Skip to content

Benchmark MAX System BMX9690MSIM125-ARecalls & reported problems

Penumbra, Inc. · BMX96 90cm, MP Tip, 6F SIM 125cm Select · Class 2

All models of Benchmark MAX System (13)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00815948022683), received between Nov 25, 2020 and May 27, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Suction Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 27, 2025Fracture; Fracture

    THE PATIENT WAS UNDERGOING A MEDICAL PROCEDURE IN THE RIGHT COMMON CAROTID ARTERY (CCA) USING A BENCHMARK BMX96 ACCESS SYSTEM (BMX96), A NON-PENUMBRA SHEATH, AND A GUIDEWIRE. DURING THE PROCEDURE, THE PHYSICIAN ADVANCED THE BMX96 AND THE GUIDEWIRE INTO THE RIGHT CCA. THE BMX96 WAS TORQUED IN THE COMMON ILIAC ARTERY AND BECAME DAMAGED. THE PHYSICIAN WAS UNABLE TO ADVANCE A GUIDEWIRE THROUGH THE BMX96. THE BMX96 THEN FRACTURED UPON RETRACTION BUT WAS STUCK IN THE SHEATH. THE PHYSICIAN THEN DECIDED TO REMOVE THE BMX96 AND THE SHEATH TOGETHER FROM THE PATIENT. THE PHYSICIAN CLAMPED THE BMX96 UPON REMOVING IT WITH THE SHEATH AND MADE ANOTHER FRACTURE IN ORDER TO MAINTAIN WIRE ACCESS. THE PROCEDURE WAS COMPLETED USING A NEW BMX96 AND A NEW NON-PENUMBRA SHEATH. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22092256 ↗
  • MalfunctionSep 4, 2024Fracture; Fracture

    THE PATIENT WAS UNDERGOING A MEDICAL PROCEDURE IN THE MIDDLE CEREBRAL ARTERY (MCA) USING A BENCHMARK BMX96 ACCESS SYSTEM (BMX96). DURING THE PROCEDURE, THE PHYSICIAN ADVANCED THE BMX96 INTO THE AORTIC ARCH. WHILE PERFORMING AN INJECTION, THE PHYSICIAN FRACTURED THE BMX96 AT THE HUB. THEREFORE, THE BMX96 WAS REMOVED. THE PROCEDURE WAS COMPLETED USING A NEURON MAX 6F 088 LONG SHEATH (NEURON MAX). THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20138525 ↗
  • MalfunctionNov 25, 2020Suction Failure; Suction Failure

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE INTERNAL CAROTID ARTERY (ICA) USING A PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER (3MAXC), A PENUMBRA SYSTEM JET 7 REPERFUSION CATHETER (JET7) AND BENCHMARK BMX96 ACCESS SYSTEM (BENCHMARK MAX). DURING THE PROCEDURE, THE PHYSICIAN ATTEMPTED TO ASPIRATE THROUGH THE BENCHMARK MAX WITH A 10CC SYRINGE BUT WAS UNABLE TO ASPIRATE FLUID. THEREFORE, THE PHYSICIAN DID NOT ATTEMPT TO INJECT CONTRAST THROUGH THE BENCHMARK MAX AROUND THE JET7. THE JET7 WAS SUBSEQUENTLY REMOVED. UPON REMOVAL, THE DISTAL END OF THE JET7 WAS OBSERVED TO BE KINKED. THE PROCEDURE ENDED AT THIS POINT. A THROMBOLYSIS IN CEREBRAL INFARCTION (TICI) GRADE 2B RECANALIZATION WAS ACHIEVED WITH TWO PASSES USING THE JET7. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10907083 ↗