UroLift UL400-4 (UDI 00814932020049)Recalls & reported problems
Neotract, Inc. · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00814932020049), received between Mar 5, 2020 and Sep 2, 2020 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K173087); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jun 24, 2022Class IIOpen, Classified
There is the potential that during implant deployment, the device may not properly deliver a implant.
Z-1502-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture3
Patient problems
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 2, 2020Fracture; Fracture
ON 22 JULY 2020, NEOTRACT WAS INFORMED OF A PROSTATIC URETHRAL LIFT (PUL) PROCEDURE DURING WHICH THE PHYSICIAN NOTED THAT THREE DEVICES APPEARED TO NOT DEPLOY PROPERLY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND NO PATIENT HARM OR INJURY WAS REPORTED. ON 3 AUGUST 2020, NEOTRACT'S INVESTIGATION OF THE RETURNED DEVICES INDICATED THAT ONE DEVICE WAS MISSING 1MM OF THE NEEDLE TIP. NEOTRACT NOTIFIED THE PHYSICIAN OF THE INVESTIGATION RESULTS. ON (B)(6) 2020, THE PHYSICIAN ACKNOWLEDGED THE MISSING NEEDLE PIECE AND STATED THAT HE DID NOT FEEL FURTHER EVALUATION WAS WARRANTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10484981 ↗MalfunctionMar 20, 2020Fracture; Fracture
ON (B)(6) 2020, NEOTRACT WAS INFORMED OF A SUCCESSFUL PROSTATIC URETHRAL LIFT (PUL) PROCEDURE DURING WHICH THE PHYSICIAN COULD NOT VISUALIZE THE SUTURE AFTER NEEDLE DEPLOYMENT. NO PATIENT INJURY OR HARM WAS REPORTED. ON 20 FEB 2020, NEOTRACT'S INVESTIGATION OF THE DEVICE INDICATED THAT 2MM OF THE NEEDLE TIP WAS MISSING. NEOTRACT NOTIFIED THE PHYSICIAN OF THE INVESTIGATION RESULTS. ON 11 MARCH 2020, THE PHYSICIAN ACKNOWLEDGED THE MISSING NEEDLE PIECE. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME DESPITE MULTIPLE OUTREACH ATTEMPTS. .
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9859631 ↗MalfunctionMar 5, 2020Fracture; Fracture
ON 10 JAN 2020, NEOTRACT WAS INFORMED OF SUCCESSFUL PROSTATIC URETHRAL LIFT (PUL) PROCEDURE DURING WHICH A DELIVERY DEVICE DID NOT DEPLOYED PROPERLY, THERE WAS NO PATIENT INJURY OR HARM REPORTED, AND THE DEVICE WAS RETURNED FOR INVESTIGATION. ON 04 FEB 2020, NEOTRACT'S DEVICE ANALYSIS CONFIRMED THAT A 1 -2MM NEEDLE TIP WAS MISSING FROM THE RETURNED DEVICE. NEOTRACT HAD NOTIFIED THE PHYSICIAN OF THE MISSING NEEDLE TIP.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9793193 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K173087. One submission often covers many models, so check the product and lot details in each record.
- Z-0598-2019 · Jul 6, 2018 · TerminatedUroLift System REF Catalog Number: UL400-4 Tray Contents: UroLift System with Handle Release Tool STERILE R CE 0050 Manufactured By: neotract
Other Implantable Transprostatic Tissue Retractor System devices
- UroLift UL400-4Neotract, Inc. · Implantable Transprostatic Tissue Retractor System1 recall9 deaths144 reports
- UROLIFT IPN924184Teleflex Incorporated · Implantable transprostatic tissue retractor system1 death32 reports
- UROLIFT IPN918876Teleflex Incorporated · Implantable transprostatic tissue retractor system18 reports
- UROLIFT IPN923352Teleflex Incorporated · Implantable transprostatic tissue retractor system15 reports
- UROLIFT IPN055826Teleflex Incorporated · Implantable transprostatic tissue retractor system15 reports
- UROLIFT IPN056134Teleflex Incorporated · Implantable transprostatic tissue retractor system1 death10 reports
- UROLIFT IPN056885Teleflex Incorporated · Implantable transprostatic tissue retractor system10 reports
- UROLIFT IPN919209Teleflex Incorporated · Implantable transprostatic tissue retractor system9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page