Dovring - 1.6Recalls & reported problems
Quality Aspirators, Inc. · Dovring - 1.6 · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00814005022734), received between Apr 9, 2026 and Apr 9, 2026 — classified as 1 injury.
The most frequently reported specific device problem is “Loss of Osseointegration” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Loss of Osseointegration1
Patient problems
- Failure of Implant1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 9, 2026Loss of Osseointegration
IT WAS REPORTED THAT A PATIENT EXPERIENCED A DENTAL IMPLANT LOSS.
Patient problems: Failure of Implant
MAUDE report 24836965 ↗
FDA clearance and classification
The FDA regulates product code EIF (accessories, retractor, dental) under 21 CFR 872.4565 as a Class I device. 21 CFR 872.4565 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Dovring - 1.6
Has the Dovring - 1.6 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00814005022734), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Dovring - 1.6?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Apr 9, 2026. That report lists device problem “Loss of Osseointegration” (1); patient problem “Failure of Implant” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Dovring - 1.6?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Dovring - 1.6 FDA approved?
Devices of this type (product code EIF) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Quality Aspirators, Inc.
Is the Dovring - 1.6 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Dovring - 1.6, and is it still sold?
The FDA's device database lists Quality Aspirators, Inc. as the company responsible for it (brand Dovring - 1.6, model Dovring - 1.6). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page