Thunderbolt + RED 62 THNDRRED62SKIT-ARecalls & reported problems

Penumbra, Inc. · RED 62 Reperfusion Catheter, 138 cm + THUNDERBOLT Aspiration Tubing

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
1
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00815948024359), received between Apr 3, 2023 and Jan 10, 2024 — classified as 1 death, 3 injuries.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 24, 2026Class IIOpen, Classified

    Due to manufacturing issues reperfusion catheters may fracture.

    Z-3331-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4

Patient problems

  • Intracranial Hemorrhage2
  • Cerebral Edema1
  • Ischemia Stroke1
  • Perforation of Vessels1
  • Thromboembolism1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • DeathJan 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2023, THE PATIENT UNDERWENT A THROMBECTOMY PROCEDURE IN THE ANTERIOR CIRCULATION OF THE LEFT M2 SEGMENT OF THE MIDDLE CEREBRAL ARTERY (MCA) USING A PENUMBRA SYSTEM RED 62 REPERFUSION CATHETER (RED62), A PENUMBRA SYSTEM RED 43 REPERFUSION CATHETER (RED43), A PENUMBRA SYSTEM 3MAX REPERFUSION CATHETER (3MAXC), A BENCHMARK BMX96 ACCESS SYSTEM (BMX96), A NEURON SELECT CATHETER 5F (5F SELECT), AN ASPIRATION TUBING (TUBING), A PENUMBRA ENGINE (ENGINE), A PENUMBRA ENGINE CANISTER (CANISTER), AND A GUIDEWIRE. IT WAS REPORTED THAT THE PATIENT RECEIVED 18 MG OF THROMBOLYTIC THERAPY AND SINGLE ANTIPLATELET MEDICATION PRIOR TO THE PROCEDURE. DURING THE PROCEDURE, THE PHYSICIAN COMPLETED TWO PASSES IN THE LEFT M2 SEGMENT USING THE RED62. THE RED62 WAS RETRACTED TO PERFORM AN ANGIOGRAM IN THE LEFT INTERNAL CAROTID ARTERY (ICA). ADDITIONAL CLOT WAS NOTICED IN THE RIGHT PERICALLOSAL ARTERY. THE PHYSICIAN THEN ADVANCED A RED43 INTO THE DISTAL RIGHT ANTERIOR CEREBRAL ARTERY (ACA) AND INTO THE RIGHT PERICALLOSAL ARTERY AND COMPLETED ONE PASS USING THE RED43. THROMBOLYSIS IN CEREBRAL INFARCTION (TICI) SCORE OF 3 WAS ACHIEVED. POST-THROMBECTOMY, IT WAS REPORTED THE PATIENT CONTINUED TO HAVE PE

    Patient problems: Cerebral Edema; Ischemia Stroke

    MAUDE report 18481441 ↗
  • InjuryMay 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT UNDERWENT A THROMBECTOMY PROCEDURE IN THE LEFT M2 SEGMENT OF THE MIDDLE CEREBRAL ARTERY (MCA) USING A PENUMBRA SYSTEM RED 62 REPERFUSION CATHETER (RED62). ON (B)(6) 2023, 24 HR FOLLOW-UP COMPUTED TOMOGRAPHY (CT) DEMONSTRATED A SMALL-VOLUME SUBARACHNOID HEMORRHAGE PRESENT IN THE LEFT SYLVIAN FISSURE. THE SUBARACHNOID HEMORRHAGE WAS REPORTED TO BE A SERIOUS ADVERSE EVENT WITH A POSSIBLE RELATIONSHIP TO THE PENUMBRA SYSTEM AND A PROBABLE RELATIONSHIP TO THE INDEX PROCEDURE. THE EVENT IS CONSIDERED ONGOING. ON (B)(6) 2023, AN UPDATE WAS RECEIVED INDICATING THAT THE EVENT WAS RESOLVED AS THE PATIENT STARTED COUMADIN.

    Patient problems: Intracranial Hemorrhage

    MAUDE report 17012926 ↗
  • InjuryApr 30, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE RIGHT M2 SEGMENT OF THE MIDDLE CEREBRAL ARTERY (MCA) USING A PENUMBRA SYSTEM RED 62 REPERFUSION CATHETER (RED62), A VELOCITY DELIVERY MICROCATHETER (VELOCITY) AND A NON-PENUMBRA SHEATH. DURING THE PROCEDURE, THE PATIENT SUFFERED A VESSEL RUPTURE AFTER THE COMPLETION OF THE FIRST PASS. THEREFORE, THE VESSEL WAS TAMPONADED WITH A BALLOON; HOWEVER, THE EXTRAVASATION CONTINUED. THEREFORE, A COIL OCCLUSION WAS PERFORMED AND THE EVENT WAS CONSIDERED RESOLVED. THE VESSEL RUPTURE WAS REPORTED TO BE A SERIOUS ADVERSE EVENT WITH A POSSIBLE RELATIONSHIP TO THE PENUMBRA SYSTEM AND A DEFINITE RELATIONSHIP TO THE INDEX PROCEDURE. ON 22-DEC-2022, AN UPDATE WAS RECEIVED FROM THE SITE INDICATING THAT THE VESSEL RUPTURE WAS UNRELATED TO THE PENUMBRA SYSTEM. ON 04-APR-2023, ADDITIONAL INFORMATION INDICATED THAT THE CLINICAL EVENTS COMMITTEE (CEC) ADJUDICATED THAT THE VESSEL RUPTURE WAS DEFINITELY RELATED TO THE PENUMBRA SYSTEM AND THE INDEX PROCEDURE.

    Patient problems: Intracranial Hemorrhage; Perforation of Vessels

    MAUDE report 16841288 ↗
  • InjuryApr 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2022 , THE PATIENT UNDERWENT A THROMBECTOMY PROCEDURE IN THE RIGHT M2 SEGMENT OF THE MIDDLE CEREBRAL ARTERY (MCA) USING A PENUMBRA SYSTEM RED 62 REPERFUSION CATHETER (RED62), A NEURON MAX 6F 088 LONG SHEATH (NEURON MAX) AND A VELOCITY DELIVERY MICROCATHETER (VELOCITY). DURING THE PROCEDURE, ONE PASS WAS COMPLETED WITH MODIFIED THROMBOLYSIS IN CEREBRAL INFARCTION (MTICI3). POST PROCEDURE, ON (B)(6) 2022 , MRI OF THE BRAIN SHOWED EVIDENCE OF INFARCTS THROUGHOUT THE RIGHT MCA DISTRIBUTION INVOLVING MOSTLY THE RIGHT FRONTAL LOBE AND RIGHT BASAL GANGLIA MOVEMENT AS WELL. THERE WERE ALSO INFARCTS PRESENT IN THE RIGHT MCA DISTRIBUTION. ON (B)(6) 2022 , THE PATIENT WAS DISCHARGED TO SKILLED NURSING FACILITY. MEDICATION ELIQUIS 5 MG TWICE DAILY WAS STARTED FROM (B)(6) 2022 AND THE PATIENT WAS CONTINUED ON ASPIRIN DAILY. IT WAS REPORTED THAT THE EVENT WAS UNRESOLVED AT TIME OF STUDY EXIT OR DEATH DUE TO A DIFFERENT EVENT. ON (B)(6) 2023, IT WAS REPORTED THAT THE CEC CHAIR ADJUDICATED INFARCTS POST PROCEDURE TO BE AN ADVERSE EVENT WITH A POSSIBLE RELATIONSHIP TO THE INDEX PROCEDURE, A POSSIBLE RELATIONSHIP TO THE INDEX STROKE AND A POSSIBLE RELATIONSHIP TO THE PENUMBRA SYSTEM.

    Patient problems: Thromboembolism

    MAUDE report 16661137 ↗

FDA clearance and classification

Cleared through 510(k) K250690 on Jun 10, 2026 (decision: Substantially Equivalent). Submitted by Penumbra, Inc. as “Penumbra System (Thunderbolt Aspiration Tubing)”. The FDA review took 460 days. FDA record ↗

Common questions about the Thunderbolt + RED 62 THNDRRED62SKIT-A

Has the Thunderbolt + RED 62 THNDRRED62SKIT-A been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-3331-2026 (Class II), began on Jul 24, 2026: Due to manufacturing issues reperfusion catheters may fracture. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Thunderbolt + RED 62 THNDRRED62SKIT-A?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Apr 3, 2023 and Jan 10, 2024. In those reports, the most common patient problems are “Intracranial Hemorrhage” (2), “Cerebral Edema” (1) and “Ischemia Stroke” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Thunderbolt + RED 62 THNDRRED62SKIT-A?

1 report is classified as a death, 3 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Thunderbolt + RED 62 THNDRRED62SKIT-A FDA approved?

It was cleared through 510(k) K250690 on Jun 10, 2026. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Thunderbolt + RED 62 THNDRRED62SKIT-A, and is it still sold?

The FDA's device database lists Penumbra, Inc. as the company responsible for it (brand Thunderbolt + RED 62, model THNDRRED62SKIT-A). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.