Penumbra System PSC054KITRecalls & reported problems
Penumbra, Inc. · PSC054KIT · Class 2
All models of Penumbra System (59)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00814548016771), received between Jun 26, 2020 and Jun 26, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Physical Resistance/Sticking” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Physical Resistance/Sticking1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 26, 2020Physical Resistance/Sticking; Physical Resistance/Sticking
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE LEFT MIDDLE CEREBRAL ARTERY (MCA) USING A PENUMBRA SYSTEM 5MAX REPERFUSION CATHETER (5MAXC), A PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER (3MAXC) AND A NEURON MAX 6F 088 LONG SHEATH (NEURON MAX). IT WAS NOTED THAT THE PATIENT¿S ANATOMY WAS MILDLY TORTUOUS. DURING THE PROCEDURE, THE PHYSICIAN EXPERIENCED SLIGHT RESISTANCE WHILE ADVANCING THE 5MAXC THROUGH THE NEURON MAX AND WHILE ADVANCING THE 3MAXC THROUGH THE 5MAXC. AFTER ADVANCING THE 3MAXC INTO THE DISTAL BRANCH OF THE TARGET VESSEL, A PASS WAS COMPLETED USING A DIRECT ASPIRATION FIRST PASS TECHNIQUE (ADAPT). WITH THE CLOT ON THE TIP OF THE 3MAXC AND UNDER ASPIRATION, THE PHYSICIAN RETRACTED THE 3MAXC INTO THE 5MAXC WITH NO SIGNIFICANT RESISTANCE. HOWEVER, IT SEEMED LIKE A SECTION OF THE 3MAXC BROKE OFF. AFTER REMOVING THE 5MAXC, IT WAS NOTICED THAT APPROXIMATELY FIVE TO TEN CENTIMETERS OF THE DISTAL END OF THE 3MAXC WAS FOUND INSIDE THE 5MAXC, ALONG WITH THE THROMBUS. THE PROCEDURE WAS ENDED AT THIS POINT. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10201401 ↗
FDA clearance and classification
Cleared through 510(k) K090752 on Sep 21, 2009 (decision: Substantially Equivalent). Submitted by Penumbra, Inc. as “Penumbra System (Pennumbra Reperfusion Catheter 054), Model Psc054; (Penumbra Separator 054), Model Pss054”. The FDA review took 185 days. FDA record ↗
The FDA regulates product code NRY (catheter, thrombus retriever) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗
FDA definition of this device type: The device is intended to restore blood flow by removing thrombus/clots in patients experiencing ischemic stroke. The prior clearances in this category have been limited to simply identifying catheter placed in the peripheral, coronary, and neurovasculature. It is felt that this is a unique claim and should not be combined with previously cleared catheters under a general procode.
Common questions about the Penumbra System PSC054KIT
Has the Penumbra System PSC054KIT been recalled?
No FDA recall record in our data names this exact device identifier (DI 00814548016771), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Penumbra System PSC054KIT?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Jun 26, 2020. That report lists device problem “Physical Resistance/Sticking” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Penumbra System PSC054KIT?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Penumbra System PSC054KIT FDA approved?
It was cleared through 510(k) K090752 on Sep 21, 2009. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Penumbra System PSC054KIT, and is it still sold?
The FDA's device database lists Penumbra, Inc. as the company responsible for it (brand Penumbra System, model PSC054KIT). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, THROMBUS RETRIEVER devices
- Penumbra System 5MAXJET7KIT-BPenumbra, Inc. · Catheter, Thrombus Retriever1 recall14 deaths221 reports
- Penumbra System 3MAXCPenumbra, Inc. · CATHETER, THROMBUS RETRIEVER1 death241 reports
- EMBOTRAP ET309537Neuravi Limited · Catheter, Thrombus Retriever13 deaths177 reports
- ACE 5MAXACE068KITPenumbra, Inc. · CATHETER, THROMBUS RETRIEVER170 reports
- Embotrap ET009533Neuravi Limited · Catheter, Thrombus Retriever16 deaths153 reports
- Penumbra System 5MAXACE132Penumbra, Inc. · CATHETER, THROMBUS RETRIEVER3 deaths112 reports
- Penumbra System 5MAXJET7KIT-APenumbra, Inc. · Catheter, Thrombus Retriever1 death94 reports
- Trevo XP 90185Stryker Corporation · CATHETER, THROMBUS RETRIEVER1 recall1 death40 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.