Rhythmlink® RLSND121-1.5Recalls & reported problems

Rhythmlink International, LLC · 10 electrodes, 13mm long, 0.4mm diameter disposable paired subdermal needle with touch proof connector · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00816312020137), received between Oct 6, 2023 and Oct 6, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 6, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING A CARTOID ENDARTERECTOMY (CEA) A SUBDERMAL NEEDLE ELECTRODE (13MM NEEDLE) DISLODGED IN POSTERIOR RIGHT CALF. THE USER NOTED THAT THE NEEDLE WAS REMOVED BY FIRST REMOVING THE TAPE COVERING THE NEEDLE, FOLLOWED WITH REMOVAL OF THE NEEDLE FROM THE PATIENT. THE USER NOTED THAT THE NEEDLE TIP WAS DETACHED FROM THE ELECTRODE AND REMAINED IN THE PATIENT. A SMALL SKIN INCISION WAS MADE AFTER CLEANSING THE SKIN WITH BETADINE. LIMITED WOUND EXPLORATION AND ULTRASOUND COULD NOT LOCALIZE THE NEEDLE TO REMOVE IT. TWO SKIN CLOSURE STITCHES WITH 4-0 MONOCRYL WERE PLACED AND DERMABOND WAS APPLIED. AN XRAY WAS PERFORMED. THE DECISION WAS MADE TO LEAVE THE NEEDLE IN THE PATIENT UNTIL A FUTURE SURGERY IS PERFORMED TO ADDRESS THIS FOREIGN BODY AT A FOLLOW-UP APPOINTMENT IF THE NEEDLE TIP IS CAUSING DISCOMFORT OR PAIN. THE USER NOTED SOME POTENTIAL CONCERN FOR ELECTRODE CONTINUITY DURING STIMULATION, AND NOTED THERE WAS SOME CURRENT RETURN ISSUES IN THE RIGHT LEG DURING THE CASE WHEN PARALYTIC HAD WORN OFF AND PATIENT WAS MOVING LEGS WITH STIMULATION. WHEN ANESTHESIA HAD REDOSED PARALYTIC, THE CURRENT RETURN RECOVERED, AND THE STIMULATION APPEARED TO BE WORKING AS NORMAL THROUGHOUT REMAINDER OF

    Patient problems: Foreign Body In Patient

    MAUDE report 17886875 ↗