Skip to content

Quick-Cross Support Catheter 518-065 (UDI 00813132029395)Recalls & reported problems

Philips Image Guided Therapy Corporation · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00813132029395), received between Jul 8, 2025 and Jul 8, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 8, 2025Material Separation; Material Separation

    A PERIPHERAL ATHERECTOMY PROCEDURE COMMENCED TO TREAT A LESION IN THE SUPERFICIAL FEMORAL ARTERY (SFA) AND CHRONIC TOTAL OCCLUSION (CTO) IN THE POPLITEAL ARTERY/TIBIOPERONEAL TRUNK. A PHILIPS QUICK-CROSS SUPPORT CATHETER WAS USED TO TREAT THE PATIENT. WHILE USING A CONTRALATERAL APPROACH, HEMOSTATS WERE USED TO CLAMP DOWN THE TERUMO 0.014 GUIDEWIRE, BUT THE QUICK-CROSS GOT CLAMPED AS WELL. UPON REMOVAL OF THE QUICK-CROSS, RESISTANCE WAS MET BUT PULLING CONTINUED. AT THAT TIME, THE QUICK-CROSS BROKE, AND REMAINED ON THE GUIDE WIRE. THE HEMOSTATS WERE REMOVED, AND WHEN ATTEMPTING TO PUSH THE GUIDEWIRE FURTHER THROUGH THE INTRODUCER SHEATH AND REMOVING THE QUICK-CROSS OFF THE GUIDEWIRE, THE DEVICE CONTINUED TO BREAK INTO SEVERAL PIECES. HOWEVER, ALL OF THE FRAGMENTS WERE REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING A BALLOON AND SHOCKWAVE DEVICE. NO PATIENT INJURY REPORTED. THIS REPORT CAPTURES THE QUICK-CROSS WHICH SEPARATED AT THE SHAFT, POTENTIAL FOR HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22442833 ↗

Other Catheter, Percutaneous devices

All Catheter, Percutaneous devices →

Sources and updates