CoolSculpting System Applicator Surface 150 (S150) CS-A3-15S-00Recalls & reported problems
Zeltiq Aesthetics, Inc. · CoolSculpting System Applicator Surface 150 (S150) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jul 5, 2021Class IITerminated
An incorrect error messaging system that could potentially lead to: 1) Reporting a thermal event error causing a user to re-treating the affected anatomic area within 24 hours, 2) Not reporting a thermal event or any other error codes causing a user to continue treating without being aware that a thermal event has occurred.
Z-2383-2021 · Root cause: Software design · Recall details · FDA record ↗
Other Dermal Cooling Pack/Vacuum/Massager devices
- CoolSculpting Control Unit BRZ-CG1-CUD-110Zeltiq Aesthetics, Inc. · Dermal Cooling Pack/Vacuum/Massager1 death2,595 reports
- CoolSculpting Elite Control Unit CS-S3-002-D-00Zeltiq Aesthetics, Inc. · Dermal Cooling Pack/Vacuum/Massager1 recall207 reports
- CoolCore BRZ-AP1-063-000Zeltiq Aesthetics, Inc. · Dermal Cooling Pack/Vacuum/Massager1 recall12 reports
- CoolMax Applicator BRZ-AP1-080-000Zeltiq Aesthetics, Inc. · Dermal Cooling Pack/Vacuum/Massager1 recall11 reports
- CoolCurve Applicator BRZ-AP1-062-000Zeltiq Aesthetics, Inc. · Dermal Cooling Pack/Vacuum/Massager1 recall2 reports
- CoolFit BRZ-AP1-066-000Zeltiq Aesthetics, Inc. · Dermal Cooling Pack/Vacuum/Massager1 recall1 report
- CoolCurve+ Applicator BRZ-AP1-064-000Zeltiq Aesthetics, Inc. · Dermal Cooling Pack/Vacuum/Massager1 recall1 report
- CoolSculpting Upper Module BRZ-CG1-UMD-000Zeltiq Aesthetics, Inc. · Dermal Cooling Pack/Vacuum/Massager3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page