Su-Por Surgical Implants 4294Recalls & reported problems

Poriferous, LLC · Ear Wedge Thin - Left - 15mm x 44mm x 11mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00815285022131), received between Jan 15, 2020 and Jan 15, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Appears to Trigger Rejection1

Patient problems

  • Pocket Erosion1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 15, 2020Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    AN IMPLANT REQUIRED REMOVAL DUE TO EXPOSURE. UPON EXAMINATION OF THE REMOVED IMPLANT, NO SIGNS OF INFECTION WERE PRESENT. THE SURGEON ORDERED A CULTURE OF THE EXPLANTED DEVICE. RESULTS WERE NEGATIVE FOR INFECTION. THE AREA OF TISSUE BREAKDOWN AND EXPOSURE WERE LOCATED IN WHAT THE SURGEON DESCRIBED AS "AREAS OF POTENTIAL PRESSURE NECROSIS". IF THE SURGICAL DRESSING CUP IS DISPLACES, IT IS POSSIBLE FOR THE PRESSURE ON THE EAR TO CAUSE TISSUE BREAKDOWN. ALSO EXPRESSED BY THE SURGEON IS THE DIFFICULTY TO GET THE PATIENT TO FOLLOW POST OPERATIVE INSTRUCTIONS.

    Patient problems: Pocket Erosion

    MAUDE report 9594282 ↗

Other Material, Polytetrafluoroethylene Vitreous Carbon, For Maxillofacial Reconstruction devices

All Material, Polytetrafluoroethylene Vitreous Carbon, For Maxillofacial Reconstruction devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page