Skip to content

Turbo-Elite Laser Atherectomy Catheter 414-159Recalls & reported problems

Spectranetics Corporation · Class 2

All models of Turbo-Elite Laser Atherectomy Catheter (12)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00813132024727), received between Aug 17, 2021 and Oct 11, 2023 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Wrong Label” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 15, 2023Class IIOpen, Classified

    Incorrect product labeling. Exterior product box label does not match internal pouch label.

    Z-0509-2024 · Root cause: Packaging process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Wrong Label3
  • Break2
  • Material Split, Cut or Torn2
  • Degraded1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Unintended Radiation Exposure4

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 11, 2023Wrong Label; Wrong Label

    DURING PREPARATION FOR USE, IT WAS DISCOVERED THAT UPON OPENING A BOX LABELED WITH A SPECTRANETICS 1.4 TURBO-ELITE LASER ATHERECTOMY CATHETER, THE BOX CONTAINED A 2.0 TURBO-ELITE. THE DEVICE WAS NOT USED IN THE PATIENT, AND THERE WAS NO REPORTED INJURY. THIS REPORT CAPTURES THE LABELING DISCREPANCY THAT A LARGER SIZE DEVICE WAS CONTAINED INSIDE THE BOX, POTENTIAL FOR HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17913496 ↗
  • MalfunctionOct 11, 2023Wrong Label; Wrong Label

    DURING PREPARATION FOR USE, IT WAS DISCOVERED THAT UPON OPENING A BOX LABELED WITH A SPECTRANETICS 1.4 TURBO-ELITE LASER ATHERECTOMY CATHETER, THE BOX CONTAINED A 2.0 TURBO-ELITE. THE DEVICE WAS NOT USED IN THE PATIENT, AND THERE WAS NO REPORTED INJURY. THIS REPORT CAPTURES THE LABELING DISCREPANCY THAT A LARGER SIZE DEVICE WAS CONTAINED INSIDE THE BOX, POTENTIAL FOR HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17913492 ↗
  • MalfunctionOct 11, 2023Wrong Label; Wrong Label

    DURING PREPARATION FOR USE, IT WAS DISCOVERED THAT UPON OPENING A BOX LABELED WITH A SPECTRANETICS 1.4 TURBO-ELITE LASER ATHERECTOMY CATHETER, THE BOX CONTAINED A 2.0 TURBO-ELITE. THE DEVICE WAS NOT USED IN THE PATIENT, AND THERE WAS NO REPORTED INJURY. THIS REPORT CAPTURES THE LABELING DISCREPANCY THAT A LARGER SIZE DEVICE WAS CONTAINED INSIDE THE BOX, POTENTIAL FOR HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17913490 ↗
  • MalfunctionJul 26, 2022Material Split, Cut or Torn; Material Split, Cut or Torn

    A PERIPHERAL ATHERECTOMY PROCEDURE COMMENCED TO TREAT A SLIGHTLY CALCIFIED LESION IN THE PATIENT'S MID POPLITEAL ARTERY. A SPECTRANETICS TURBO ELITE LASER ATHERECTOMY CATHETER WAS USED TO TREAT THE PATIENT. AFTER THE PROCEDURE WAS COMPLETED SUCCESSFULLY, IT WAS NOTICED THAT THE TURBO ELITE WAS SLICED OR COMING APART. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO REPORTED PATIENT HARM. THIS EVENT IS BEING REPORTED FOR UNINTENDED RADIATION EXPOSURE, POTENTIAL FOR HARM.

    Patient problems: Unintended Radiation Exposure

    MAUDE report 15108431 ↗
  • MalfunctionJun 14, 2022Break; Break

    A PERIPHERAL ATHERECTOMY PROCEDURE COMMENCED TO TREAT A SLIGHTLY CALCIFIED, PLAQUE LESION IN THE PATIENT'S DISTAL POSTERIOR TIBIAL ARTERY. THE PHYSICIAN CHOSE TO USE A SPECTRANETICS TURBO ELITE LASER ATHERECTOMY CATHETER TO TREAT THE PATIENT. AFTER THE TREATMENT HAD BEEN COMPLETED, THE TURBO ELITE WAS REMOVED FROM THE OUTER SHEATH AND THE PHYSICIAN NOTICED THE FIBERS OF THE TURBO ELITE WERE EXPOSED. THE PROCEDURE WAS COMPLETED WITH NO REPORTED PATIENT HARM. THIS EVENT IS BEING REPORTED FOR UNINTENDED RADIATION EXPOSURE, POTENTIAL FOR HARM.

    Patient problems: Unintended Radiation Exposure

    MAUDE report 14688935 ↗
  • MalfunctionMay 11, 2022Degraded; Degraded

    A PERIPHERAL ATHERECTOMY PROCEDURE COMMENCED TO TREAT A MODERATELY CALCIFIED LESION IN THE PATIENT'S DISTAL ANTERIOR TIBIAL (AT) VESSEL. THE PHYSICIAN CHOSE A SPECTRANETICS TURBO ELITE LASER ATHERECTOMY CATHETER TO TREAT THE PATIENT. DURING USE IN THE PROCEDURE, IT WAS REPORTED THE TIP OF THE CATHETER AND THE SHEATH APPEARED BURNED. A NEW TURBO ELITE OF THE SAME MODEL WAS USED TO COMPLETE THE PROCEDURE WITH NO REPORTED PATIENT HARM. THIS EVENT IS BEING REPORTED FOR UNINTENDED RADIATION EXPOSURE, POTENTIAL FOR HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unintended Radiation Exposure

    MAUDE report 14366167 ↗
  • MalfunctionAug 17, 2021Break; Material Split, Cut or Torn; Break; Material Split, Cut or Torn

    A PERIPHERAL ATHERECTOMY PROCEDURE COMMENCED TO TREAT A SLIGHTLY CALCIFIED LESION IN THE PATIENT'S MID SUPERFICIAL FEMORAL ARTERY (SFA). THE PHYSICIAN CHOSE TO USE A SPECTRANETICS TURBO ELITE LASER ATHERECTOMY CATHETER TO TREAT THE PATIENT. AFTER THE PHYSICIAN USED THE TURBO ELITE DEVICE IN THE PATIENT AND WAS REMOVING THE DEVICE, HE NOTICED THE "OUTER SHEATH WAS TORN" AND FIBERS WERE VISIBLE. IT WAS REPORTED THAT THERE WAS DIFFICULTY GETTING THE TURBO ELITE DEVICE THROUGH THE SHEATH AND BELIEVED THAT MAY HAVE BEEN THE REASON THE DEVICE WAS DAMAGED. A NEW SPECTRANETICS TURBO ELITE LASER ATHERECTOMY CATHETER WAS USED TO FINISH THE PROCEDURE, AND THERE WAS NO REPORTED PATIENT HARM. THIS EVENT IS BEING SUBMITTED TO CAPTURE THE TURBO ELITE WHICH WAS FOUND DAMAGED DURING USE AND RESULTED IN UNINTENDED RADIATION EXPOSURE WITH POTENTIAL FOR HARM.

    Patient problems: Unintended Radiation Exposure

    MAUDE report 12333209 ↗

Other Catheter, Peripheral, Atherectomy devices

All Catheter, Peripheral, Atherectomy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page