Benchmark BMK6F115MRecalls & reported problems
Penumbra, Inc. · Bmk70 MP, 115cm · Class 2
All models of Benchmark (17)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00814548014104), received between Jan 4, 2018 and Nov 10, 2023 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation4
- Device Damaged Prior to Use2
- Failure to Advance2
- Break1
- Leak/Splash1
- Physical Resistance/Sticking1
Patient problems
- No Consequences Or Impact To Patient3
- No Clinical Signs, Symptoms or Conditions2
- No Patient Involvement2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 10, 2023Break; Break
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE MIDDLE MENINGEAL ARTERY (MMA) USING A BENCHMARK 6F 071 DELIVERY CATHETER (BENCHMARK), A NON-PENUMBRA INTRODUCER SHEATH, AND GUIDEWIRES. DURING THE PROCEDURE, WHILE ATTEMPTING TO ADVANCE THE BENCHMARK THROUGH THE INTRODUCER SHEATH, THE PHYSICIAN FRACTURED THE MID-SHAFT OF THE BENCHMARK. IT WAS REPORTED THAT THE FRACTURED PART OF THE BENCHMARK WAS OUTSIDE THE PATIENT, AND CONTRAST MEDIA WAS LEAKING FROM THE BENCHMARK. THEREFORE, THE BENCHMARK WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW BENCHMARK AND THE SAME INTRODUCER SHEATH. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18115042 ↗MalfunctionAug 4, 2021Failure to Advance; Material Deformation; Failure to Advance; Material Deformation
THE PATIENT WAS UNDERGOING A MEDICAL PROCEDURE IN THE OPHTHALMIC SEGMENT OF THE RIGHT INTERNAL CAROTID ARTERY USING A BENCHMARK 6F 071 DELIVERY CATHETER (BENCHMARK) AND A NON-PENUMBRA LONG SHEATH. DURING THE PROCEDURE, THE PHYSICIAN EXPERIENCED RESISTANCE WHILE INSERTING THE BENCHMARK INTO THE LONG SHEATH AND SUBSEQUENTLY, THE DISTAL TIP OF THE BENCHMARK BECAME DAMAGED. THEREFORE, THE BENCHMARK AND CATHETER WERE REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW BENCHMARK AND ANOTHER SHEATH. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12277948 ↗MalfunctionJun 19, 2020Material Deformation; Material Deformation
THE PATIENT WAS UNDERGOING A MEDICAL PROCEDURE USING A BENCHMARK 6F 071 DELIVERY CATHETER (BENCHMARK). DURING THE PROCEDURE, WHILE ADVANCING THE BENCHMARK INTO THE PATIENT, THE BENCHMARK KINKED; THEREFORE, IT WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW BENCHMARK. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10175449 ↗MalfunctionApr 12, 2019Material Deformation; Material Deformation
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE ANTERIOR COMMUNICATING ARTERY (ACOM) USING BENCHMARK DELIVERY CATHETERS (BENCHMARKS). DURING THE PROCEDURE, THE DISTAL END OF A BENCHMARK BECAME KINKED WHILE BEING ADVANCED INTO A NEURON MAX 6F 088 LONG SHEATH (NEURON MAX) AND, THEREFORE, THE BENCHMARK WAS REMOVED. THE PHYSICIAN THEN ATTEMPTED TO ADVANCE A NEW BENCHMARK INTO THE NEURON MAX, HOWEVER THE DISTAL END OF THE BENCHMARK FRACTURED WHILE BEING ADVANCED INTO THE NEURON MAX. THE PHYSICIAN THEN SUCCESSFULLY COMPLETED THE PROCEDURE USING A THIRD BENCHMARK WITH THE SAME NEURON MAX. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8510261 ↗MalfunctionMar 8, 2019Device Damaged Prior to Use; Material Deformation; Device Damaged Prior to Use; Material Deformation
DURING PREPARATION FOR A MEDICAL PROCEDURE, THE PHARMACIST NOTICED THAT THE DISTAL TIP OF A BENCHMARK 6F 071 DELIVERY CATHETER (BENCHMARK) WAS FLATTENED UPON OPENING OF THE PACKAGING. THE DAMAGE TO THE BENCHMARK WAS FOUND PRIOR TO USE AND, THEREFORE, IT WAS NOT USED IN THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING ANOTHER BENCHMARK.
Patient problems: No Patient Involvement
MAUDE report 8404445 ↗MalfunctionOct 9, 2018Failure to Advance; Physical Resistance/Sticking; Failure to Advance; Physical Resistance/Sticking
THE PATIENT WAS UNDERGOING A MEDICAL PROCEDURE USING A BENCHMARK 6F 071 DELIVERY CATHETER (BENCHMARK). DURING THE PROCEDURE, WHILE ATTEMPTING TO INSERT A BENCHMARK INTO A NEURON MAX 6F 088 LONG SHEATH (NEURON MAX), THE PHYSICIAN EXPERIENCED RESISTANCE AND JUST THE TIP OF THE BENCHMARK WOULD ONLY ADVANCE INTO THE NEURON MAX; THEREFORE, THE BENCHMARK WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER BENCHMARK AND THE SAME NEURON MAX. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7949933 ↗MalfunctionJan 4, 2018Leak/Splash; Device Damaged Prior to Use; Leak/Splash; Device Damaged Prior to Use
WHILE PREPARING A BENCHMARK 6F 071 DELIVERY CATHETER (BENCHMARK) FOR THROMBECTOMY PROCEDURE, THE PHYSICIAN FLUSHED THE BENCHMARK AND NOTICED A LEAK AT THE BASE OF THE PROXIMAL END. THEREFORE, THE BENCHMARK WAS SET ASIDE AND THE PROCEDURE WAS COMPLETED USING ANOTHER BENCHMARK.
Patient problems: No Patient Involvement
MAUDE report 7164167 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.