoring Balloon Catheter 2092-6040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring41 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2076-5040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring31 reports
- 2cm Peripheral Cutting Balloon® M001PCB8020900Boston Scientific Corporation · Catheter, percutaneous, cutting/scoring18 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2076-6040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring17 reports
- AngioSculpt OTW PTA Scoring Balloon Catheter 2105-6040The Spectranetics Corporation · Catheter, Percutaneous, Cutting/Scoring17 reports
- ULTRASCORE US14150530Bard Peripheral Vascular, Inc. · Catheter, percutaneous, cutting/scoring15 reports
- ULTRASCORE US14150615Bard Peripheral Vascular, Inc. · Catheter, percutaneous, cutting/scoring14 reports
- ULTRASCORE US3513058Bard Peripheral Vascular, Inc. · Catheter, percutaneous, cutting/scoring14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page