NeuroBlate Optic SideFire® Laser Probe 20891-00Recalls & reported problems
Monteris Medical Inc · 3.3mm Directional Laser Probe · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00816589020878), received between Jan 4, 2019 and Dec 9, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Radiation Leak” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Radiation Leak2
- Fracture1
- Insufficient Heating1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 9, 2024Radiation Leak; Radiation Leak
DURING AN ABLATION PROCEDURE, IT WAS REPORTED THAT WHILE ABLATING, HEAT WAS NO LONGER VISIBLE IN THE TREATMENT AREA DESPITE THE LASER BEING FIRED, AND THE TEMPERATURE READING AT THE HOTTEST POINT HAD DECREASED. THE PROBE WAS CHECKED AND IT WAS NOTICED THAT THE OUTER SHEATH OF THE CABLE WAS THERMALLY DAMAGED. THE PROBE WAS REPLACED AND THE TREATMENT RESUMED WITH NO FURTHER ISSUES. THERE ARE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20883276 ↗MalfunctionDec 6, 2023Radiation Leak; Radiation Leak
DURING AN ABLATION PROCEDURE, IT WAS REPORTED THAT, PRIOR TO INSERTING THE PROBE THROUGH THE ROBOTIC PROBE DRIVER, THE COOLING AND PILOT LIGHT WERE CHECKED. NO ISSUES WERE NOTED AT THIS TIME. THE VOLUMETRIC MRI SCAN WAS RUN, COORDINATES WERE ENTERED, AND THE ABLATION STARTED. FOR THE FIRST MINUTE, THE ABLATION WENT AS PLANNED AND THE PROBE FUNCTIONED AS INTENDED. AFTER THE FIRST MINUTE PASSED, IT WAS NOTED THAT THE SIDEFIRE PROBE BEING USED WAS NOT OBSERVED TO BE HEATING THE TARGETTED LESION. SINCE THE STOPPING POINT HAS BEEN REACHED, A NEW BURN WAS STARTED AT THE SAME POINT. THE USER BEGAN ABLATING AGAIN AFTER STARTING THE MRI, WITH NO HEAT BEING NOTED ON THE NOISE MASK. THE ABLATION WAS PAUSED, AND UPON WALKING INTO THE MRI, THE PROBE WAS INSPECTED AND A RED PILOT LIGHT WAS VISIBLE HALFWAY UP THE PROBE WIRE. THE PROBE WAS REMOVED AND WAS SENT BACK TO THE MONTERIS HEADQUARTERS. A NEW PROBE WAS INSERTED, AND THE ABLATION CONTINUED WITHOUT ISSUE. THERE WERE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18274399 ↗MalfunctionJan 4, 2019Fracture; Insufficient Heating; Fracture; Insufficient Heating
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, THE RED LIGHT AT THE TIP WAS FAINT, AND THERE WAS VISIBLE LIGHT EXITING THROUGH THE UMBILICAL CLOSE TO THE DEPTH STOP. THIS WAS NOTED AFTER REMOVAL FROM THE PACKAGE, THE LASER FIBER WAS PLUGGED IN. THERE WERE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8218955 ↗
Other Powered laser surgical instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page