NeuroBlate Optic SideFire® Laser Probe 20891-1Recalls & reported problems
Monteris Medical Inc · 3.3mm Directional Laser Probe · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00816589020892), received between Sep 17, 2019 and Oct 15, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Insufficient Heating3
- Break1
- Radiation Leak1
Patient problems
- No Known Impact Or Consequence To Patient4
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 15, 2025Radiation Leak; Radiation Leak
DURING AN ABLATION PROCEDURE, IT WAS REPORTED THAT AFTER THE LASER PROBE WAS INSERTED AND ABLATION HAD STARTED, NO HEAT WAS BEING SHOWN ON THE SOFTWARE. IT WAS THEN NOTICED THAT THERE WAS AN EVIDENT KINK ON THE LASER PROBE UMBILICALS WHERE THE LASER FIBER EXTENDS. THE LASER PROBE WAS REPLACED WITH A NEW ONE AND THE CASE RESUMED WITH NO FURTHER ISSUES REPORTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23290534 ↗MalfunctionOct 23, 2019Fracture; Insufficient Heating; Fracture; Insufficient Heating
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, WHILE TESTING THE LASER OUTPUT PRIOR TO INSERTION, THE LASER LIGHT WAS VISIBLY EMITTING WITHIN THE WHITE CORD THAT LEADS TO THE LASER CONNECTOR. IT WAS NOTED IT THE DAMAGE MAY HAVE OCCURRED WHILE BEING REMOVED FROM THE PACKAGING. THERE WERE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9226688 ↗MalfunctionSep 23, 2019Break; Break
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, THE PHYSICIAN NOTICED THE TIP OF THE PROBE WAS NOT SEEN UPON REMOVAL FROM THE BOLT. SCANS WERE TAKEN TO TRY TO IDENTIFY THE SAPPHIRE TIP BUT IT WAS LATER FOUND ON THE FLOOR IN THE OPERATING ROOM. THE PHYSICIAN REPORTED THAT THE ABLATION ZONES LOOKED "GOOD" IN THE SOFTWARE BUT THAT AFTER THE CASE, POST-OPERATIVE SCANS WERE TAKEN AND THE ABLATION ZONE LOOKED MINIMAL. THERE WERE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9106385 ↗MalfunctionSep 18, 2019Fracture; Insufficient Heating; Fracture; Insufficient Heating
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, THE LASER LIGHT WAS SHINING IN THE UMBILICALS AFTER HOOKING IT UP TO TEST THE LIGHT. THE USER REPLACED THE PROBE. THERE WERE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9083188 ↗MalfunctionSep 17, 2019Fracture; Insufficient Heating; Fracture; Insufficient Heating
IT WAS REPORTED THAT DURING AN ABLATION PROCEDURE, IT WAS NOTICED THAT THE FIBER WAS BROKEN. THIS WAS DISCOVERED UPON CONNECTING LASER; THE PROBE WAS NEVER INSERTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED IN RELATION TO THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9078968 ↗
Other Powered laser surgical instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.