Skip to content

Indigo System LITNG12XTORQ115-ARecalls & reported problems

Penumbra, Inc. · CAT12 115 New XTORQ shape plus Lightning · Class 2

All models of Indigo System (64)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00815948022591), received between Nov 5, 2020 and Nov 5, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Circuit Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Circuit Failure1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 5, 2020Circuit Failure; Obstruction of Flow; Circuit Failure; Obstruction of Flow

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE FEMORAL AND ILIAC VEIN USING A LIGHTNING ASPIRATION TUBING (LIGHTNING), AN INDIGO SYSTEM ASPIRATION CATHETER 12 (CAT12) AND A PENUMBRA ENGINE (ENGINE). DURING THE PROCEDURE, THERE WAS AN INITIAL STRONG ASPIRATION WHEN THE PHYSICIAN TURNED THE ENGINE ON. SUBSEQUENTLY, THE LIGHTNING CLOGGED AND THE INDICATOR LIGHT ON THE LIGHTNING UNIT TURNED FROM GREEN TO RED. THEREFORE, THE PHYSICIAN ATTEMPTED TO USE A SYRINGE TO FLUSH THE THROMBUS THROUGH THE LIGHTNING. THE PHYSICIAN WAS ABLE TO FLUSH THE MAJORITY OF THE THROMBUS OUT OF THE LIGHTNING; HOWEVER, THE INDICATOR LIGHT REMAINED RED. NEXT, THE PHYSICIAN UNPLUGGED AND RE-PLUGGED THE LIGHTNING UNIT FROM THE ENGINE A FEW TIMES BUT THE INDICATOR LIGHT REMAINED RED. LASTLY, THE PHYSICIAN POWER CYCLED THE ENGINE AND THE INDICATOR LIGHT REMAINED RED; THEREFORE, THE LIGHTNING WAS REMOVED AND NOT USED FOR THE REMAINDER OF THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW LIGHTNING AND THE SAME CAT12. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10794415 ↗

Other Peripheral Mechanical Thrombectomy With Aspiration devices

All Peripheral Mechanical Thrombectomy With Aspiration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page