Skip to content

Benchmark MAX System BMX7F105M-ARecalls & reported problems

Penumbra, Inc. · BMX81 105cm, MP Tip · Class 2

All models of Benchmark MAX System (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00815948024618), received between May 8, 2023 and May 8, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Fracture1
  • Material Deformation1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 8, 2023Fracture; Material Deformation; Physical Resistance/Sticking; Fracture; Material Deformation; Physical Resistance/Sticking

    THE PATIENT WAS UNDERGOING A THROMBECTOMY PROCEDURE IN THE MIDDLE CEREBRAL ARTERY (MCA) USING A BENCHMARK BMX 81 ACCESS SYSTEM (BMX81) AND A NON-PENUMBRA MICROCATHETER. IT WAS REPORTED THAT THE PATIENT¿S ANATOMY WAS TORTUOUS. DURING THE PROCEDURE, WHILE ADVANCING THE BMX81 THROUGH THE RADIAL ARTERY, THE PHYSICIAN EXPERIENCED RESISTANCE AFTER ENCOUNTERING A RADIAL LOOP OF THE ARTERY. NEXT, UPON REMOVAL, THE PHYSICIAN NOTICED THAT THE BMX81 WAS KINKED AND FRACTURED AT THE DISTAL LOCATION. THE PROCEDURE WAS COMPLETED USING A BENCHMARK BMX96 ACCESS SYSTEM (BMX96) AND THE SAME MICROCATHETER. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16887498 ↗

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page