Reform Pedicle Screw System 39-SP-0730Recalls & reported problems
Precision Spine, Inc. · Polyaxial Driver Assembly · Class 1
All models of Reform Pedicle Screw System (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00815362022016), received between Jul 12, 2021 and Aug 18, 2025 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 18, 2025Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT A PROCEDURE ON (B)(6) 2024, UTILIZING THE REFORM PEDICLE SCREW SYSTEM. A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2025 TO EXTEND THE EXISTING CONSTRUCT. DURING THE PROCEDURE IT WAS NOTED THAT ONE OF THE Ø6.5MM X 50MM REFORM TI MODULAR NON-CANNULATED - NON-FLUTED SCREW (39-SG-6550) WAS FRACTURED. THE BROKEN SCREW WAS REMOVED AND REPLACED WITH NEW HARDWARE. IT WAS ALSO NOTED THAT WHILE IMPLANTING A SCREW THE TIP OF THE T20 POLYAXIAL DRIVER REFORM PEDICLE SCREW SYSTEM (39-SP-0730) BROKE. THE TIP WAS REMOVED AND THE PROCEDURE COMPLETED WITH NO FURTHER ISSUE AND NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22834571 ↗InjuryApr 18, 2022Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2022, TO ADDRESS A BROKEN REFORM PEDICLE SCREW, ATTRIBUTED TO NON-FUSION. THE ORIGINAL IMPLANTATION DATE WAS NOT PROVIDED. THE EXISTING CONSTRUCT WAS REMOVED AND UPON ATTEMPTED INSERTION OF A OF A REFORM 10.5 X 50MM POLYAXIAL PEDICLE SCREW (39-PA-1050) IN A NEWLY PREPARED PEDICLE, THE TIP OF THE REFORM POLYAXIAL DRIVER (HXX-39-0001) BROKE OFF IN THE HEAD OF THE SCREW. THE SCREW WAS THEN REMOVED AND SUCCESSFULLY REPLACED WITH ANOTHER SCREW USING A SECOND DRIVER IN THE SET. SUBSEQUENTLY, WHILE IMPLANTING ANOTHER REFORM 10.5 X 50MM POLYAXIAL PEDICLE SCREW (39-PA-1050) THE SCREW WAS 85% SEATED WHEN THE HEAD OF THE SCREW SHEARED OFF AT THE NECK BETWEEN THE SCREW HEAD AND THREADED SHAFT. THE THREADED SHAFT WAS LEFT IN THE PATIENT'S BONE AND THE PEDICLE WAS LEFT UN-FIXATED. THE CONSTRUCT WAS SUCCESSFULLY COMPLETED WITH NO HARM TO THE PATIENT. THERE WAS A DELAY OF ONE HOUR BECAUSE OF THE REPORTED MALFUNCTIONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14138140 ↗InjuryJul 12, 2021Break
IT WAS REPORTED THAT A PROCEDURE WAS PERFORMED ON AN UNKNOWN DATE, UTILIZING THE REFORM PEDICLE SCREW SYSTEM. WHILE ATTEMPTING TO INSERT A REFORM PEDICLE SCREW HA COATED, THE TIP OF THE DRIVER BROKE. THE BROKEN TIP WAS LEFT IN THE HEAD OF THE SCREW, IMPLANTED IN THE PATIENT. THERE WAS NO DELAY TO THE PROCEDURE.
Patient problems: Insufficient Information
MAUDE report 12150550 ↗
Other Screwdriver devices
- DELTA XTEND 2307-92-003Depuy (Ireland) · SCREWDRIVER751 reports
- EXPEDIUM 279712400Medos International Sàrl · SCREWDRIVER551 reports
- NA 314.467Synthes GmbH · SCREWDRIVER498 reports
- N/A 5584111Medtronic Sofamor Danek, Inc. · SCREWDRIVER492 reports
- QUICKSET 2274-47-000Depuy (Ireland) · SCREWDRIVER487 reports
- NA 03.130.010Synthes GmbH · SCREWDRIVER1 recall422 reports
- QUICKSET 2274-63-000Depuy (Ireland) · SCREWDRIVER404 reports
- EVOS 71175072Smith & Nephew, Inc. · SCREWDRIVER336 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page