Benchmark MAX System BMX7F105-ARecalls & reported problems

Penumbra, Inc. · BMX81 105cm · Class 2

All models of Benchmark MAX System (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00815948024601), received between Oct 4, 2024 and Sep 17, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Physical Resistance/Sticking2

Patient problems

  • Foreign Body In Patient1
  • Perforation of Vessels1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2025Fracture; Physical Resistance/Sticking; Fracture; Physical Resistance/Sticking

    THE PATIENT WAS UNDERGOING A THROMBECTOMY IN THE COMMON CAROTID ARTERY (CCA) USING A BENCHMARK BMX81 ACCESS SYSTEM (BMX81), A NON-PENUMBRA SHEATH (7F), A NON-PENUMBRA ASPIRATION CATHETER, AND A GUIDEWIRE. DURING THE PROCEDURE, THE PHYSICIAN GAINED ACCESS TO THE RADIAL ARTERY USING THE SHEATH. WHILE ADVANCING THE BMX81 THROUGH THE SHEATH OVER THE GUIDEWIRE TOWARD THE SUBCLAVIAN ARTERY, THE PHYSICIAN EXPERIENCED RESISTANCE. IT WAS NOTED THAT THE PHYSICIAN WAS ABLE TO ADVANCE THE BMX81 TO THE CCA. THE PHYSICIAN THEN ADVANCED THE ASPIRATION CATHETER THROUGH THE BMX81 OVER THE GUIDEWIRE TO THE TARGET LOCATION AND PERFORMED ASPIRATION. AFTER COMPLETION OF THE PROCEDURE, THE PHYSICIAN REMOVED THE ASPIRATION CATHETER. WHILE REMOVING THE BMX81 FROM THE PATIENT'S BODY, THE PHYSICIAN EXPERIENCED RESISTANCE AND NOTICED THAT THE BMX81 BECAME DAMAGED AT THE MIDSHAFT OUTSIDE OF THE PATIENT'S BODY. THE PHYSICIAN THEN USED A HEMOSTAT TO CLAMP THE BMX81 AND THE SHEATH TOGETHER TO PREVENT AIR FROM ENTERING. WHILE SLOWLY REMOVING THE BMX81 AND THE SHEATH USING THE HEMOSTAT, THE PHYSICIAN EXPERIENCED RESISTANCE AND NOTICED THAT THE RADIAL ARTERY CAME OUT ALONG WITH THE DAMAGED BMX81. THE PHYSICIAN HELD

    Patient problems: Perforation of Vessels

    MAUDE report 23073868 ↗
  • MalfunctionOct 4, 2024Fracture; Physical Resistance/Sticking; Fracture; Physical Resistance/Sticking

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE MIDDLE MENINGEAL ARTERY USING A BENCHMARK BMX81 ACCESS SYSTEM (BMX81) AND A NON-PENUMBRA SHEATH (7FR). DURING THE PROCEDURE, THE PHYSICIAN ACCESSED THE TARGET VESSEL THROUGH THE RIGHT RADIAL ARTERY AND ADMINISTERED MEDICATIONS (VERAPAMIL, NITROGLYCERIN, AND HEPARIN). AFTER PLACING THE SHEATH, THE BMX81 WAS ADVANCED INTO THE VESSEL. WHILE ADVANCING THE BMX81 INTO THE VESSEL, THE PHYSICIAN ENCOUNTERED RESISTANCE AND DECIDED TO REMOVE THE BMX81. AFTER REMOVAL, THE PHYSICIAN NOTICED THAT THE BMX81 HAD FRACTURED ON THE DISTAL END NEAR THE MARKERBAND. IT WAS NOTED THAT THE TIP OF THE BMX81 HAD MOVED INTO AN ADJACENT ARTERY. AFTER ASSESSING WITH A SECOND PHYSICIAN, IT WAS DECIDED TO LEAVE THE FRACTURED TIP IN THE PATIENT''S VESSEL. THE PROCEDURE WAS COMPLETED WITH A NEW BMX81 WITH RIGHT FEMORAL ACCESS.

    Patient problems: Foreign Body In Patient

    MAUDE report 20377689 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →