Theragenics Corporation · TheraSleeve with x = 1-12 seeds and 0 or more markers · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00813787020433), received between Oct 17, 2024 and Oct 17, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Device Markings/Labelling Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Markings/Labelling Problem2
  • Wrong Label2

Patient problems

  • Unintended Radiation Exposure2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 17, 2024Device Markings/Labelling Problem; Wrong Label; Device Markings/Labelling Problem; Wrong Label

    THE NON-STERILIZED SEEDS AND PAPERWORK INCLUDED IN THE ORDERS WERE LABELED CORRECTLY, BUT THE PRE-STERILIZED NEEDLE TRAYS WERE ALL LABELED INCORRECTLY. THE LABELING OF THE PRE-STERILIZED NEEDLE TRAYS FOR ORDERS (B)(6) WERE INADVERTENTLY SWITCHED WHEN THE LABELING FOR THE NEEDLE TRAYS WERE APPLIED. THERE WERE TWO (2) PRE-STERILIZED NEEDLE TRAYS PER ORDER THAT WERE LABELED INCORRECTLY. SEE ORDERS BELOW: ORDER (B)(6) PATIENT (B)(6). COMPLAINANT SUBMITTED VARISEED NEEDLE LOADING REPORT/PLAN: AGX100 I-125. 0.237 MCI PER SEED. (B)(4) SEEDS, (B)(4) PRE-LOADED NEEDLES. TOTAL ACTIVITY PER PLAN: 20.856 MCI. ORDER (B)(6) PATIENT (B)(6). COMPLAINANT SUBMITTED VARISEED NEEDLE LOADING REPORT/PLAN: AGX100 I-125. 0.357 MCI PER SEED. (B)(4) SEEDS, (B)(4) PRE-LOADED NEEDLES. TOTAL ACTIVITY PER PLAN: 27.846 MCI.

    Patient problems: Unintended Radiation Exposure

    MAUDE report 20470426 ↗
  • InjuryOct 17, 2024Device Markings/Labelling Problem; Wrong Label; Device Markings/Labelling Problem; Wrong Label

    THE NON-STERILIZED SEEDS AND PAPERWORK INCLUDED IN THE ORDERS WERE LABELED CORRECTLY, BUT THE PRE-STERILIZED NEEDLE TRAYS WERE ALL LABELED INCORRECTLY. THE LABELING OF THE PRE-STERILIZED NEEDLE TRAYS FOR ORDERS (B)(6) WERE INADVERTENTLY SWITCHED WHEN THE LABELING FOR THE NEEDLE TRAYS WERE APPLIED. THERE WERE TWO (2) PRE-STERILIZED NEEDLE TRAYS PER ORDER THAT WERE LABELED INCORRECTLY. SEE ORDERS BELOW: ORDER (B)(6) PATIENT (B)(6). COMPLAINANT SUBMITTED VARISEED NEEDLE LOADING REPORT/PLAN: AGX100 I-125. 0.357 MCI PER SEED. 78 SEEDS, 28 PRE-LOADED NEEDLES. TOTAL ACTIVITY PER PLAN: 27.846 MCI . ORDER (B)(6) PATIENT (B)(6). COMPLAINANT SUBMITTED VARISEED NEEDLE LOADING REPORT/PLAN: AGX100 I-125. 0.237 MCI PER SEED. 88 SEEDS, 28 PRE-LOADED NEEDLES. TOTAL ACTIVITY PER PLAN: 20.856 MCI.

    Patient problems: Unintended Radiation Exposure

    MAUDE report 20470401 ↗

Other Source, Brachytherapy, Radionuclide devices