Trevo 80052Recalls & reported problems
Stryker Corporation · Trevo XP Provue System
All models of Trevo (7)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00815742000528), received between May 11, 2020 and Sep 8, 2020 — classified as 2 injuries.
The most frequently reported specific device problem is “Difficult to Remove” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Apr 23, 2024Open, Classified
Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.
Z-1902-2024 · Root cause: Process design · Recall details · FDA record ↗
Most reported problems
Device problems
- Difficult to Remove1
- Fracture1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjurySep 8, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING THE PROCEDURE OF THROMBECTOMY, THE RETRIEVER (SUBJECT DEVICE) DETACHED DURING PROCEDURE. THE PHYSICIAN ATTEMPTED TO REMOVE THE SUBJECT RETRIEVER USING A SNARE DEVICE. ALL ATTEMPTS TO RETRIEVE THE SUBJECT DEVICE WERE UNSUCCESSFUL CAUSING A SURGICAL DELAY. THE SUBJECT RETRIEVER REMAINS IN M1 INSIDE THE PATIENT¿S ANATOMY. IT WAS REPORTED THAT THE PATIENT IS UNDER ADDED MEDICATION. THERE WERE NO CLINICAL CONSEQUENCES REPORTED TO THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 10504669 ↗InjuryMay 11, 2020Difficult to Remove; Difficult to Remove
IT WAS REPORTED THAT DURING THE THROMBECTOMY PROCEDURE, THE PHYSICIAN TRIED TO MOVE THE HIGH CALCIFICATION CLOT OUT OF THE PATIENT¿S ANATOMY WITH THE SUBJECT STENT RETRIEVER. DUE TO A NARROWING IN THE PATIENT¿S ANATOMY, THE PHYSICIAN COULD NOT MOVE THE RETRIEVER THROUGH THE INTERNAL CAROTID ARTERY(ICA). THE PHYSICIAN THEN CUT THE PUSHER WIRE AND LEFT THE SUBJECT RETRIEVER INSIDE THE VESSEL OF THE PATIENT¿S ANATOMY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO FURTHER INFORMATION IS AVAILABLE AT THE MOMENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10044707 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page