Convatec, Purchasing Department · Uncoated Male, 16”, Coude Tip · Class 2

Quick summary

This Cure urethral catheter is an uncoated male coude tip device. Reports mostly describe sharp edges on the catheter causing pain and tissue injury, sometimes minor bleeding. There are 95 adverse event reports: no deaths and 3 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
95
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 95 adverse event reports that list this exact device (DI 00815947020093), received between Sep 5, 2019 and Jun 3, 2026 — classified as 3 injuries, 92 malfunctions.

The most frequently reported specific device problem is “Sharp Edges” (92 reports).

1 other recall cover devices cleared under the same FDA submission (K072539); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Sharp Edges92
  • Adverse Event Without Identified Device or Use Problem3

Patient problems

  • Pain61
  • Unspecified Tissue Injury32
  • Urinary Tract Infection3
  • Fever1

Recent adverse event reports

The 25 most recent of 95 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 3, 2026Adverse Event Without Identified Device or Use Problem

    (B)(6) YEAR OLD END USER REPORTS HE HAS HAD "INFECTIONS IN THE BLADDER" IN HIS PAST AND HIS DOCTOR HAS NEEDED TO GIVE HIM MEDICINE FOR AT TIME BUT DOESN'T KNOW IF POSSIBLY THIS COULD BE CAUSED BY USING CATHETERS. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Urinary Tract Infection

    MAUDE report 25378462 ↗
  • MalfunctionApr 29, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14251039 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243796 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243787 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243637 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243629 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243622 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243422 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243415 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243387 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243379 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243372 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243370 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN JULY/AUG 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTIO

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243364 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243256 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243249 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243246 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243237 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243229 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN JULY/AUG 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTIO

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243222 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243209 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243201 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENT

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243142 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14243136 ↗
  • MalfunctionApr 28, 2022Sharp Edges; Sharp Edges

    CONSUMER STATES THE CATHETERS M16C FEEL LIKE SOME OF THEM HAVE A "BUR" ON THE END OF THEM THAT SCRATCH HIM AT THE VERY TIP WHEN THEY ENTER HIS BLADDER AND SOMETIMES CAUSE HIM TO HAVE MINOR BLEEDING AFTER. HE HAS USED THESE CATHETERS SINCE HE STARTED CATHING IN (B)(6) 2017 AND FEELS THE PROBLEM IS GETTING WORSE. HE THINKS IT IS A QUALITY CONTROL ISSUE. AT TIMES, HE HAS SMALL QUANTITIES OF BLOOD AS WELL THAT CAN BE VISIBLE AS HE RETRACTED THE CATHETER AND SOMETIMES HE NOTES IT WHEN HE CLEANS HIS MEATUS OR EVEN A SHORT TIME AFTER. HE SAID IT IS NOT A LOT OF BLOOD AND NOT NOTED IN HIS URINE. HE IS NOT ON BLOOD THINNERS. HE NOTES NO ABNORMALITY WITH THE CATHETER ITSELF VISIBLY PRIOR TO CATHING. HE SAID HE SEES NO ABNORMALITY WITH THE EYELETS. HE DOES SAY THAT HE SOMETIMES PUTS THE COUDE TIP IN HIS MEATUS TO THE RIGHT/ LEFT OR DOWN, NOT ALWAYS UPWARDS. I RELAYED TO HIM THE PROPER WAY TO PLACE A COUDE TIP IS UPWARDS AND HE SAID HE WILL TRY THAT. HE HAS NOT SEEN HIS UROLOGIST IN A LONG TIME OVER A YEAR STATING THEY JUST WANT TO GIVE HIM MEDICATIONS AND HE CAN TEACH HIS DOCTOR ABOUT HIS CONDITION. HE SAID HE HAS AN APPOINTMENT NEXT WEEK WITH HIS PCP. HE STARTED CATHING BECAUSE OF RETENTION

    Patient problems: Unspecified Tissue Injury

    MAUDE report 14242991 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K072539. One submission often covers many models, so check the product and lot details in each record.

  • Z-0004-2013 · Sep 19, 2012 · Terminated
    Male, 16Fr, Straight-tip Intermittent Urinary Catheters M16. The product is distributed individually or in intermediate boxes containing 30 individual, sterile packaged catheters (

Other Catheter, Urethral devices

All Catheter, Urethral devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.