Indigo System CAT7XTORQ130KIT-ARecalls & reported problems
Penumbra, Inc. · Aspiration Catheter 7, 130cm XTORQ Kit · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00815948021648), received between Feb 17, 2021 and Mar 26, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Fracture1
- Material Deformation1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 26, 2021Material Deformation; Physical Resistance/Sticking; Material Deformation; Physical Resistance/Sticking
THE PATIENT UNDERWENT A THROMBECTOMY PROCEDURE IN THE POPLITEAL USING AN INDIGO SYSTEM ASPIRATION CATHETER 7 (CAT7), A NON-PENUMBRA SHEATH, AND A GUIDEWIRE. IT SHOULD BE NOTED THAT THE PATIENT HAD TORTUOUS ILIACS AND BIFURCATIONS. DURING THE PROCEDURE, THE PHYSICIAN COMPLETED A FEW PASSES USING THE CAT7. WHILE WITHDRAWING THE CAT7 FROM THE PATIENT, THE PHYSICIAN ENCOUNTERED RESISTANCE. AFTER THE CAT7 WAS REMOVED FROM THE PATIENT, THE PHYSICIAN NOTICED THE CAT7 HAD KINKED MID-SHAFT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AT THIS POINT AND NO ADDITIONAL PASSES WERE REQUIRED. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11574916 ↗MalfunctionFeb 17, 2021Fracture
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE OF AN OCCLUDED STENT IN THE SUPERFICIAL FEMORAL ARTERY (SFA) USING AN INDIGO SYSTEM ASPIRATION CATHETER 7 (CAT7) AND NON-PENUMBRA SHEATH. DURING THE PROCEDURE, THE TIP OF THE CAT7 KINKED WHEN ADVANCED OUT OF THE SHEATH WITHOUT A GUIDEWIRE. IT WAS REPORTED THAT THIS MAY HAVE BEEN DUE TO AN INTERACTION WITH THE STENT. AN 0.035" GUIDE WIRE WAS THEN ADVANCED THROUGH THE CAT7 TO STRAIGHTEN OUT THE TIP. THE GUIDEWIRE WAS THEN REMOVED. THE PHYSICIAN THEN TORQUED THE CAT7 ONE FULL ROTATION AND NOTICED THE DISTAL TIP WAS NOT RESPONDING. THE CATHETER HAD FRACTURED PROXIMAL TO THE VASCULAR SHEATH VALVE, OUTSIDE OF THE PATIENT. SUBSEQUENTLY, THE CAT7 WAS REMOVED. THE PROCEDURE WAS COMPLETED USING AN INDIGO SYSTEM ASPIRATION CATHETER 8 (CAT8) AND THE SAME SHEATH. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11337835 ↗
Other Peripheral Mechanical Thrombectomy With Aspiration devices
- Indigo System LITNG12HTORQ115-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration4 deaths105 reports
- Indigo System LITNG7XTORQ130-APenumbra, Inc. · Peripheral mechanical thrombectomy with aspiration100 reports
- Indigo System LITNG12XTORQ100-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration1 death80 reports
- ClotTriever XL Catheter AS-0235Inari Medical, Inc. · Peripheral Mechanical Thrombectomy With Aspiration1 recall9 deaths20 reports
- Indigo System LITNG12HTORQ100-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration1 death46 reports
- Indigo System LITNGBT7TQ130-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration46 reports
- Triever 24 22-101Inari Medical, Inc. · Peripheral Mechanical Thrombectomy With Aspiration20 deaths34 reports
- ClotTriever Catheter, 16 mm 40-102Inari Medical, Inc. · Peripheral Mechanical Thrombectomy With Aspiration11 deaths27 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.