Penumbra System PST2Recalls & reported problems

Penumbra, Inc. · PS MAX Aspiration Tubing · Class 2

All models of Penumbra System (59)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00814548012025), received between Aug 28, 2015 and Nov 3, 2021 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Device Handling Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Device Handling Problem2
  • Out-Of-Box Failure2
  • Aspiration Issue1
  • Connection Problem1
  • Device Damaged Prior to Use1
  • Device Operational Issue1
  • Fracture1

Patient problems

  • No Patient Involvement4
  • No Consequences Or Impact To Patient2
  • No Known Impact Or Consequence To Patient2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 3, 2021Fracture; Suction Failure; Fracture; Suction Failure

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE USING PENUMBRA SYSTEM MAX ASPIRATION TUBING (TUBING). DURING THE PROCEDURE, THE TUBING DID NOT ASPIRATE AT ALL AND WAS FOUND TO BE CRACKED. THEREFORE, THE TUBING WAS REMOVED. THE PROCEDURE WAS COMPLETED USING NEW TUBING. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12744690 ↗
  • MalfunctionSep 7, 2016Misconnection; Device Handling Problem; Misconnection; Device Handling Problem

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE FOR A CEREBRAL INFARCTION IN THE RIGHT MIDDLE CEREBRAL ARTERY (MCA) USING A PENUMBRA SYSTEM ASPIRATION PUMP MAX 110V (PUMP MAX) AND THE PENUMBRA SYSTEM MAX ASPIRATION TUBING (TUBING). DURING THE PROCEDURE, ALTHOUGH THE PATIENT HAD SLIGHTLY TORTUOUS VESSELS, THE PHYSICIAN WAS ABLE TO APPROACH THE RIGHT MCA WITH A PENUMBRA SYSTEM ACE REPERFUSION CATHETER (ACE). THE PHYSICIAN'S ASSISTANT PREPARED FOR THE PROCEDURE BY CONNECTING THE CLEAN SIDE OF THE TUBING TO THE HUB OF THE ACE AND SETTING UP THE PUMP MAX AND CANISTER FOR ASPIRATION. THE PUMP MAX WAS THEN TURNED ON AND ASPIRATION WAS INITIATED. HOWEVER, AFTER WAITING ABOUT NINETY SECONDS, THE PHYSICIAN NOTICED A GRADUAL BACK FLOW OF BLOOD INTO THE PUMP MAX AND THEN NOTICED THAT THE ASSISTANT HAD INAPPROPRIATELY CONNECTED THE BACKSIDE END OF THE TUBING BY MISTAKE. THEREFORE, THE PUMP MAX WAS IMMEDIATELY TURNED OFF AND THE DEVICES WERE RECONNECTED APPROPRIATELY. THE PHYSICIAN THEN RE-INITIATED ASPIRATION AND NOTICED THAT THE ASPIRATION PRESSURE WAS NORMAL. ADDITIONALLY, THERE WAS NO LONGER A BACK FLOW OF BLOOD INTO THE PUMP MAX. THEREAFTER, THE PROCEDURE WAS COMPLETED USING THE SAME PE

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5934636 ↗
  • MalfunctionAug 2, 2016Device Handling Problem; Device Handling Problem

    DURING PREPARATION FOR A THROMBECTOMY PROCEDURE, THE HOSPITAL TECHNICIAN INADVERTENTLY DROPPED AN ASPIRATION TUBING ONTO THE FLOOR UPON REMOVING IT FROM THE PACKAGING. THE ASPIRATION TUBING DROPPED PRIOR TO USE AND THEREFORE, WAS NOT USED FOR THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW ASPIRATION TUBING.

    Patient problems: No Patient Involvement

    MAUDE report 5841017 ↗
  • MalfunctionApr 29, 2016Hole In Material; Device Damaged Prior to Use; Out-Of-Box Failure; Hole In Material; Device Damaged Prior to Use; Out-Of-Box Failure

    DURING PREPARATION FOR A THROMBECTOMY PROCEDURE, THE PHYSICIAN NOTICED THAT THE PENUMBRA SYSTEM MAX ASPIRATION TUBING (TUBING) HAD A HOLE AND THEREFORE, SUCTION WAS NOT POSSIBLE. THE DAMAGED TUBING WAS FOUND PRIOR TO USE AND WAS NOT USED FOR THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW ASPIRATION TUBING.

    Patient problems: No Patient Involvement

    MAUDE report 5621688 ↗
  • MalfunctionFeb 8, 2016Occlusion Within Device; Out-Of-Box Failure; Aspiration Issue; Device Operational Issue; Occlusion Within Device; Out-Of-Box Failure; Aspiration Issue; Device Operational Issue

    DURING PREPARATION FOR A THROMBECTOMY PROCEDURE, THE HOSPITAL STAFF NOTICED THAT THE PENUMBRA SYSTEM MAX ASPIRATION TUBING (TUBING) WAS OCCLUDED AND DID NOT ASPIRATE. THE DEFECTIVE TUBING WAS FOUND PRIOR TO USE AND WAS NOT USED FOR THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW TUBING.

    Patient problems: No Patient Involvement

    MAUDE report 5417563 ↗
  • MalfunctionDec 15, 2015Adverse Event Without Identified Device or Use Problem

    DURING PREPARATION FOR A THROMBECTOMY PROCEDURE, THE TECHNOLOGIST HAD DIFFICULTY LOCATING THE PENUMBRA ASPIRATION TUBING (TUBING) BECAUSE THE PACKAGING OF THE TUBING LOOKED SIMILAR TO THE PACKAGING OF OTHER PENUMBRA DEVICES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED BY MANUAL ASPIRATION USING A PENUMBRA SYSTEM ACE 64 REPERFUSION CATHETER, A PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER AND A SYRINGE.

    Patient problems: No Patient Involvement

    MAUDE report 5298134 ↗
  • MalfunctionOct 29, 2015Use of Device Problem; Connection Problem

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE FOR ACUTE-STAGE CEREBRAL INFARCTION USING A PENUMBRA SYSTEM ASPIRATION PUMP MAX 110 (PUMP MAX) AND A PENUMBRA SYSTEM MAX ASPIRATION TUBING (TUBING). DURING PREPARATION FOR THE PROCEDURE, THE NURSE MISTAKENLY CONNECTED THE BLUE END TERMINAL OF THE TUBING DIRECTLY TO THE SUCTION PORT OF THE PUMP MAX. DURING THE PROCEDURE, THE PHYSICIAN BEGAN ASPIRATING THE THROMBUS AND NOTICED THE PUMP MAX WAS NOT CONNECTED PROPERLY. THE PHYSICIAN CORRECTLY RECONNECTED THE TUBING AND THE PUMP MAX BUT THE NEGATIVE PRESSURE DID NOT GO DOWN. THE PROCEDURE CONTINUED USING A SYRINGE.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 5187565 ↗
  • MalfunctionAug 28, 2015Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE USING A PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER (3MAX). UPON REMOVAL FROM THE PACKAGING, A PENUMBRA SYSTEM MAX ASPIRATION TUBING WAS FOUND INSIDE THE BOX LABELLED FOR A PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER. THE PROCEDURE CONTINUED USING A NEW 3MAX CATHETER AND THE ASPIRATION TUBING WAS SAVED FOR USE AT A LATER DATE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5040340 ↗

FDA clearance and classification

Cleared through Special 510(k) K113163 on Nov 23, 2011 (decision: Substantially Equivalent). Submitted by Penumbra, Inc. as “Penumbra System Max”. The FDA review took 28 days. FDA record ↗

The FDA regulates product code NRY (catheter, thrombus retriever) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗

FDA definition of this device type: The device is intended to restore blood flow by removing thrombus/clots in patients experiencing ischemic stroke. The prior clearances in this category have been limited to simply identifying catheter placed in the peripheral, coronary, and neurovasculature. It is felt that this is a unique claim and should not be combined with previously cleared catheters under a general procode.

Common questions about the Penumbra System PST2

Has the Penumbra System PST2 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00814548012025), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Penumbra System PST2?

The FDA's MAUDE database lists 8 adverse event reports for this device, received between Aug 28, 2015 and Nov 3, 2021. In those reports, the most common specific device problems are “Device Handling Problem” (2), “Out-Of-Box Failure” (2) and “Aspiration Issue” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Penumbra System PST2?

None of the reports are classified as deaths, 0 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Penumbra System PST2 FDA approved?

It was cleared through Special 510(k) K113163 on Nov 23, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Penumbra System PST2, and is it still sold?

The FDA's device database lists Penumbra, Inc. as the company responsible for it (brand Penumbra System, model PST2). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.