Indigo System SEP7-ARecalls & reported problems
Penumbra, Inc. · Separator 7, 190cm · Class 2
All models of Indigo System (64)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00815948024014), received between Sep 3, 2021 and May 14, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation2
- Break1
- Fracture1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 14, 2024Fracture; Fracture
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY USING AN INDIGO SYSTEM SEPARATOR 7 (SEP7) AND AN INDIGO SYSTEM ASPIRATION CATHETER 7 (CAT7). IT SHOULD BE NOTED THAT THE PATIENT''S ANATOMY WAS TORTUOUS. DURING THE PROCEDURE, THE PHYSICIAN PERFORMED MULTIPLE PASSES IN THE TARGET LOCATION USING THE CAT7 AND SEP7. WHILE MAKING ANOTHER PASS, THE PHYSICIAN OBSERVED UNDER FLUOROSCOPY THAT THE SEP7 HAD FRACTURED IN THE POPLITEAL ARTERY. THEREFORE, THE SEP7 WAS REMOVED FROM THE PATIENT AND THE FRACTURED PORTION WAS SUCCESSFULLY ASPIRATED OUT USING THE CAT7. THE PROCEDURE WAS COMPLETED USING A NEW SEP7 AND THE SAME CAT7. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19315742 ↗MalfunctionJan 31, 2022Break; Break
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE TO TREAT A BRACHIOCEPHALIC FISTULA USING AN INDIGO SYSTEM SEPARATOR 7 (SEP7) AND AN INDIGO SYSTEM ASPIRATION CATHETER 7 (CAT7). DURING THE PROCEDURE, THE PHYSICIAN COMPLETED TWO PASSES USING THE SEP7 AND CAT7. DURING THE THIRD PASS, THE BULB OF THE SEP7 BROKE OFF INTO THE FISTULA. THEREFORE, THE PHYSICIAN USED A SNARE DEVICE TO REMOVE THE BROKEN BULB OF THE SEP7. THE PROCEDURE WAS COMPLETED AT THIS POINT. IT WAS REPORTED THAT THE THROMBUS WAS OLD AND CHRONIC, AND THE FISTULA NEEDED TO BE ABANDONED. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13405546 ↗MalfunctionDec 3, 2021Material Deformation; Physical Resistance/Sticking; Material Deformation; Physical Resistance/Sticking
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY (SFA) USING AN INDIGO SYSTEM SEPARATOR 7 (SEP7). DURING THE PROCEDURE, WHILE ATTEMPTING TO ADVANCE THE SEP7 INTO A ROTATING HEMOSTASIS VALVE (RHV), THE PHYSICIAN EXPERIENCED RESISTANCE AND KINKED THE SEP7 AT THE PROXIMAL END. THEREFORE, THE SEP7 WAS REMOVED. THE PROCEDURE WAS COMPLETED WITHOUT A SEP7 USING THE SAME RHV. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12926996 ↗MalfunctionSep 3, 2021Material Deformation; Material Deformation
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA) USING AN INDIGO SYSTEM SEPARATOR 7 (SEP7), AN INDIGO SYSTEM ASPIRATION CATHETER 7 (CAT7), AND A NON-PENUMBRA SHEATH. DURING THE PROCEDURE, THE PHYSICIAN COMPLETED ONE PASS USING THE SEP7 AND CAT7. WHILE ADVANCING THE SEP7 THROUGH THE SHEATH TO MAKE THE NEXT PASS AND THE PROXIMAL END OF THE SEP7 WIRE BECAME KINKED. THEREFORE, THE SEP7 WAS REMOVED. THE PROCEDURE WAS COMPLETED USING THE CAT7 AND THE SAME SHEATH. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12424956 ↗
FDA clearance and classification
Cleared through Special 510(k) K210083 on Feb 9, 2021 (decision: Substantially Equivalent). Submitted by Penumbra, Inc. as “Indigo System Aspiration Catheter 7, Indigo Aspiration System Separator 7, Lightning 7”. The FDA review took 28 days. FDA record ↗
The FDA regulates product code QEW (peripheral mechanical thrombectomy with aspiration) under 21 CFR 870.5150 as a Class II device. 21 CFR 870.5150 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Indigo System SEP7-A
Has the Indigo System SEP7-A been recalled?
No FDA recall record in our data names this exact device identifier (DI 00815948024014), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Indigo System SEP7-A?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Sep 3, 2021 and May 14, 2024. In those reports, the most common specific device problems are “Material Deformation” (2), “Break” (1) and “Fracture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Indigo System SEP7-A?
None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Indigo System SEP7-A FDA approved?
It was cleared through Special 510(k) K210083 on Feb 9, 2021. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Indigo System SEP7-A, and is it still sold?
The FDA's device database lists Penumbra, Inc. as the company responsible for it (brand Indigo System, model SEP7-A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Peripheral Mechanical Thrombectomy With Aspiration devices
- Indigo System LITNG12HTORQ115-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration4 deaths105 reports
- Indigo System LITNG7XTORQ130-APenumbra, Inc. · Peripheral mechanical thrombectomy with aspiration100 reports
- Indigo System LITNG12XTORQ100-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration1 death80 reports
- ClotTriever XL Catheter AS-0235Inari Medical, Inc. · Peripheral Mechanical Thrombectomy With Aspiration1 recall9 deaths21 reports
- Indigo System LITNGBT7TQ130-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration46 reports
- Indigo System LITNG12HTORQ100-APenumbra, Inc. · Peripheral Mechanical Thrombectomy With Aspiration1 death46 reports
- Triever 24 22-101Inari Medical, Inc. · Peripheral Mechanical Thrombectomy With Aspiration20 deaths34 reports
- ClotTriever Catheter, 16 mm 40-102Inari Medical, Inc. · Peripheral Mechanical Thrombectomy With Aspiration11 deaths27 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page