Skip to content

CorPath GRX Robotic Drive 303Recalls & reported problems

Corindus, Inc. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00816280023031), received between Jun 19, 2018 and Sep 9, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Device Handling Problem1
  • Device Issue1
  • Display or Visual Feedback Problem1
  • Electrical /Electronic Property Problem1
  • Electrical Shorting1
  • Material Deformation1
  • Thermal Decomposition of Device1

Patient problems

  • No Known Impact Or Consequence To Patient3
  • No Clinical Signs, Symptoms or Conditions2
  • Electric Shock1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2024Device Handling Problem; Device Handling Problem

    IT WAS REPORTED THAT - "SYSTEM ERROR WHEN TRYING TO POWER UP. REPORTS OF ELECTRICAL SPARKS WHEN CONNECTING, ERROR CODE 1001."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20176561 ↗
  • MalfunctionJul 1, 2021Break; Material Deformation

    SERVICE REPORTS THAT DURING CONNECTION OF THE RD CABLE BY THE USER DURING SETUP, A PIN OF THE CONNECTOR WAS BROKEN. THIS RESULTED IN A BURNT CONNECTOR AND A DAMAGED PCB.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12102214 ↗
  • MalfunctionJul 9, 2020Thermal Decomposition of Device

    SUSPECT WIRING ISSUE AS ROBOTIC DRIVE CONNECTOR HAS A BURN MARK.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10254445 ↗
  • MalfunctionSep 24, 2019Electrical Shorting; Electrical Shorting

    IT WAS REPORTED THAT THE ROBOTIC DRIVE DID NOT POWER ON. UPON INVESTIGATION FROM THE CORINDUS SERVICE TEAM, IT WAS NOTED THAT THE ELECTRICAL CABLE WAS BURNT AND THE PINS WERE MISSALIGNED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9109643 ↗
  • MalfunctionApr 17, 2019Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    DURING A SERVICE EVENT, IT WAS DISCOVERED THAT THE ELECTRICAL CABLES OF THE ROBOTIC DRIVE AND EXTENDED REACH ARM WERE DAMAGED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8522776 ↗
  • MalfunctionJul 24, 2018Display or Visual Feedback Problem; Display or Visual Feedback Problem

    THE ROBOTIC DRIVE TOUCHSCREEN BECAME UNRESPONSIVE WHEN THE TECHNICIAN SELECTED "LOAD/EXCHANGE DEVICE" TO REMOVE THE GUIDEWIRE. THE SYSTEM WAS REBOOTED BUT THE "CONTINUE CASE" BUTTON COULD NOT BE SELECTED DUE TO THE UNRESPONSIVE TOUCHSCREEN. THE EMERGENCY RELEASE BUTTON ON THE CASSETTE HAD TO BE PRESSED TO DISASSEMBLE THE CASSETTE AND REMOVE THE INTERVENTIONAL DEVICES FROM THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7713953 ↗
  • MalfunctionJun 19, 2018Device Issue; Device Issue

    WHEN ATTEMPTING TO CONNECT THE CABLE BETWEEN THE ROBOTIC DRIVE AND EXTENDED REACH ARM, SPARKS WERE SEEN.

    Patient problems: Electric Shock

    MAUDE report 7616354 ↗

Other System, Catheter Control, Steerable devices

All System, Catheter Control, Steerable devices →

Sources and updates