Kiwi® OmniCup® VAC-6000MTRecalls & reported problems
Clinical Innovations, LLC · Kiwi® Vacuum Delivery System, OmniCup® with Traction Force Indicator · Class 2
All models of Kiwi OmniCup (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 1
- Injuries reported
- 5
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00814247020642), received between Sep 18, 2017 and Nov 22, 2024 — classified as 1 death, 5 injuries, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K981260); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 9, 2026Class IIOpen, Classified
Due to complaints of device breakage at the traction force gauge to handle joint.
Z-1945-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information4
- Adverse Event Without Identified Device or Use Problem1
- Human-Device Interface Problem1
- Use of Device Problem1
Patient problems
- Abrasion1
- Encephalopathy1
- Hematoma1
- Hemorrhage/Blood Loss/Bleeding1
- Injury1
- Insufficient Information1
- Intracranial Hemorrhage1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 22, 2024
WHEN APPLYING THE VACUUM, AND SUCTION WAS ATTEMPTED, THE DEVICE "FELL APART".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20757030 ↗InjuryNov 5, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
PROLONGED 2ND STAGE. 2 PULLS GIVEN. KIWI CAME OFF BEGINNING OF 2ND PULL. SUBGALEAL HEMATOMA.
Patient problems: Hematoma
MAUDE report 20612399 ↗InjuryNov 5, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ABRASIONS WERE MADE ON THE ADSORPTION AREA (HEAD) OF THE CHILD.
Patient problems: Abrasion
MAUDE report 20612225 ↗DeathMar 15, 2023Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
IN THE CUSTOMER'S WORDS: "OPERATIVE VAGINAL DELIVERY VIA KIWI VACUUM. INFANT WITH SUBGALEAL BLEED".
Patient problems: Hemorrhage/Blood Loss/Bleeding; Insufficient Information
MAUDE report 16546287 ↗InjuryAug 30, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
IN THE CUSTOMERS WORD: "THE KIWI VACCUM DEVICE WAS NOT ABLE TO MAINTAIN SUCTION PRESSURE AFTER PUMPED UP TO 65KPA, CREATES HISSING SOUND AND SUCTION PRESSURE QUICKLY REDUCE DOWN TO 15KPA". "2ND KIWI VACCUM DEVICE NEED TO BE USED TO EFFECT DELIVERY OF THE BABY AND THIS CAUSED EXTENSIVE SUBAPONEUROTIC HAEMORRHAGE (SAH) AND HYPOXIC ISCHAEMIC ENCEPHALOPATHY (HIE)" "VACCUM ASSISTED DELIVERY FOR FETAL BRADYCARDIA". "1ST KIWI VACCUUM DEVICE APPLIED AT 0215H, TESTED PRIOR TO APPLICATION ON FETAL HEAD AND SUCTION PRESENT, ABLE TO CREATE SUCTION PRESSURE UP TO 65KPA HOWEVER UPON TRACTION NOTED HISSING SOUND AND NOTICED SUCTION PRESSURE DROPPED TO 15KPA, REPEATED PUMP SUCTION ATTEMPTED BUT SIMILAR PROBLEM. 2ND KIWI DEVICE WAS OBTAINED, TESTED AND FUNCTIONING WELL, APPLIED AT 0218H, SUCTION MAINTAINED AND SUCCESSFUL DELIVERY AT 0222H". "BABY WAS BORN PALE REQUIRING INOTROPE SUPPORT DUE TO EXTENSIVE SAH, ALSO REQUIRED PACKED CELL AND FRESH FROZEN PLASMA TRANSFUSION. ALSO SUSTAINED HIE GRADE 2".
Patient problems: Encephalopathy; Intracranial Hemorrhage
MAUDE report 15321973 ↗InjuryMay 22, 2019Human-Device Interface Problem; Human-Device Interface Problem
KIWI APPLIED CORRECTLY THEN DEFLATED AT FIRST ATTEMPT TO PULL. NOT A POP-OFF. THERE WAS CAPUT THIS TIME SO VACUUM HAD TO BE ACHIEVED. IT WAS REPORTED THE PATIENT WAS TRAUMATIZED BUT NO FURTHER DETAIL PROVIDED. THE PATIENT UNDERWENT A C-SECTION UNDER GENERAL ANESTHESIA. NO SAMPLE AVAILABLE.
Patient problems: Injury
MAUDE report 8633595 ↗InjurySep 18, 2017Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE BABY HAD A SKULL FRACTURE POST USE.
Patient problems: Skull Fracture
MAUDE report 6874427 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K981260. One submission often covers many models, so check the product and lot details in each record.
- Z-2072-2020 · Aug 6, 2019 · TerminatedKiwi Complete Vacuum Delivery System, Model No. VAC-6000MTE - Product Usage: The Kiwi is a disposable vacuum-assisted fetal delivery system.
Other Extractor, Vacuum, Fetal devices
- Kiwi® OmniCup® VAC-6000MTEClinical Innovations, LLC · Extractor, Vacuum, Fetal1 recall1 death20 reports
- Mystic 10057Coopersurgical, Inc. · Extractor, Vacuum, Fetal19 reports
- Kiwi® OmniCup® VAC-6000MClinical Innovations, LLC · Extractor, Vacuum, Fetal14 reports
- Kiwi® OmniCup® VAC-6000MEClinical Innovations, LLC · Extractor, Vacuum, Fetal10 reports
- MityOne 10067Coopersurgical, Inc. · Extractor, Vacuum, Fetal10 reports
- Mityvac 10137Coopersurgical, Inc. · Extractor, Vacuum, Fetal4 reports
- Kiwi® ProCup® VAC-6000SClinical Innovations, LLC · Extractor, Vacuum, Fetal3 reports
- MityOne 10068Coopersurgical, Inc. · Extractor, Vacuum, Fetal2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page