Indigo System SEP7-ARecalls & reported problems

Penumbra, Inc. · Separator 7, 190cm · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00815948024014), received between Sep 3, 2021 and May 14, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation2
  • Break1
  • Fracture1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 14, 2024Fracture; Fracture

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY USING AN INDIGO SYSTEM SEPARATOR 7 (SEP7) AND AN INDIGO SYSTEM ASPIRATION CATHETER 7 (CAT7). IT SHOULD BE NOTED THAT THE PATIENT''S ANATOMY WAS TORTUOUS. DURING THE PROCEDURE, THE PHYSICIAN PERFORMED MULTIPLE PASSES IN THE TARGET LOCATION USING THE CAT7 AND SEP7. WHILE MAKING ANOTHER PASS, THE PHYSICIAN OBSERVED UNDER FLUOROSCOPY THAT THE SEP7 HAD FRACTURED IN THE POPLITEAL ARTERY. THEREFORE, THE SEP7 WAS REMOVED FROM THE PATIENT AND THE FRACTURED PORTION WAS SUCCESSFULLY ASPIRATED OUT USING THE CAT7. THE PROCEDURE WAS COMPLETED USING A NEW SEP7 AND THE SAME CAT7. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19315742 ↗
  • MalfunctionJan 31, 2022Break; Break

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE TO TREAT A BRACHIOCEPHALIC FISTULA USING AN INDIGO SYSTEM SEPARATOR 7 (SEP7) AND AN INDIGO SYSTEM ASPIRATION CATHETER 7 (CAT7). DURING THE PROCEDURE, THE PHYSICIAN COMPLETED TWO PASSES USING THE SEP7 AND CAT7. DURING THE THIRD PASS, THE BULB OF THE SEP7 BROKE OFF INTO THE FISTULA. THEREFORE, THE PHYSICIAN USED A SNARE DEVICE TO REMOVE THE BROKEN BULB OF THE SEP7. THE PROCEDURE WAS COMPLETED AT THIS POINT. IT WAS REPORTED THAT THE THROMBUS WAS OLD AND CHRONIC, AND THE FISTULA NEEDED TO BE ABANDONED. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13405546 ↗
  • MalfunctionDec 3, 2021Material Deformation; Physical Resistance/Sticking; Material Deformation; Physical Resistance/Sticking

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY (SFA) USING AN INDIGO SYSTEM SEPARATOR 7 (SEP7). DURING THE PROCEDURE, WHILE ATTEMPTING TO ADVANCE THE SEP7 INTO A ROTATING HEMOSTASIS VALVE (RHV), THE PHYSICIAN EXPERIENCED RESISTANCE AND KINKED THE SEP7 AT THE PROXIMAL END. THEREFORE, THE SEP7 WAS REMOVED. THE PROCEDURE WAS COMPLETED WITHOUT A SEP7 USING THE SAME RHV. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12926996 ↗
  • MalfunctionSep 3, 2021Material Deformation; Material Deformation

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA) USING AN INDIGO SYSTEM SEPARATOR 7 (SEP7), AN INDIGO SYSTEM ASPIRATION CATHETER 7 (CAT7), AND A NON-PENUMBRA SHEATH. DURING THE PROCEDURE, THE PHYSICIAN COMPLETED ONE PASS USING THE SEP7 AND CAT7. WHILE ADVANCING THE SEP7 THROUGH THE SHEATH TO MAKE THE NEXT PASS AND THE PROXIMAL END OF THE SEP7 WIRE BECAME KINKED. THEREFORE, THE SEP7 WAS REMOVED. THE PROCEDURE WAS COMPLETED USING THE CAT7 AND THE SAME SHEATH. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12424956 ↗

Other Peripheral Mechanical Thrombectomy With Aspiration devices

All Peripheral Mechanical Thrombectomy With Aspiration devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.