Stellarex AB35SX060200135 (UDI 00813132026981)Recalls & reported problems
Spectranetics Corporation · Stellarex 0.035 OTW Drug-coated Angioplasty Balloon (6mm x 200mm) x 135cm · Class 3
All models of Stellarex (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00813132026981), received between May 21, 2021 and Jun 15, 2022 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Rupture4
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 15, 2022Material Rupture; Material Rupture
TWO STELLAREX DEVICES WERE USED IN A SLIGHTLY TORTUOUS AND MODERATELY CALCIFIED SFA. DURING INFLATION OF THE FIRST STELLAREX DEVICE (MDR 3011416935-2022-00010), THE BALLOON RUPTURED AT 6 ATM. THE BALLOON WAS REMOVED IN ONE PIECE AND A NEW STELLAREX DEVICE WAS USED. DURING INFLATION OF THE SECOND STELLAREX DEVICE, THE BALLOON RUPTURED AT 5 ATM. THE BALLOON WAS ALSO REMOVED IN ONE PIECE AND THE PROCEDURE WAS COMPLETED WITH A NON-STELLAREX DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED PER FDA REQUEST BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14709309 ↗MalfunctionJun 1, 2022Material Rupture; Material Rupture
THE STELLAREX DEVICE WAS USED IN A SEVERELY CALCIFIED PROXIMAL SFA. DURING INFLATION AT 2 ATM, THE BALLOON RUPTURED. THE BALLOON WAS REMOVED IN ONE PIECE AND THE PROCEDURE WAS COMPLETED WITH A NEW STELLAREX DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED PER FDA REQUEST BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14569454 ↗MalfunctionOct 26, 2021Material Rupture; Material Rupture
THE STELLAREX DEVICE WAS USED TO TREAT A MODERATELY TORTUOUS AND A SEVERELY CALCIFIED MID SFA. DURING INFLATION AT 2 ATM, THE BALLOON BURST DUE TO HEAVILY CALCIFIED EDGES. THE BALLOON WAS REMOVED IN ONE PIECE AND THE PROCEDURE WAS COMPLETED WITH A NEW STELLAREX DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING SUBMITTED PER FDA REQUEST BECAUSE THE BALLOON RUPTURED BELOW RBP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12701517 ↗MalfunctionMay 21, 2021Material Rupture; Material Rupture
FROM A RETROGRADE FEMORAL APPROACH, THE STELLAREX DEVICE WAS USED TO TREAT A SEVERELY CALCIFIED MID SFA. DURING INFLATION AT 8 ATM, THE DEVICE BURST. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11864273 ↗
Other Drug-Eluting Peripheral Transluminal Angioplasty Catheter devices
- LUTONIX® Drug Coated Balloon LX351305150Lutonix, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter117 reports
- LUTONIX® Drug Coated Balloon LX351306150Lutonix, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter75 reports
- Ranger™ H74939419502010Boston Scientific Corporation · Drug-Eluting Peripheral Transluminal Angioplasty Catheter71 reports
- Ranger™ H74939419602010Boston Scientific Corporation · Drug-Eluting Peripheral Transluminal Angioplasty Catheter71 reports
- Lutonix® Drug Coated Balloon LX351305100Lutonix, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter66 reports
- IN.PACT™ Admiral™ ADM06004013PMedtronic, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter1 recall12 reports
- IN.PACT™ Admiral™ ADM06015013PMedtronic, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter1 recall11 reports
- IN.PACT™ Admiral™ ADM04015013PMedtronic, Inc. · Drug-Eluting Peripheral Transluminal Angioplasty Catheter1 recall8 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page