Skip to content

Stellarex AB35SX060150135Recalls & reported problems

Spectranetics Corporation · Stellarex 0.035 OTW Drug-coated Angioplasty Balloon (6mm x 150mm) x 135cm · Class 3

All models of Stellarex (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00813132026974), received between Apr 12, 2021 and Jun 15, 2022 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Rupture2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 15, 2022Material Rupture; Material Rupture

    TWO STELLAREX DEVICES WERE USED IN A SLIGHTLY TORTUOUS AND MODERATELY CALCIFIED SFA. DURING INFLATION OF THE FIRST STELLAREX DEVICE, THE BALLOON RUPTURED AT 6 ATM. THE BALLOON WAS REMOVED IN ONE PIECE AND A NEW STELLAREX DEVICE WAS USED. DURING INFLATION OF THE SECOND STELLAREX DEVICE (MDR 3011416935-2022-00011), THE BALLOON RUPTURED AT 5 ATM. THE BALLOON WAS ALSO REMOVED IN ONE PIECE AND THE PROCEDURE WAS COMPLETED WITH A NON-STELLAREX DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED PER FDA REQUEST BECAUSE THE BALLOON RUPTURED BELOW RBP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14709407 ↗
  • MalfunctionApr 12, 2021Material Rupture; Material Rupture

    FROM A CONTRALATERAL FEMORAL APPROACH, THE STELLAREX DEVICE WAS USED TO TREAT THE PROXIMAL POPLITEAL ARTERY. DURING INFLATION AT 1 ATM, CONTRAST WAS VISIBLY LEAKING OUT OF THE BALLOON. THE BALLOON WAS REMOVED FROM THE BODY, INFLATED ON THE TABLE AND A PINHOLE ON THE BALLOON WAS OBSERVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11653902 ↗