Neuron PND6F1056MRecalls & reported problems
Penumbra, Inc. · Neur6F DC MP, 105x6 · Class 2
All models of Neuron (36)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00814548010113), received between May 19, 2016 and Feb 22, 2018 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Device Damaged Prior to Use” (4 reports).
1 other recall cover devices cleared under the same FDA submission (K082290); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Damaged Prior to Use4
- Out-Of-Box Failure4
- Kinked3
- Bent1
- Break1
- Hole In Material1
Patient problems
- No Patient Involvement4
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 22, 2018Kinked; Device Damaged Prior to Use; Out-Of-Box Failure; Kinked; Device Damaged Prior to Use; Out-Of-Box Failure
DURING PREPARATION FOR A MEDICAL PROCEDURE, THE HOSPITAL STAFF FOUND THAT THE NEURON 6F 053 DELIVERY CATHETER (NEURON 053) WAS KINKED UPON REMOVAL FROM THE PACKAGING. THE DAMAGE TO THE NEURON 053 WAS FOUND PRIOR TO USE AND THEREFORE IT WAS NOT USED IN THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW NEURON 053.
Patient problems: No Patient Involvement
MAUDE report 7289236 ↗MalfunctionJan 5, 2018Kinked; Device Damaged Prior to Use; Out-Of-Box Failure; Kinked; Device Damaged Prior to Use; Out-Of-Box Failure
DURING PREPARATION FOR MEDICAL PROCEDURE, THE HOSPITAL STAFF FOUND THAT THE TIP OF A NEURON 6F 053 DELIVERY CATHETER (NEURON 053) WAS KINKED UPON REMOVAL FROM THE PACKAGING. THE DAMAGED NEURON 053 WAS FOUND PRIOR TO USE AND THEREFORE, WAS NOT USED IN THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW NEURON 053.
Patient problems: No Patient Involvement
MAUDE report 7166889 ↗MalfunctionOct 19, 2017Break; Hole In Material; Break; Hole In Material
THE PATIENT WAS UNDERGOING A MEDICAL PROCEDURE USING A NEURON 6F 053 DELIVERY CATHETER (NEURON 053). DURING THE PROCEDURE, THE PHYSICIAN ADVANCED THE NEURON 053 THROUGH A NON-PENUMBRA SHEATH AND INTO THE TARGET LOCATION. THE PHYSICIAN THEN NOTICED A HOLE OR BREAK IN THE NEURON 053 WHEN CONTRAST WAS INJECTED THROUGH IT. THE PHYSICIAN WAS UNSURE HOW THE BREAK OCCURRED BUT BELIEVED THAT IT HAPPENED AT A KINK IN THE PROXIMAL INTERNAL CAROTID ARTERY (ICA). AFTER THAT, THE PHYSICIAN NOTED A ¿BREAK DOWN¿ AT THE TIP OF A NON-PENUMBRA GUIDEWIRE, WHICH WAS RETRIEVED USING A SNARE DEVICE. IT WAS REPORTED THAT THE BREAKAGE OF THE GUIDEWIRE HAD NOTHING TO DO WITH ANY PENUMBRA DEVICES. THE PROCEDURE WAS THEN COMPLETED USING THE SAME NEURON 053 DESPITE THE FACT THAT IT WAS LEAKING. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6963211 ↗MalfunctionApr 20, 2017Kinked; Device Damaged Prior to Use; Out-Of-Box Failure; Kinked; Device Damaged Prior to Use; Out-Of-Box Failure
DURING PREPARATION FOR A THROMBECTOMY PROCEDURE, THE HOSPITAL STAFF NOTICED THAT THE DISTAL END OF A NEURON 6F 053 DELIVERY CATHETER (NEURON 053) WAS KINKED UPON REMOVAL FROM THE PACKAGING. THE DAMAGE TO THE NEURON 053 WAS FOUND PRIOR TO USE AND THEREFORE, IT WAS NOT USED IN THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW NEURON 053.
Patient problems: No Patient Involvement
MAUDE report 6509133 ↗MalfunctionMay 19, 2016Bent; Device Damaged Prior to Use; Out-Of-Box Failure; Bent; Device Damaged Prior to Use; Out-Of-Box Failure
DURING PREPARATION FOR A MEDICAL PROCEDURE, THE HOSPITAL STAFF NOTICED THAT THE TIP OF A NEURON 6F 053 DELIVERY CATHETER (NEURON 053) WAS BENT UPON REMOVAL FROM THE PACKAGING. THE BENT TIP OF THE NEURON 053 WAS FOUND PRIOR TO USE AND THEREFORE, THE NEURON 053 WAS NOT USED FOR THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW NEURON 053.
Patient problems: No Patient Involvement
MAUDE report 5667467 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K082290. One submission often covers many models, so check the product and lot details in each record.
- Z-0001-2010 · Sep 22, 2009 · TerminatedPenumbra Neuron Delivery Catheter 070, percutaneous catheter, catalog numbers PND6F0701058, PND6F070956, PND6F0701058M, and PND6F070956M; manufactured by Penumbra, Inc., Alameda,
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page