Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00813954020273), received between Oct 5, 2017 and Jul 22, 2025 — classified as 3 injuries.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 13, 2021Terminated

    Outer packaging (outer pouch) is not sealed compromising sterility and result in potential risk of patient infection, which could lead to revision surgery

    Z-0668-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • Abscess1
  • Cerebrospinal Fluid Leakage1
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    COMPLAINT SUBMITTED THROUGH CLINICAL DATA COLLECTION, WITH LIMITED DETAILS PROVIDED. FURTHER INFORMATION REGARDING THE PATIENT OR EVENT IS UNLIKELY TO BECOME AVAILABLE. THE PRODUCT INVOLVED IS DURAMATRIX-ONLAY®. A PATIENT EXPERIENCED AN ADVERSE EVENT (AE) CHARACTERIZED AS WOUND DEHISCENCE WITH PURULENT DISCHARGE. THE INITIAL SURGERY WAS (B)(6) 2025 AND THE LOCATION OF THE IMPLANT WAS THE ANTERIOR FOSSA. NO SUTURES, SEALANTS OR OTHER DURAL CLOSURE PRODUCTS WERE USED DURING THE PROCEDURE. NO CSF LEAK WAS OBSERVED POST-PLACEMENT. A DRAINAGE SHUNT WAS USED. A ROUTINE FOLLOW UP VISIT ON (B)(6) 2025 NOTED NO ADVERSE EVENTS. HOWEVER, A SUBSEQUENT FOLLOW-UP VISIT ON (B)(6) 2025 INDICATED WOUND DEHISCENCE ON THE "SCALP, WOUND". THE PATIENT WAS INITIALLY PRESCRIBED CEPHALEXIN WHICH WAS SWITCHED TO DOXYCYCLINE DUE TO SENSITIVITIES. CLINICAL DATA COLLECTION INDICATED THAT THE ADVERSE EVENT WAS UNRELATED TO THE INTERVENTION. NO SURGICAL INTERVENTION WAS REQUIRED. NO FURTHER INFORMATION SURROUNDING PATIENT STATUS OR PATIENT OUTCOME WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 22568587 ↗
  • InjuryJul 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    COMPLAINT SUBMITTED THROUGH CLINICAL DATA COLLECTION. LIMITED INFORMATION PROVIDED AND IT IS UNLIKELY FURTHER INFORMATION SURROUNDING THE PATIENT OR THE EVENT WILL BE PROVIDED. DURAMATRX-ONLAY COLLAGEN DURA MEMBRANE USED ON A PATIENT ON (B)((6) 2025 IN THE ANTERIOR FOSSA. PATIENT WAS PART OF CLINICAL STUDY FOR MENINGIOMA WHO GRADE 1. AT TIME OF INITIAL SURGERY, 3-0 AND 4-0 SYNTHETIC, NON-ABSORBABLE SUTURES WERE USED (NOT COLLAGEN MATRIX, INC. PRODUCTS). NO OTHER DURAL PRODUCTS WERE UTILIZED. A DRAINAGE SHUNT WAS ALSO USED (NOT A COLLAGEN MATRIX, INC. PRODUCT). THERE WERE NO COMPLICATIONS NOTED DURING THE INITIAL SURGERY. THERE WAS NO CSF LEAKAGE AFTER INITIAL PLACEMENT. THE CLINICIAN CLASSIFIED THE HANDLING PROPERTIES OF THE DURAMATRIX-ONLAY PRODUCT AS "GOOD" AND ALSO CLASSIFIED THE ADAPTABILITY OF THE PRODUCT TO ANATOMIC STRUCTURE AS "GOOD." AN ADVERSE EVENT WAS REPORTED ON 5/31/2025, IN WHICH AN INTRACRANIAL ABSCESS POST CRANIECTOMY IN THE BONE FLAP WAS NOTED. CLINICIAN CLASSIFIED THIS AS A SERIOUS, BUT UNEXPECTED, ADVERSE EVENT. HOSPITALIZATION WAS REQUIRED. INTERVENTION REQUIRED AS ADDITIONAL SURGERY WAS PERFORMED TO RE-DO THE CRANIECTOMY AND EVACUATION OF ABSCESS AND WASHOUT.

    Patient problems: Abscess

    MAUDE report 22560140 ↗
  • InjuryOct 5, 2017Appropriate Device Problem Term/Code Not Available

    IT WAS INITIALLY REPORTED THAT THE DURAMATRIX ONLAY (CDSLM45) PRODUCT IN COMBINATION WITH DURAMATRIX SUTURABLE (DMS45; EMDR REPORT 2249852-2017-00017) WAS USED AS PART OF A POST AREA FOSSA CRANIOTOMY PROCEDURE WHICH RESULTED IN FOLLOW-UP SURGERY TO FIX A CEREBROSPINAL FLUID LEAK. BASED ON ADDITIONAL INFORMATION RECEIVED ON 10/11/2018, IT HAS BEEN DETERMINED THAT THE DURAMATRIX ONLAY (CDSML45) PRODUCT WAS NOT USED OR ASSOCIATED WITH THE INITIALLY REPORTED INCIDENT.

    Patient problems: Cerebrospinal Fluid Leakage

    MAUDE report 6915042 ↗