FullFire® Select Diffusing Tip Laser Probe 20624-5Recalls & reported problems
Monteris Medical Inc · 2.2mm Diffusing Tip Laser Probe · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00816589020427), received between May 24, 2017 and Sep 22, 2017 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Operates Differently Than Expected1
- Device Operational Issue1
Patient problems
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, THERE WAS UNINTENDED PROBE HEATING. THERE WERE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6887399 ↗MalfunctionMay 24, 2017Device Operational Issue; Device Operational Issue
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, THE PROBE WAS PLACED AND THE THERMOCOUPLE WAS NOT REGISTERING PROPERLY. IT WAS NOTED THAT THE PROBE WAS DISCARDED AND A SECOND PROBE WAS PLACED. IT WAS NOTED THAT THE SECOND THERMOCOUPLE WAS MEASURING APPROPRIATELY. THERE WERE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE SUBMITTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6587324 ↗
Other Powered Laser Surgical Instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page