Skip to content

Indigo System CAT8XTORQ115-ARecalls & reported problems

Penumbra, Inc. · Asp Cat 8, 115cm XTORQ · Class 2

All models of Indigo System (64)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00815948022478), received between Dec 17, 2020 and Sep 7, 2021 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Material Deformation3
  • Break1
  • Physical Resistance/Sticking1
  • Suction Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 7, 2021Material Deformation; Material Deformation

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE USING AN INDIGO SYSTEM ASPIRATION CATHETER 8 (CAT8) AND NON-PENUMBRA SHEATH. DURING THE PROCEDURE, THE CAT8 KINKED WHILE ADVANCING THROUGH THE SHEATH. THEREFORE, THE CAT8 WAS REMOVED. THE PROCEDURE COMPLETED USING A NEW CAT8 WITH A PEEL AWAY SHEATH. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12435983 ↗
  • MalfunctionMay 30, 2021Break; Break

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY (SFA) USING AN INDIGO SYSTEM ASPIRATION CATHETER 8 (CAT8). DURING THE PROCEDURE, THE HUB OF THE CAT8 BROKE WHILE TORQUING. THEREFORE, THE CAT8 WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEW CAT8. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11905884 ↗
  • MalfunctionMar 16, 2021Material Deformation; Suction Failure; Material Deformation; Suction Failure

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE DEEP FEMORAL VEIN USING AN INDIGO SYSTEM ASPIRATION CATHETER 8 (CAT8) AND A NON-PENUMBRA SHEATH. DURING THE PROCEDURE, THE CAT8 DID NOT ASPIRATE AND SUBSEQUENTLY, THE PHYSICIAN NOTICED THE CAT8 WAS FLATTENED IN TWO LOCATIONS MID-SHAFT. IT WAS UNKNOWN WHETHER THE DAMAGE OCCURRED UPON REMOVAL FROM THE PACKAGING OR WHILE ATTEMPTING TO ASPIRATE. THEREFORE, THE CAT8 WAS NO LONGER USED. THE PROCEDURE WAS COMPLETED USING ANOTHER CAT8 AND THE SAME SHEATH. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11497930 ↗
  • MalfunctionDec 17, 2020Material Deformation; Physical Resistance/Sticking; Material Deformation; Physical Resistance/Sticking

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE RIGHT COMMON FEMORAL ARTERY (CFA) USING AN INDIGO SYSTEM ASPIRATION CATHETER 8, A NON-PENUMBRA SHEATH AND A GUIDEWIRE. IT WAS NOTED THAT THE PATIENT ANATOMY WAS TORTUOUS. DURING THE PROCEDURE, WHILE THE PHYSICIAN ADVANCED THE CAT8 OVER THE GUIDEWIRE TO THE BIFURCATION, THE PHYSICIAN EXPERIENCED RESISTANCE AND SUBSEQUENTLY, APPROXIMATELY FIFTEEN CENTIMETERS FORM THE HUB OF THE CAT8 KINKED. THEREFORE, THE CAT8 WAS REMOVED. THE PROCEDURE WAS COMPLETED USING AN INDIGO SYSTEM ASPIRATION CATHETER 12 (CAT12) AND A NON-PENUMBRA SHEATH. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11032494 ↗

FDA clearance and classification

Cleared through Special 510(k) K200771 on Apr 22, 2020 (decision: Substantially Equivalent). Submitted by Penumbra, Inc. as “Indigo Aspiration System Lightning Aspiration Tubing”. The FDA review took 28 days. FDA record ↗

Cleared through 510(k) K192833 on Dec 20, 2019 (decision: Substantially Equivalent). Submitted by Penumbra, Inc. as “Indigo Aspiration System”. The FDA review took 79 days. FDA record ↗

The FDA regulates product code QEW (peripheral mechanical thrombectomy with aspiration) under 21 CFR 870.5150 as a Class II device. 21 CFR 870.5150 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Indigo System CAT8XTORQ115-A

Has the Indigo System CAT8XTORQ115-A been recalled?

No FDA recall record in our data names this exact device identifier (DI 00815948022478), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Indigo System CAT8XTORQ115-A?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Dec 17, 2020 and Sep 7, 2021. In those reports, the most common specific device problems are “Material Deformation” (3), “Break” (1) and “Physical Resistance/Sticking” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Indigo System CAT8XTORQ115-A?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Indigo System CAT8XTORQ115-A FDA approved?

It was cleared through 510(k) K192833 on Dec 20, 2019; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Indigo System CAT8XTORQ115-A, and is it still sold?

The FDA's device database lists Penumbra, Inc. as the company responsible for it (brand Indigo System, model CAT8XTORQ115-A). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Peripheral Mechanical Thrombectomy With Aspiration devices

All Peripheral Mechanical Thrombectomy With Aspiration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page