Penumbra System 3MAXCKITRecalls & reported problems
Penumbra, Inc. · 3MAXCKIT · Class 2
All models of Penumbra System (59)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00814548016757), received between Jan 11, 2017 and Feb 27, 2023 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Physical Resistance/Sticking” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Physical Resistance/Sticking4
- Material Deformation3
- Break2
- Device Damaged Prior to Use2
- Fracture2
- Out-Of-Box Failure2
- Failure to Advance1
- Kinked1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Consequences Or Impact To Patient3
- No Patient Involvement2
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 27, 2023Fracture; Physical Resistance/Sticking; Fracture; Physical Resistance/Sticking
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE M1 SEGMENT OF THE MIDDLE CEREBRAL ARTERY (MCA) USING A PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER (3MAXC), A PENUMBRA ENGINE (ENGINE), A NON-PENUMBRA CATHETER, A NON-PENUMBRA SHEATH, AND A GUIDEWIRE. IT SHOULD BE NOTED THAT THE PATIENT''S ANATOMY WAS TORTUOUS AND HAD ARTERIOSCLEROSIS. DURING THE PROCEDURE, THE PHYSICIAN COMPLETED A DIRECT ASPIRATION FIRST PASS (ADAPT) USING THE 3MAXC AND PARTIAL RECANALIZATION OF THE TARGET VESSEL WAS ACHIEVED. WHEN THE 3MAXC PASSED THROUGH THE LESION A SECOND TIME, RESISTANCE WAS ENCOUNTERED. IT WAS ALSO REPORTED THAT THERE WAS NO REVERSE FLOW AFTER ASPIRATING THE THROMBUS. WHILE RETRACTING THE 3MAXC AND CATHETER INTO THE SHEATH UNDER NEGATIVE PRESSURE, RESISTANCE WAS ENCOUNTERED. UPON REMOVAL, THE 3MAXC WAS FOUND FRACTURED AND THE DISTAL END OF THE 3MAXC REMAINED IN THE M1 TO THE INTERNAL CAROTID ARTERY (ICA). NEXT, A SNARE DEVICE WAS USED TO RETRIEVE THE FRACTURED PIECE OF THE 3MAXC. ON THE BACK TABLE, THE PHYSICIAN NOTED THAT THE DISTAL END OF THE 3MAXC WAS KINKED NEAR THE FRACTURE AND WAS LIKELY CAUSED BY THE SNARE DEVICE. AT THIS POINT, THE PROCEDURE WAS ENDED. IT WAS ALSO NOTED THAT T
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16444262 ↗MalfunctionJan 17, 2023Break; Material Deformation; Physical Resistance/Sticking; Break; Material Deformation; Physical Resistance/Sticking
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE M2 SEGMENT OF THE MIDDLE CEREBRAL ARTERY (MCA) USING A PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER (3MAXC), A PENUMBRA SYSTEM ACE 68 REPERFUSION CATHETER (ACE68), AND A GUIDEWIRE. DURING THE PROCEDURE, THE 3MAXC WAS ADVANCED TO THE TARGET LOCATION AND USED TO PERFORM ASPIRATION. WHILE WITHDRAWING THE 3MAXC FROM THE ACE68, RESISTANCE WAS ENCOUNTERED AND THE 3MAXC BECAME STUCK WHEN THE DISTAL END OF THE 3MAXC WAS APPROXIMATELY ONE MILLIMETER OUTSIDE OF THE ACE68. THE PHYSICIAN CONTINUED TO RETRACT THE 3MAXC OVER THE GUIDEWIRE AND SUBSEQUENTLY, THE 3MAXC BROKE IN TWO PIECES AT THE PROXIMAL END. THEREFORE, THE ACE68 CONTAINING THE BROKEN 3MAXC WAS REMOVED FROM THE PATIENT. IT WAS ALSO REPORTED THAT THE 3MAXC WAS CRUSHED AND KINKED NEAR THE LOCATION OF THE BREAK AT THE PROXIMAL END. AT THIS POINT, THE PROCEDURE WAS SUCCESSFULLY COMPLETED AND NO ADDITIONAL PASSES WERE REQUIRED. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16188067 ↗MalfunctionMay 3, 2021Physical Resistance/Sticking; Physical Resistance/Sticking
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE POSTERIOR CEREBRAL ARTERY USING A PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER (3MAXC), NON-PENUMBRA GUIDE CATHETER, AND A GUIDEWIRE. DURING THE PROCEDURE, THE PHYSICIAN INSERTED THE 3MAXC INTO THE GUIDE CATHETER AND THEN INSERTED THE GUIDEWIRE INTO THE 3MAXC; HOWEVER, THE GUIDEWIRE BECAME STUCK IN THE MIDDLE SHAFT OF THE 3MAXC AND GUIDE CATHETER. THE 3MAXC WAS THEREFORE, REMOVED OUT FROM THE GUIDE CATHETER. SUBSEQUENTLY, THE PHYSICIAN FLUSHED THE 3MAXC ON THE BACK TABLE AND THEN INSERTED THE SAME GUIDEWIRE INTO THE 3MAXC, BUT AGAIN, IT BECAME STUCK INSIDE THE 3MAXC. THEREFORE, THE 3MAXC WAS NOT USED FOR THE REMAINDER OF THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW 3MAXC WITH THE SAME GUIDE CATHETER AND GUIDEWIRE. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11760476 ↗MalfunctionApr 21, 2021Material Deformation; Material Deformation
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE POSTERIOR CEREBRAL ARTERY USING A PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER (3MAXC), NON-PENUMBRA GUIDE CATHETER, GUIDE, AND GUIDEWIRE. IT WAS REPORTED THAT THE 3MAXC AND GUIDE CATHETER WERE FLUSHED WITH SALINE PRIOR TO USE. DURING THE PROCEDURE, THE PHYSICIAN EXPERIENCED RESISTANCE WHEN ADVANCING THE 3MAXC THROUGH THE MIDDLE OF THE GUIDE CATHETER. THE 3MAXC WAS REMOVED AND UPON FLUSHING ON THE BACK TABLE USING A SYRINGE, THE DISTAL TIP OF THE 3MAXC WAS OVALIZED. THEREFORE, THE 3MAXC WAS NOT USED FOR THE REMAINDER OF THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW 3MAXC AND THE SAME GUIDE CATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11704127 ↗MalfunctionMar 8, 2021Stretched; Stretched
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE MIDDLE CEREBRAL ARTERY (MCA) USING A PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER (3MAXC), A PENUMBRA SYSTEM ACE 68 REPERFUSION CATHETER (ACE68), A NON-PENUMBRA GUIDE CATHETER, AND A GUIDEWIRE. DURING THE PROCEDURE, THE PHYSICIAN INJECTED TISSUE PLASMINOGEN ACTIVATOR (TPA) AND SUBSEQUENTLY PLACED THE GUIDE CATHETER IN THE INTERNAL CAROTID ARTERY (ICA) AND THE ACE68 WAS ADVANCED INTO THE CAROTID SIPHON. NEXT, THE 3MAXC WAS PASSED THROUGH THE CLOT TOWARDS THE TARGET LOCATION. THE PHYSICIAN REMOVED THE 3MAXC AND GUIDEWIRE FROM THE ACE68 IN ORDER TO BEGIN ASPIRATION AND NOTICED THAT THE 3MAXC WAS STRETCHED. THEREFORE, THE 3MAXC WAS NO LONGER USED IN THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING TWO OTHER STENT RETRIEVERS, THE SAME GUIDE CATHETER, AND THE SAME ACE68. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11441264 ↗MalfunctionDec 8, 2020Break; Break
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE POSTERIOR CEREBRAL ARTERY (PCA) USING A PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER (3MAXC), PENUMBRA SYSTEM ACE 68 REPERFUSION CATHETER (ACE68), PENUMBRA ENGINE (ENGINE), AND NON-PENUMBRA CATHETER. DURING THE PROCEDURE, THE PHYSICIAN ADVANCED THE DISTAL TIP OF THE 3MAXC INTO THE TARGET VESSEL AND STARTED ASPIRATION WITH THE ENGINE USING ADAPT TECHNIQUE. SUBSEQUENTLY, 90 SECONDS AFTER THE BLOOD INFLOW STOPPED, THE PHYSICIAN ATTEMPTED TO REMOVE THE 3MAXC. IT APPEARED THAT THE MARKER BAND OF THE 3MAXC BROKE WHILE RETRACTING THE 3MAXC. THEREFORE, THE PHYSICIAN TURNED OFF THE ENGINE AND REMOVED THE 3MAXC. UPON REMOVAL, THE 3MAXC WAS OBSERVED TO BE BROKEN. IT WAS REPORTED THAT THERE WAS A BREAK 55MM FROM THE DISTAL TIP OF THE 3MAXC ON THE BACK TABLE. HOWEVER, THE PHYSICIAN CONFIRMED THAT THE MARKER BAND TIP WAS NOT INSIDE THE PATIENT. THE PHYSICIAN CONTINUED THE PROCEDURE USING AN ACE68 WITH THE SAME CATHETER, ENGINE, AND CANISTER BUT WAS UNABLE TO ACHIEVE RECANALIZATION. THE PROCEDURE ENDED AT THIS POINT. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10968794 ↗MalfunctionSep 4, 2020Failure to Advance; Physical Resistance/Sticking; Failure to Advance; Physical Resistance/Sticking
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE M2 SEGMENT OF THE MIDDLE CEREBRAL ARTERY (MCA) USING A PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER (3MAXC), A PENUMBRA SYSTEM ACE 60 REPERFUSION CATHETER (ACE60), A NON-PENUMBRA GUIDE, AND A GUIDEWIRE. DURING THE PROCEDURE, WHILE INTRODUCING THE DEVICES TO THE PATIENT, THE PHYSICIAN ENCOUNTERED RESISTANCE AND THE GUIDEWIRE BECAME STUCK AFTER BEING ADVANCED APPROXIMATELY NINETY CENTIMETERS THROUGH THE 3MAXC. IT WAS REPORTED THAT THE SAME ISSUE HAD PREVIOUSLY OCCURRED ON THE BACK TABLE, PRIOR TO INTRODUCTION TO THE PATIENT. THEREFORE, THE 3MAXC WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED USING A NON-PENUMBRA MICROCATHETER, THE SAME ACE60, AND A STENT RETRIEVER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10497098 ↗MalfunctionJul 24, 2020Material Deformation; Material Deformation
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE M2 SEGMENT OF THE MIDDLE CEREBRAL ARTERY (MCA) USING A PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER (3MAXC). DURING THE PROCEDURE, WHILE FLUSHING THE 3MAXC ON THE BACK TABLE, THE 3MAXC WAS OBSERVED TO BE DAMAGED. THE PROCEDURE WAS COMPLETED USING A PENUMBRA SYSTEM 4MAX REPERFUSION CATHETER (4MAXC) AND NON-PENUMBRA STENT RETRIEVER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10323836 ↗MalfunctionFeb 21, 2018Kinked; Device Damaged Prior to Use; Out-Of-Box Failure; Kinked; Device Damaged Prior to Use; Out-Of-Box Failure
DURING PREPARATION FOR MEDICAL PROCEDURE, THE HOSPITAL TECHNICIAN FOUND THAT THE PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER (3MAXC) WAS KINKED IN TWO PLACES UPON REMOVAL FROM THE PACKAGING OF THE PENUMBRA SYSTEM 3MAX HI-FLOW KIT. THE DAMAGE TO THE 3MAXC WAS FOUND PRIOR TO USE AND THEREFORE THE 3MAXC WAS NOT USED IN THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW 3MAXC.
Patient problems: No Patient Involvement
MAUDE report 7286225 ↗MalfunctionJan 11, 2017Fracture; Device Damaged Prior to Use; Out-Of-Box Failure; Fracture; Device Damaged Prior to Use; Out-Of-Box Failure
DURING PREPARATION FOR A THROMBECTOMY PROCEDURE, THE HOSPITAL STAFF NOTICED THAT THE PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER (3MAX) WAS FRACTURED ALONG ITS SIDE UPON REMOVAL FROM ITS PACKAGING. THE FRACTURED 3MAX WAS FOUND PRIOR TO USE AND THEREFORE, WAS NOT USED IN THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW 3MAX.
Patient problems: No Patient Involvement
MAUDE report 6244398 ↗
FDA clearance and classification
Cleared through Special 510(k) K113163 on Nov 23, 2011 (decision: Substantially Equivalent). Submitted by Penumbra, Inc. as “Penumbra System Max”. The FDA review took 28 days. FDA record ↗
The FDA regulates product code NRY (catheter, thrombus retriever) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗
FDA definition of this device type: The device is intended to restore blood flow by removing thrombus/clots in patients experiencing ischemic stroke. The prior clearances in this category have been limited to simply identifying catheter placed in the peripheral, coronary, and neurovasculature. It is felt that this is a unique claim and should not be combined with previously cleared catheters under a general procode.
Common questions about the Penumbra System 3MAXCKIT
Has the Penumbra System 3MAXCKIT been recalled?
No FDA recall record in our data names this exact device identifier (DI 00814548016757), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Penumbra System 3MAXCKIT?
The FDA's MAUDE database lists 10 adverse event reports for this device, received between Jan 11, 2017 and Feb 27, 2023. In those reports, the most common specific device problems are “Physical Resistance/Sticking” (4), “Material Deformation” (3) and “Break” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Penumbra System 3MAXCKIT?
None of the reports are classified as deaths, 0 as injuries and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Penumbra System 3MAXCKIT FDA approved?
It was cleared through Special 510(k) K113163 on Nov 23, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Penumbra System 3MAXCKIT, and is it still sold?
The FDA's device database lists Penumbra, Inc. as the company responsible for it (brand Penumbra System, model 3MAXCKIT). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, THROMBUS RETRIEVER devices
- Penumbra System 5MAXJET7KIT-BPenumbra, Inc. · Catheter, Thrombus Retriever1 recall14 deaths221 reports
- Penumbra System 3MAXCPenumbra, Inc. · CATHETER, THROMBUS RETRIEVER1 death241 reports
- EMBOTRAP ET309537Neuravi Limited · Catheter, Thrombus Retriever13 deaths177 reports
- ACE 5MAXACE068KITPenumbra, Inc. · CATHETER, THROMBUS RETRIEVER170 reports
- Embotrap ET009533Neuravi Limited · Catheter, Thrombus Retriever16 deaths153 reports
- Penumbra System 5MAXACE132Penumbra, Inc. · CATHETER, THROMBUS RETRIEVER3 deaths112 reports
- Penumbra System 5MAXJET7KIT-APenumbra, Inc. · Catheter, Thrombus Retriever1 death94 reports
- Trevo XP 90185Stryker Corporation · CATHETER, THROMBUS RETRIEVER1 recall1 death40 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page