Turbo-Power Laser Atherectomy Catheter 423-050Recalls & reported problems
Spectranetics Corporation · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00813132026776), received between Aug 31, 2022 and Nov 9, 2022 — classified as 2 injuries.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Restenosis1
- Thromboembolism1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 9, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS PART OF A JAPAN TURBO POSTMARKET STUDY, A PERIPHERAL ATHERECTOMY PROCEDURE BEGAN TO TREAT AN IN-STENT RESTENOSIS (ISR) IN THE PATIENT'S MID SUPERFICIAL FEMORAL ARTERY (SFA) TO THE POPLITEAL REGION. THE LESION TO BE TREATED WAS 400 MM IN LENGTH, WITH HEAVY THROMBUS PRESENT WITHIN THE STENTS. IT WAS PREDICTED THAT TREATMENT WOULD RESULT IN CLOTS AND POTENTIAL EMBOLISM; THEREFORE, A FILTER WAS PLACED PRE-PROCEDURE, DISTAL TO THE LESION. THE PHYSICIAN CHOSE TO TREAT THE PATIENT WITH A SPECTRANETICS TURBO-POWER LASER ATHERECTOMY CATHETER. DURING USE, AND AS EXPECTED, PORTIONS OF THE THROMBUS WERE OBSERVED. THE FILTER DID NOT CATCH ALL THE THROMBUS, RESULTING IN A PERIPHERAL EMBOLISM. THE THROMBUS THAT WAS NOT CAPTURED BY THE FILTER WAS REMOVED SUCCESSFULLY USING A MECHANICAL THROMBECTOMY CATHETER. THE PROCEDURE WAS COMPLETED WITH NO REPORTED PATIENT HARM. THIS REPORT CAPTURES THE TURBO-POWER IN USE WITHIN THE LESION WHEN THE PERIPHERAL EMBOLISM OCCURRED, REQUIRING INTERVENTION. THERE WAS NO ALLEGED MALFUNCTION OF ANY SPECTRANETICS DEVICES IN USE DURING THE PROCEDURE.
Patient problems: Thromboembolism
MAUDE report 15765107 ↗InjuryAug 31, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A PERIPHERAL ATHERECTOMY PROCEDURE COMMENCED ON (B)(6) 2022 TO TREAT A SEVERELY CALCIFIED LESION IN THE PATIENT'S LEFT PROXIMAL SUPERFICIAL FEMORAL ARTERY (SFA). A STENT HAD BEEN PLACED IN THIS AREA PRIOR TO THE PROCEDURE, AND THE PATIENT WAS EXPERIENCING A WOUND AND NUMBNESS IN THE LEFT LOWER EXTREMITY. THE PHYSICIAN CHOSE A SPECTRANETICS TURBO-POWER LASER ATHERECTOMY CATHETER TO TREAT THE PATIENT, FOLLOWED BY BALLOON THERAPY, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. FOLLOWING THE INITIAL PROCEDURE, THE PATIENT WAS TREATED WITH ANTIMICROBIALS AND ANTICOAGULANTS. HOWEVER, SOME NUMBNESS REMAINED AFTER THE PROCEDURE, SO THE PATIENT WAS CHECKED DAILY BY DOPPLER ULTRASONOGRAPHY AND PALPATION. OVER THE NEXT FEW DAYS, TESTING SHOWED A DECREASED LEFT ANKLE/BRACHIAL INDEX (ABI) WHICH CAN INDICATE WORSENING PERIPHERAL ARTERY DISEASE (PAD). ON (B)(6) 2022, AN ECHO WAS PERFORMED AND THE FOLLOWING DAY, THE PHYSICIAN DIAGNOSED AN IN-STENT RESTENOSIS (ISR) IN THE SAME REGION TREATED INITIALLY. ON(B)(6) 2022, A PROCEDURE COMMENCED TO TREAT THE ISR, USING A TURBO-POWER, A PLAIN OLD BALLOON ANGIOPLASTY (POBA), AND A STENT GRAFT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO REPORTED PATIENT
Patient problems: No Clinical Signs, Symptoms or Conditions; Restenosis
MAUDE report 15332257 ↗
Other Catheter, Peripheral, Atherectomy devices
- HawkOne™ H1-MCovidien LP · Catheter, peripheral, atherectomy1 recall367 reports
- JETSTREAM™ XC 112266-001Boston Scientific Corporation · Catheter, peripheral, atherectomy288 reports
- JETSTREAM™ XC 112264-001Boston Scientific Corporation · Catheter, peripheral, atherectomy262 reports
- SET Rotarex®S 6F 135cm (Straub Medical®) Version 1.0Straub Medical AG · Catheter, Peripheral, Atherectomy2 deaths206 reports
- SET Rotarex™S 6 F x 135 cm 80237Straub Medical AG · Catheter, Peripheral, Atherectomy1 recall79 reports
- HawkOne™ H1-SCovidien LP · Catheter, peripheral, atherectomy1 recall70 reports
- Turbo-Elite Laser Atherectomy Catheter 420-006Spectranetics Corporation · Catheter, Peripheral, Atherectomy2 recalls19 reports
- Turbo-Elite Laser Atherectomy Catheter 417-152Spectranetics Corporation · Catheter, Peripheral, Atherectomy2 recalls16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page