CorPath GRX Robotic Drive 303Recalls & reported problems
Corindus, Inc. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00816280023031), received between Jun 19, 2018 and Sep 9, 2024 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Device Handling Problem1
- Device Issue1
- Display or Visual Feedback Problem1
- Electrical /Electronic Property Problem1
- Electrical Shorting1
- Material Deformation1
- Thermal Decomposition of Device1
Patient problems
- No Known Impact Or Consequence To Patient3
- No Clinical Signs, Symptoms or Conditions2
- Electric Shock1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 9, 2024Device Handling Problem; Device Handling Problem
IT WAS REPORTED THAT - "SYSTEM ERROR WHEN TRYING TO POWER UP. REPORTS OF ELECTRICAL SPARKS WHEN CONNECTING, ERROR CODE 1001."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20176561 ↗MalfunctionJul 1, 2021Break; Material Deformation
SERVICE REPORTS THAT DURING CONNECTION OF THE RD CABLE BY THE USER DURING SETUP, A PIN OF THE CONNECTOR WAS BROKEN. THIS RESULTED IN A BURNT CONNECTOR AND A DAMAGED PCB.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12102214 ↗MalfunctionJul 9, 2020Thermal Decomposition of Device
SUSPECT WIRING ISSUE AS ROBOTIC DRIVE CONNECTOR HAS A BURN MARK.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10254445 ↗MalfunctionSep 24, 2019Electrical Shorting; Electrical Shorting
IT WAS REPORTED THAT THE ROBOTIC DRIVE DID NOT POWER ON. UPON INVESTIGATION FROM THE CORINDUS SERVICE TEAM, IT WAS NOTED THAT THE ELECTRICAL CABLE WAS BURNT AND THE PINS WERE MISSALIGNED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9109643 ↗MalfunctionApr 17, 2019Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
DURING A SERVICE EVENT, IT WAS DISCOVERED THAT THE ELECTRICAL CABLES OF THE ROBOTIC DRIVE AND EXTENDED REACH ARM WERE DAMAGED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8522776 ↗MalfunctionJul 24, 2018Display or Visual Feedback Problem; Display or Visual Feedback Problem
THE ROBOTIC DRIVE TOUCHSCREEN BECAME UNRESPONSIVE WHEN THE TECHNICIAN SELECTED "LOAD/EXCHANGE DEVICE" TO REMOVE THE GUIDEWIRE. THE SYSTEM WAS REBOOTED BUT THE "CONTINUE CASE" BUTTON COULD NOT BE SELECTED DUE TO THE UNRESPONSIVE TOUCHSCREEN. THE EMERGENCY RELEASE BUTTON ON THE CASSETTE HAD TO BE PRESSED TO DISASSEMBLE THE CASSETTE AND REMOVE THE INTERVENTIONAL DEVICES FROM THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7713953 ↗MalfunctionJun 19, 2018Device Issue; Device Issue
WHEN ATTEMPTING TO CONNECT THE CABLE BETWEEN THE ROBOTIC DRIVE AND EXTENDED REACH ARM, SPARKS WERE SEEN.
Patient problems: Electric Shock
MAUDE report 7616354 ↗
Other System, Catheter Control, Steerable devices
- CorPath GRX Control Console 301Corindus, Inc. · System, Catheter Control, Steerable7 reports
- CorPath GRX Extended Reach Arm 302Corindus, Inc. · System, Catheter Control, Steerable5 reports
- CorPath GRX Cassette 305Corindus, Inc. · System, Catheter Control, Steerable4 reports
- CorPath GRX Extended Reach Arm 312Corindus, Inc. · System, Catheter Control, Steerable1 report
- CorPath GRX Drape, Robotic Drive and Arm 309Corindus, Inc. · System, Catheter Control, Steerable1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page