8000229Recalls & reported problems

Laser Peripherals, LLC · 272u Laser Fiber PL · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00816469021315), received between Jan 24, 2023 and Aug 7, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Material Disintegration1
  • Material Too Soft/Flexible1

Patient problems

  • Foreign Body In Patient2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2025Break

    A LASER FIBER WIRE BROKE OFF INSIDE OF THE PATIENT'S RIGHT KIDNEY. AFTER SPEAKING WITH THE SURGEON ABOUT IT, HE FELT THAT NO OTHER INTERVENTION WAS NECESSARY INCLUDING AN XRAY. HE SAID THE REASON HE FELT AN X-RAY WAS UNNECESSARY WAS BECAUSE "LASER FIBER TIPS ARE NON DETECTABLE THROUGH X-RAY". SURGEON ALSO STATED THAT AFTER LOOKING IN THE PATIENT'S URETER AND KIDNEY THAT THE FIBER WAS UNDETECTABLE AND "DID NOT SEE IT IN ORDER TO RETRIEVE IT". UPON FURTHER CONVERSATION, SURGEON DID STATE THAT HE WOULD "INFORM THE PATIENT OF WHAT HAPPENED" REGARDING THE FIBER WIRE TIP. THIS IS NOT THE FIRST TIME LASER FIBER WIRE TIPS OF THIS PARTICULAR BRAND AND BATCH HAVE BROKEN OFF DURING SURGICAL PROCEDURES. TIP OF THE FIBER WHILE IT WAS INSERTED THROUGH THE SCOPE. TIP OF THE FIBER BROKE WHILE RUNNING UP THE SCOPE. THE FIBER WAS STILL USEABLE. SETTINGS: ENERGY: 0.3J, FREQUENCY: 50HZ, POWER: 24W.

    Patient problems: Foreign Body In Patient

    MAUDE report 22737162 ↗
  • MalfunctionMay 7, 2025Break; Material Disintegration

    AT 0840 IT WAS TIME FOR LASER LITHOTRIPSY AND THE LASER REP, PASSED OFF THE AGILITI 272U LASER FIBER PL WITH LOT# DB25006. THE LASER FIBER BROKE, TWO PIECES BROKE OFF INSIDE THE PATIENT, TWO BROKEN PIECES WERE RETRIEVED AND BAGGED. A SECOND LASER FIBER, AGILITI 272U LASER FIBER PL ALSO WITH LOT# DB25006 WAS OPENED AND PASSED TO SURGEON. THE SECOND LASER FIBER BROKE, ONE PIECE BROKE OFF IN THE PATIENT, ONE PIECE WAS RETRIEVED AND BAGGED. A THIRD LASER FIBER, AGILITI 272U LASER FIBER PL WITH NEW LOT# DB25017 WAS OPENED AND PASSED TO SURGEON. THE THIRD LASER FIBER BROKE, ONE PIECE BROKE OFF IN THE PATIENT, ONE PIECE WAS RETRIEVED AND BAGGED. LASER SERVICES WERE DISCONTINUED. ALL FIBERS AND BROKEN PIECES WERE BAGGED AND LABLED. SURGEON REQUESTED DIFFERENT LASER MACHINE TO BE BROUGHT IN. STENT WAS PLACED AND PATIENT PROCEDURE CONCLUDED. AGILITI LASER REP, PUT IN AN INCIDENT REPORT WITH HIS COMPANY WHICH WAS TO BE SENT TO OR MANAGER.

    Patient problems: Foreign Body In Patient

    MAUDE report 21976055 ↗
  • MalfunctionJun 16, 2023Material Too Soft/Flexible

    MD NOTICED THAT THE FIBER WIRE WAS WEAK AND WAS GOING TO BREAK IF USED. IF THE FIBER WIRE HAD BEEN USED THERE WAS POTENTIAL THAT THE FIBER COULD HAVE BROKEN OFF INSIDE THE PATIENT. MD OPENED A NEW FIBER WIRE TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17142558 ↗
  • MalfunctionJan 24, 2023Break

    ONE FIBER BROKE AND WORE OUT FASTER THAN EXPECTED. CONTINUED TO OPEN ANOTHER FIBER THAT WORKED CORRECTLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16232205 ↗