RED 68 RED68-SRecalls & reported problems
Penumbra, Inc. · Reperfusion Catheter RED 68, 132cm, CE · Class 2
All models of RED 68 (3)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00815948024090), received between Nov 21, 2022 and Jul 8, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Physical Resistance/Sticking” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K211411); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Physical Resistance/Sticking2
- Break1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 8, 2024Fracture; Physical Resistance/Sticking; Fracture; Physical Resistance/Sticking
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE LEFT INTERNAL CAROTID ARTERY (ICA) TERMINUS USING A PENUMBRA SYSTEM RED 68 REPERFUSION CATHETER (RED68), A PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER (3MAXC), A NON-PENUMBRA SHEATH, A STENT RETRIEVER, AND A GUIDEWIRE. IT WAS NOTED THAT THE PATIENT'S ANATOMY WAS TORTUOUS. DURING THE PROCEDURE, THE PHYSICIAN COMPLETED PASSES IN THE TARGET LOCATION USING THE RED68 AND 3MAXC. THE PHYSICIAN THEN RE-ADVANCED THE RED68 AND 3MAXC TO THE CLOT AND ATTEMPTED TO ADVANCE THE STENT RETRIEVER THROUGH THE RED68. IT WAS REPORTED THAT RESISTANCE WAS ENCOUNTERED WHEN PLACING THE RED68 AND STENT RETRIEVER DUE TO THE PATIENT¿S ANATOMY, CAUSING THE PHYSICIAN TO USE ADDITIONAL FORCE. AFTER SEVERAL ATTEMPTS TO ADVANCE THE STENT RETRIEVER, THE PHYSICIAN DECIDED TO REMOVE IT WITH THE RED68 AND 3MAXC AND AGAIN EXPERIENCED RESISTANCE. UPON REMOVAL, IT WAS NOTED THAT BOTH THE RED68 AND THE 3MAXC FRACTURED INTO TWO PIECES APPROXIMATELY FIVE CENTIMETERS FROM THE PROXIMAL END. THEREFORE, THE RED68 AND THE 3MAXC WERE NOT USED FOR THE REMAINDER OF THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NON-PENUMBRA ASPIRATION CATHETER AND THE SAME SHEATH. THERE W
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19696754 ↗MalfunctionNov 21, 2022Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking
THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE LEFT MIDDLE CEREBRAL ARTERY (MCA) USING A PENUMBRA SYSTEM RED 68 REPERFUSION CATHETER (RED68) AND A NON-PENUMBRA SHEATH. DURING THE PROCEDURE, THE PHYSICIAN PLACED THE SHEATH IN THE PATIENT USING RIGHT RADIAL ACCESS. WHILE ADVANCING THE RED68 THROUGH THE SHEATH, THE PHYSICIAN EXPERIENCED RESISTANCE AND SUBSEQUENTLY, BROKE THE RED68 INTO TWO PIECES AT THE MID-SHAFT AREA. THEREFORE, THE PHYSICIAN USED A SNARE DEVICE TO SUCCESSFULLY REMOVE THE BROKEN RED68. THE PROCEDURE WAS COMPLETED USING THE SAME SHEATH, ANOTHER CATHETER, AND A STENT RETRIEVER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15840083 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K211411. One submission often covers many models, so check the product and lot details in each record.
- Z-2770-2026 · Jun 3, 2026 · Open, ClassifiedPenumbra System RED 68 Reperfusion Catheter REF: RED68KIT
Other Catheter, Thrombus Retriever devices
- Penumbra System 5MAXJET7KIT-BPenumbra, Inc. · Catheter, Thrombus Retriever1 recall14 deaths221 reports
- Penumbra System 3MAXCPenumbra, Inc. · CATHETER, THROMBUS RETRIEVER1 death241 reports
- EMBOTRAP ET309537Neuravi Limited · Catheter, Thrombus Retriever13 deaths177 reports
- ACE 5MAXACE068KITPenumbra, Inc. · CATHETER, THROMBUS RETRIEVER170 reports
- Embotrap ET009533Neuravi Limited · Catheter, Thrombus Retriever16 deaths153 reports
- Penumbra System 5MAXACE132Penumbra, Inc. · CATHETER, THROMBUS RETRIEVER3 deaths112 reports
- Penumbra System 5MAXJET7KIT-APenumbra, Inc. · Catheter, Thrombus Retriever1 death94 reports
- Trevo XP 90185Stryker Corporation · CATHETER, THROMBUS RETRIEVER1 recall1 death40 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page