SideFire® Directional Laser Probe 20368-000 (REF NBP-000-01)Recalls & reported problems
Monteris Medical Inc · 3.3mm Directional Laser Probe · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00816589020816), received between Oct 24, 2018 and Nov 7, 2018 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture3
- Insufficient Heating3
Patient problems
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 7, 2018Fracture; Insufficient Heating; Fracture; Insufficient Heating
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, UPON CHECKING THE LASER TIP PRIOR TO INSERTING THE PROBE, THE LIGHT LOOKED DIM. THE CLINICAL SPECIALIST THEN INSPECTED THE UMBILICALS AND NOTED RED LASER LIGHT WAS VISIBLY NOTICEABLE CLOSE TO THE WHITE AND YELLOW DEPTH LOCK. THERE WERE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8046516 ↗MalfunctionOct 25, 2018Fracture; Insufficient Heating; Fracture; Insufficient Heating
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, THE LASER LIGHT COULD NOT BE SEEN AT THE TIP OF THE PROBE AFTER THE FIBER WAS PLUGGED IN. IT WAS ALSO NOTED THERE WAS LASER LIGHT SEEN ALONG THE FIBER CABLE. THERE WERE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8005204 ↗MalfunctionOct 24, 2018Fracture; Insufficient Heating; Fracture; Insufficient Heating
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, THE LASER FIBER OF THE PROBE BURNED ABOVE THE PORTABLE CONNECTOR MODULE (PCM) CONNECTION. UPON STARTING THE LASER ABLATION, THE HEAT WAS INCREASING AT A SLOW RATE ON THE NOISE MASK. THE USER THEN STOPPED LASING AND THE CLINICAL SPECIALIST (CS) WENT INTO THE MAGNET BAY AND IMMEDIATELY SMELLED BURNING PLASTIC. IT WAS NOTED THAT ABOVE THE LASER FIBER PLUG, THE FIBER WAS CHARRED. THE PROBE WAS THEN DISCONNECTED AND A NEW PROBE WAS INSERTED. THE USER WAS ABLE TO CONTINUE WITH THE ABLATION WITHOUT ISSUE. THERE WERE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8001619 ↗
Other Powered Laser Surgical Instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.