BioGuard 00711784Recalls & reported problems
US Endoscopy · The single use BioGuard air/water valve (4-piece) is used to control the air / water function of an endoscope during GI endoscopic procedures. The single use BioGuard suction valve is used to control the suction function of an endoscope during GI endoscopic procedures. The single use BioShield biopsy valve is used to cover the opening to the biopsy/suction channel of the gastrointestinal endoscope. The BioShield biopsy valve provides access for endoscopic device passage and exchange, helps maintain insufflation, minimizes leakage of biomaterial from the biopsy port thought the endoscopic procedure and provides access for irrigation. The single use Torrent irrigation scope connector is used in conjunction with the Torrent irrigation tubing to provide irrigation via irrigation fluids such as sterile water supplied during gastrointestinal endoscopic procedures when used in conjunction with an irrigation pump (or electrosurgical unit). · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00816765013007), received between Oct 13, 2017 and Feb 13, 2019 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak4
- Detachment Of Device Component1
- Device Operates Differently Than Expected1
- Leak/Splash1
- Self-Activation or Keying1
- Structural Problem1
Patient problems
- No Known Impact Or Consequence To Patient2
- No Patient Involvement2
- Exposure to Body Fluids1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 13, 2019Structural Problem
THE USER FACILITY REPORTED THE LID OF THE BIOPSY VALVE OPENED, WITH RESULTANT FLUID SPRAY FROM THE DEVICE DURING PROCEDURAL USE. THE FLUID DID NOT CONTACT THE USER, AND THERE WAS NO REPORT OF HARM TO THE PATIENT OR USER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8337355 ↗MalfunctionMay 10, 2018Fluid/Blood Leak
THE USER FACILITY REPORTED FLUID SPRAY FROM THE BIOSHIELD BIOPSY VALVE. THE SPRAY DID NOT CONTACT THE USER, AND THERE WAS NO REPORT OF HARM TO THE PATIENT OR USER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7504004 ↗MalfunctionMay 9, 2018Fluid/Blood Leak; Self-Activation or Keying
THE USER FACILITY REPORTED THE LID OF THE BIOSHIELD BIOPSY VALVE OPENED AND SPRAYED FLUID. THERE WAS NO REPORT OF HARM TO THE PATIENTS OR USERS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7500556 ↗MalfunctionDec 8, 2017Fluid/Blood Leak
THE BIOSHIELD BIOPSY VALVE IS USED TO COVER THE OPENING TO THE BIOPSY/SUCTION CHANNEL INLET OF A GASTROINTESTINAL ENDOSCOPE. THE BIOSHIELD BIOPSY VALVE PROVIDES ACCESS FOR ENDOSCOPIC DEVICE PASSAGE AND EXCHANGE, HELPS MAINTAIN INSUFFLATION AND MINIMIZES LEAKAGE OF BIOMATERIAL FROM THE BIOPSY PORT THROUGHOUT THE GASTROINTESTINAL ENDOSCOPIC PROCEDURE. THE USER FACILITY REPORTED THAT POST PROCEDURE, THE BIOPSY VALVE IN THE BIOGUARD AIR/WATER AND SUCTION VALVE KIT SPRAYED FLUID. THERE WAS NO CONTACT WITH PATIENT OR USER.
Patient problems: No Patient Involvement
MAUDE report 7099659 ↗MalfunctionOct 27, 2017Detachment Of Device Component; Fluid/Blood Leak
THE BIOPSY VALVE IS USED TO COVER THE OPENING TO THE BIOPSY/SUCTION CHANNEL INLET OF A GASTROINTESTINAL ENDOSCOPE. THE BIOPSY VALVE PROVIDES ACCESS FOR ENDOSCOPIC DEVICE PASSAGE AND EXCHANGE, HELPS MAINTAIN INSUFFLATION AND MINIMIZES LEAKAGE OF BIOMATERIAL FROM THE BIOPSY PORT THROUGHOUT THE GASTROINTESTINAL ENDOSCOPIC PROCEDURE. THE USER FACILITY REPORTED FOLLOWING A PROCEDURE THE LID OF THE BIOPSY VALVE OPENED AND SPRAYED FLUID. AS THE EVENT TOOK PLACE AFTER A PROCEDURE THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 6983404 ↗MalfunctionOct 13, 2017Leak/Splash; Device Operates Differently Than Expected
THE BIOSHIELD® BIOPSY VALVE IS USED TO COVER THE OPENING TO THE BIOPSY/SUCTION CHANNEL INLET OF A GASTROINTESTINAL ENDOSCOPE. THE BIOSHIELD® BIOPSY VALVE PROVIDES ACCESS FOR ENDOSCOPIC DEVICE PASSAGE AND EXCHANGE, HELPS MAINTAIN INSUFFLATION AND MINIMIZES LEAKAGE OF BIOMATERIAL FROM THE BIOPSY PORT THROUGHOUT THE GASTROINTESTINAL ENDOSCOPIC PROCEDURE. THE USER FACILITY REPORTED THE BIOPSY VALVE IN THE BIOGUARD AIR/WATER AND SUCTION VALVE KIT SPRAYED FLUID THAT CONTACTED A PHYSICIAN'S AND A NURSE'S PERSONAL PROTECTIVE EQUIPMENT. THERE IS NO REPORT OF PATIENT OR USER HARM.
Patient problems: Exposure to Body Fluids
MAUDE report 6948829 ↗
Other Endoscope introducer kit devices
- Medline Industries, Inc. DYNJ58648JMedline Industries, Inc. · Endoscope introducer kit2 recalls0 reports
- ACS BUCY78EAmerican Contract Systems, Inc. · Endoscope introducer kit2 recalls0 reports
- Medline Industries, Inc. DYNJ44673GMedline Industries, Inc. · Endoscope introducer kit2 recalls0 reports
- Medline Industries, Inc. DYNJ66447Medline Industries, Inc. · Endoscope introducer kit2 recalls0 reports
- Merit Medical® 00884450360408Merit Medical Systems, Inc. · Endoscope introducer kit1 recall0 reports
- CARDINAL HEALTH SEN15SERM5Cardinal Health 200, LLC · Endoscope introducer kit1 recall0 reports
- CARDINAL HEALTH PGVMDVB26Cardinal Health 200, LLC · Endoscope introducer kit1 recall0 reports
- CARDINAL HEALTH SBA73CLK26Cardinal Health 200, LLC · Endoscope introducer kit1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page