FullFire® Select Diffusing Tip Laser Probe 20624-1Recalls & reported problems
Monteris Medical Inc · 2.2mm Diffusing Tip Laser Probe · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00816589020700), received between Oct 24, 2018 and Oct 25, 2018 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture3
- Insufficient Heating3
Patient problems
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 25, 2018Fracture; Insufficient Heating; Fracture; Insufficient Heating
IT WAS REPORTED THAT A PROBE WAS OPENED DURING AN ABLATION PROCEDURE AND CONTAINED A BROKEN FIBER. THE PROBE WAS INSERTED AND FIRED, HOWEVER NO HEATING WAS OBSERVED. THE PROBE WAS THEN REMOVED AND NO PILOT LIGHT WAS VISIBLE AT THE PROBE TIP. THE PROBE WAS REPLACED AND THE CASE PROCEEDED SUCCESSFULLY. THERE WERE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8005201 ↗MalfunctionOct 25, 2018Fracture; Insufficient Heating; Fracture; Insufficient Heating
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, THE USER PULLED THE PROBE OUT OF THE PACKAGING AND SET TO THE CORRECT DEPTH. THE USER THEN PLUGGED IN THE E2K CONNECTOR FULLY TO CHECK THE PILOT LIGHT AND IT WAS NOTED THAT THERE WAS NO PILOT LIGHT PRESENT. THE USER THEN SWITCHED PROBES OUT. THERE WERE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8005198 ↗MalfunctionOct 24, 2018Fracture; Insufficient Heating; Fracture; Insufficient Heating
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, THE PROBE DID NOT HAVE HEAT OUTPUT. THE CLINICAL SPECIALIST (CS) CHECKED THE LASER OUTPUT FOR THE PROBE BEFORE INSERTING INTO THE BOLT AND NOTICED THE RED BEAM. HOWEVER, WHEN FIRING THE LASER THE USER EXPERIENCED COLD PIXELS AND NO HEATING. THE PHYSICIAN THEN REMOVED THE PROBE AND THERE WAS NO RED BEAM ANYMORE. THIS CAUSED AN APPROXIMATE DELAY OF 10 MINUTES. THE USER THEN SWITCHED OUT PROBES. THERE WERE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8001511 ↗
Other Powered Laser Surgical Instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page