Skip to content

ClearView Total Uterine Manipulator UM750Recalls & reported problems

Clinical Innovations, LLC · ClearView Total Uterine Manipulator, regular tip (7cm) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00814247020499), received between May 13, 2019 and May 16, 2019 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Component Missing” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Component Missing2
  • Manufacturing, Packaging or Shipping Problem1
  • Structural Problem1
  • Therapeutic or Diagnostic Output Failure1

Patient problems

  • No Consequences Or Impact To Patient4
  • Injury1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 16, 2019Therapeutic or Diagnostic Output Failure

    BEFORE USE, THERE IS NO GROOVE FOR FIXING THE COLPOTOMY CUP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8617037 ↗
  • MalfunctionMay 16, 2019Manufacturing, Packaging or Shipping Problem

    NO GROOVE TO ATTACH THE COLPOTOMY CUP. OUR QA PEOPLE CHECKED ALL CLEARVIEW TOTAL DEVICES AND FOUND 1 THAT HAD EXACTLY THE SAME PROBLEM THAT WE REPORTED LAST WEEK. THIS DEVICE WAS STORED IN THE WAREHOUSE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8616997 ↗
  • MalfunctionMay 16, 2019Component Missing

    NO GROOVE TO ATTACH THE COLPOTOMY CUP. OUR QA PEOPLE CHECKED ALL CLEARVIEW TOTAL DEVICES AND FOUND 1 THAT HAD EXACTLY THE SAME PROBLEM THAT WE REPORTED LAST WEEK. THIS DEVICE WAS STORED IN THE WAREHOUSE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8616956 ↗
  • MalfunctionMay 16, 2019Structural Problem

    NO GROOVE TO ATTACH THE COLPOTOMY CUP. OUR QA PEOPLE CHECKED ALL CLEARVIEW TOTAL DEVICES AND FOUND 1 THAT HAD EXACTLY THE SAME PROBLEM THAT WE REPORTED LAST WEEK. THIS DEVICE WAS STORED IN THE WAREHOUSE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8616917 ↗
  • InjuryMay 13, 2019Component Missing; Component Missing

    THIS IS A REPORT ON A HEALTH INJURY CAUSED BY YOUR DEFECTIVE CLEARVIEW TOTAL WITH NO NOTCH (A GROOVE TO FIX THE COLPOTOMY CUP TIGHTLY). THE DOCTOR TRIED TO INSERT THE DEVICE WITH THE CUP BEING INSUFFICIENTLY FIXED INTO A PATIENT AND HURT HER VAGINAL PORTION OF CERVIX DUE TO NO SULCUS FIXATION APPLIED.

    Patient problems: Injury

    MAUDE report 8606170 ↗

Other Cannula, Manipulator/Injector, Uterine devices

All Cannula, Manipulator/Injector, Uterine devices →