Skip to content

Indigo System CAT7XTORQ130KIT-ARecalls & reported problems

Penumbra, Inc. · Aspiration Catheter 7, 130cm XTORQ Kit · Class 2

All models of Indigo System (64)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00815948021648), received between Feb 17, 2021 and Mar 26, 2021 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Fracture1
  • Material Deformation1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 26, 2021Material Deformation; Physical Resistance/Sticking; Material Deformation; Physical Resistance/Sticking

    THE PATIENT UNDERWENT A THROMBECTOMY PROCEDURE IN THE POPLITEAL USING AN INDIGO SYSTEM ASPIRATION CATHETER 7 (CAT7), A NON-PENUMBRA SHEATH, AND A GUIDEWIRE. IT SHOULD BE NOTED THAT THE PATIENT HAD TORTUOUS ILIACS AND BIFURCATIONS. DURING THE PROCEDURE, THE PHYSICIAN COMPLETED A FEW PASSES USING THE CAT7. WHILE WITHDRAWING THE CAT7 FROM THE PATIENT, THE PHYSICIAN ENCOUNTERED RESISTANCE. AFTER THE CAT7 WAS REMOVED FROM THE PATIENT, THE PHYSICIAN NOTICED THE CAT7 HAD KINKED MID-SHAFT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AT THIS POINT AND NO ADDITIONAL PASSES WERE REQUIRED. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11574916 ↗
  • MalfunctionFeb 17, 2021Fracture

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE OF AN OCCLUDED STENT IN THE SUPERFICIAL FEMORAL ARTERY (SFA) USING AN INDIGO SYSTEM ASPIRATION CATHETER 7 (CAT7) AND NON-PENUMBRA SHEATH. DURING THE PROCEDURE, THE TIP OF THE CAT7 KINKED WHEN ADVANCED OUT OF THE SHEATH WITHOUT A GUIDEWIRE. IT WAS REPORTED THAT THIS MAY HAVE BEEN DUE TO AN INTERACTION WITH THE STENT. AN 0.035" GUIDE WIRE WAS THEN ADVANCED THROUGH THE CAT7 TO STRAIGHTEN OUT THE TIP. THE GUIDEWIRE WAS THEN REMOVED. THE PHYSICIAN THEN TORQUED THE CAT7 ONE FULL ROTATION AND NOTICED THE DISTAL TIP WAS NOT RESPONDING. THE CATHETER HAD FRACTURED PROXIMAL TO THE VASCULAR SHEATH VALVE, OUTSIDE OF THE PATIENT. SUBSEQUENTLY, THE CAT7 WAS REMOVED. THE PROCEDURE WAS COMPLETED USING AN INDIGO SYSTEM ASPIRATION CATHETER 8 (CAT8) AND THE SAME SHEATH. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11337835 ↗

FDA clearance and classification

Cleared through 510(k) K193595 on Aug 19, 2020 (decision: Substantially Equivalent). Submitted by Penumbra, Inc. as “Indigo Aspiration System - Aspiration Catheter 7 and Separator 7”. The FDA review took 240 days. FDA record ↗

The FDA regulates product code QEW (peripheral mechanical thrombectomy with aspiration) under 21 CFR 870.5150 as a Class II device. 21 CFR 870.5150 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Indigo System CAT7XTORQ130KIT-A

Has the Indigo System CAT7XTORQ130KIT-A been recalled?

No FDA recall record in our data names this exact device identifier (DI 00815948021648), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Indigo System CAT7XTORQ130KIT-A?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Feb 17, 2021 and Mar 26, 2021. In those reports, the most common specific device problems are “Fracture” (1), “Material Deformation” (1) and “Physical Resistance/Sticking” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Indigo System CAT7XTORQ130KIT-A?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Indigo System CAT7XTORQ130KIT-A FDA approved?

It was cleared through 510(k) K193595 on Aug 19, 2020. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Indigo System CAT7XTORQ130KIT-A, and is it still sold?

The FDA's device database lists Penumbra, Inc. as the company responsible for it (brand Indigo System, model CAT7XTORQ130KIT-A). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Peripheral Mechanical Thrombectomy With Aspiration devices

All Peripheral Mechanical Thrombectomy With Aspiration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page