ACCUPULSE SINGLE LUMEN CONSERVING REGULATOR IN6505-L-BLUERecalls & reported problems
Inovo, Inc. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00814470020051), received between May 31, 2017 and May 31, 2017 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Fire” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Fire1
Patient problems
- Abrasion1
- Injury1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 31, 2017Fire; Fire
IT WAS REPORTED THAT THE CUSTOMER WAS IN A VEHICLE AT THE TIME OF THE INCIDENT AND WAS SWITCHING OVER THE DEVICE FROM AN EMPTY CYLINDER TO A FULL CYLINDER. THE DEVICE WAS CONNECTED AND AFTER TURNING ON THE CYLINDER VALVE, A LEAK WAS NOTICED. THE CYLINDER VALVE WAS TURNED OFF AND RECONNECTED TO THE DEVICE TWO MORE TIMES. ON THE LAST ATTEMPT, A HISS WAS HEARD THAT GOT LOUDER, AND THEN FLASHED WITH A FLAME. THE CUSTOMER JUMPED OUT OF THE CAR AND THEN QUICKLY TURNED THE CYLINDER VALVE OFF TO PREVENT A FIRE. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6602668 ↗InjuryMay 31, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
(B)(6) (A THIRD PARTY ADMINISTRATOR OF INSURANCE FOR A DURABLE MEDICAL EQUIPMENT (DME) COMPANY) REPORTED THAT (B)(6) (CLAIMANT) WAS REMOVING AN INOVO ACCUPULSE REGULATOR DEVICE FROM AN OXYGEN CYLINDER WHEN THE O-RING BLEW OFF, STRIKING HER IN THE HAND. THE O-RING MENTIONED BY THE COMPLAINANT IS KNOWN AS THE SEAL WASHER AND JOINS SYSTEMS OR MECHANISMS TOGETHER BY PREVENTING LEAKAGE. THE EXPLANATION OF THE EVENT FROM THE ESIS REPORT INDICATES THAT THE ACCUPULSE REGULATOR WAS STILL UNDER PRESSURE WHEN MS. (B)(6) ATTEMPTED TO LOOSEN/REMOVE THE DEVICE FROM THE CYLINDER. THE LOOSENING OF THE REGULATOR CONNECTION FROM THE CYLINDER 870 POST VALVE CAUSED THE IMMEDIATE RELEASE OF HIGH PRESSURE GAS FROM THE OUTLET ORIFICE OF THE OXYGEN CYLINDER. THE HIGH FLOW AND PRESSURE OF GAS ESCAPING FROM THE CYLINDER DISLODGED THE SEAL WASHER FROM THE REGULATOR, AND CAUSED THE LOUD "POP" SOUND DESCRIBED IN THE REPORT. ATTEMPTING TO REMOVE THE REGULATOR WHILE THE CYLINDER IS STILL ON AND UNDER PRESSURE IS IN DIRECT CONFLICT WITH THE GUIDANCE AND WARNINGS IN THE ACCUPULSE USER MANUAL, WHICH STATES 'I WARNING: FAILURE TO CLOSE THE CYLINDER VALVE BEFORE REMOVING THE REGULATOR FROM THE CYLINDER WILL RESULT IN
Patient problems: Abrasion; Injury
MAUDE report 6600819 ↗
Other Conserver, Oxygen devices
- AirLife™ SFT2601Vyaire Medical, Inc. · Conserver, oxygen4 reports
- Drive OM-900AInovo, Inc. · Conserver, Oxygen2 reports
- Evolution™ OM-900Inovo, Inc. · Conserver, Oxygen2 reports
- AirLife™ FM2699Vyaire Medical, Inc. · Conserver, oxygen2 reports
- AirLife™ SFT2699Vyaire Medical, Inc. · Conserver, oxygen1 report
- EVOLUTION OM-900AInovo, Inc. · Conserver, Oxygen1 report
- AirLife™ SFT2692Vyaire Medical, Inc. · Conserver, oxygen1 report
- Drive P-224Drive Devilbiss Healthcare · Conserver, oxygen1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page