Bio-Med Devices, Inc. 2003FFBRecalls & reported problems

Bio-Med Devices, Inc. · Air / Oxygen Blender low flow with flowmeters · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00813841021192), received between Nov 3, 2016 and Nov 3, 2016 — classified as 1 malfunction.

The most frequently reported specific device problem is “Bent” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Bent1
  • Improper Device Output1
  • Physical Property Issue1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 3, 2016Bent; Improper Device Output; Physical Property Issue

    BLENDER WAS PULLED OUT OF SERVICE WHEN IT WAS DETERMINED THAT THE OXYGEN % OUTPUT DID NOT MATCH KNOB SET. HOSPITAL CLINICAL ENGINEERING FOUND THE KNOB DAMAGED, WITH SHAFT BENT. BLENDER WAS RETURNED TO MFR FOR REPAIR.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6075988 ↗

Other Accessory To Continuous Ventilator (Respirator) devices

All Accessory To Continuous Ventilator (Respirator) devices →