Skip to content

Penumbra System PSC054L125-BRecalls & reported problems

Penumbra, Inc. · PS MAX RC5MAX 125cm · Class 2

All models of Penumbra System (59)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00814548012759), received between Mar 14, 2017 and Apr 10, 2020 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Device Damaged Prior to Use” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Device Damaged Prior to Use1
  • Fracture1
  • Kinked1
  • Material Deformation1
  • Out-Of-Box Failure1
  • Physical Resistance/Sticking1

Patient problems

  • Hemorrhage, Subarachnoid1
  • Neurological Deficit/Dysfunction1
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1
  • Perforation of Vessels1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 10, 2020Material Deformation; Physical Resistance/Sticking; Material Deformation; Physical Resistance/Sticking

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE RIGHT VERTEBRAL ARTERY USING A PENUMBRA SYSTEM 5MAX REPERFUSION CATHETER (5MAXC), A NEURON MAX 6F 088 LONG SHEATH (NEURON MAX) AND A NON-PENUMBRA BALLOON CATHETER. IT WAS NOTED THAT THE PATIENT¿S VERTEBRAL VESSELS WERE ELONGATED. DURING THE PROCEDURE, THE PHYSICIAN ADVANCED A 5MAXC THROUGH THE NEURON MAX. THEN, WHILE ADVANCING THE BALLOON CATHETER APPROXIMATELY 5 CENTIMETERS FROM THE DISTAL END OF THE 5MAXC, THE PHYSICIAN EXPERIENCED RESISTANCE; THEREFORE, THE BALLOON CATHETER AND 5MAXC WERE REMOVED. UPON REMOVAL, IT WAS NOTICED THAT THE BALLOON CATHETER AND 5MAXC WERE DAMAGED. IT WAS REPORTED THAT 2 TO 5 CENTIMETERS FROM THE DISTAL END OF THE 5MAXC WAS DAMAGED. THE PROCEDURE WAS COMPLETED USING A NEW 5MAXC AND NEW BALLOON CATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9948530 ↗
  • MalfunctionMar 29, 2017Kinked; Device Damaged Prior to Use; Out-Of-Box Failure; Kinked; Device Damaged Prior to Use; Out-Of-Box Failure

    DURING PREPARATION FOR A THROMBECTOMY PROCEDURE, THE HOSPITAL STAFF NOTICED THAT THE PENUMBRA SYSTEM 5MAX REPERFUSION CATHETER (5MAX) WAS KINKED AND COMING APART UPON REMOVAL FROM THE PACKAGING HOOP. THE DAMAGED 5MAX WAS NOTICED PRIOR TO USE AND THEREFORE, WAS NOT USED FOR THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW 5MAX.

    Patient problems: No Patient Involvement

    MAUDE report 6443042 ↗
  • InjuryMar 14, 2017Fracture; Fracture

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE M2 SECTION OF THE MIDDLE CEREBRAL ARTERY (MCA) USING A PENUMBRA SYSTEM 5MAX REPERFUSION CATHETER (5MAX) AND A PENUMBRA SYSTEM ASPIRATION PUMP MAX 220V (PUMP MAX). DURING THE PROCEDURE, THE PHYSICIAN ATTEMPTED TO REMOVE THROMBUS FOR THE FIRST AND SECOND PASS USING A STENT RETRIEVER DEVICE AND A 5MAX; HOWEVER, THE THROMBUS WAS PULLED PROXIMAL AND NO REPERFUSION WAS ACHIEVED. THE STENT RETRIEVER WAS THEN REMOVED, ANOTHER STENT DEVICE WAS ADVANCED WITH THE 5MAX FOR THE THIRD PASS; HOWEVER, NO REPERFUSION WAS ACHIEVED. THEREFORE, THE STENT RETRIEVER AND 5MAX WERE REMOVED. UPON REMOVAL, THE PHYSICIAN NOTICED THAT THE 5MAX WAS BROKEN IN TWO PIECES. THE 5MAX WAS THEN DISCONNECTED FROM THE PUMP MAX AND A NEURON MAX 6F 088 (NEURON MAX) WAS CONNECTED TO THE PUMP MAX; HOWEVER, ONLY SLOW REVERSAL FLOW OCCURRED. UPON CHECKING THE PUMP MAX, THE PHYSICIAN NOTICED THAT NO VACUUM WAS DETECTED. IT WAS ALSO REPORTED THAT THE PUMP MAX INTERMITTENTLY DID NOT PRODUCE ENOUGH VACUUM DURING THE FIRST, SECOND AND THIRD PASS. THE PHYSICIAN THEN PERFORMED MANUAL ASPIRATION WITH A 20ML SYRINGE AND AN ANGIOGRAPHY. ACCORDING TO THE PHYSICIAN, IN CONSEQUEN

    Patient problems: Hemorrhage, Subarachnoid; Neurological Deficit/Dysfunction; Perforation of Vessels

    MAUDE report 6404654 ↗

FDA clearance and classification

Cleared through 510(k) K090752 on Sep 21, 2009 (decision: Substantially Equivalent). Submitted by Penumbra, Inc. as “Penumbra System (Pennumbra Reperfusion Catheter 054), Model Psc054; (Penumbra Separator 054), Model Pss054”. The FDA review took 185 days. FDA record ↗

The FDA regulates product code NRY (catheter, thrombus retriever) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗

FDA definition of this device type: The device is intended to restore blood flow by removing thrombus/clots in patients experiencing ischemic stroke. The prior clearances in this category have been limited to simply identifying catheter placed in the peripheral, coronary, and neurovasculature. It is felt that this is a unique claim and should not be combined with previously cleared catheters under a general procode.

Common questions about the Penumbra System PSC054L125-B

Has the Penumbra System PSC054L125-B been recalled?

No FDA recall record in our data names this exact device identifier (DI 00814548012759), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Penumbra System PSC054L125-B?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Mar 14, 2017 and Apr 10, 2020. In those reports, the most common specific device problems are “Device Damaged Prior to Use” (1), “Fracture” (1) and “Kinked” (1); the most common patient problems are “Hemorrhage, Subarachnoid” (1), “Neurological Deficit/Dysfunction” (1) and “Perforation of Vessels” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Penumbra System PSC054L125-B?

None of the reports are classified as deaths, 1 as an injury and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Penumbra System PSC054L125-B FDA approved?

It was cleared through 510(k) K090752 on Sep 21, 2009. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Penumbra System PSC054L125-B, and is it still sold?

The FDA's device database lists Penumbra, Inc. as the company responsible for it (brand Penumbra System, model PSC054L125-B). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETER, THROMBUS RETRIEVER devices

All CATHETER, THROMBUS RETRIEVER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page