THERMAGE FLX TG-3ARecalls & reported problems
Solta Medical · Thermage FLX RF System · Class 1
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00816995020639), received between Apr 19, 2024 and Apr 19, 2024 — classified as 1 injury.
The most frequently reported specific device problem is “Excessive Heating” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Excessive Heating1
- Radiation Output Problem1
Patient problems
- Blister1
- Burn(s)1
- Erythema1
- Skin Inflammation/ Irritation1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 19, 2024Radiation Output Problem; Excessive Heating; Radiation Output Problem; Excessive Heating
A USER FACILITY REPORTED THAT A PATIENT EXPERIENCED A BURN WITH BLISTERING ON THE LEFT LOWER EYELID DURING A THERMAGE FLX TREATMENT. AVAILABLE PICTURES WERE REVIEWED BY THE MEDICAL REVIEWER AND ERYTHEMA AND INFLAMMATION WERE VISIBLE ON THE PATIENT''S FACE. IN A SECOND PICTURE PROVIDED, A BLISTER WAS VISIBLE ON THE LOWER EYELID OF ONE EYE. AN EYE SHIELD AND CLOROPH EYE SHIELD LUBRICATION WERE USED. ENCAIN TOPICAL NUMBING CREAM WAS APPLIED TO THE PATIENT. SECONDARY INTERVENTION INCLUDES LUCAS PAPAW/OINTMENT. THE PATIENT''S CURRENT STATUS IS REPORTED THAT THE AFFECTED AREA IS FORMING SCAR TISSUE; HOWEVER, IT IS UNKNOWN IF THERE WILL BE ANY PERMANENT DAMAGE OR SCARRING. NO OTHER TREATMENTS (BESIDES THE ONE REPORTED) WERE BEING PERFORMED IN THE SAME AREA WHERE THE SYMPTOMS WERE REPORTED. THE PATIENT HAS NOT UNDERGONE ANY OTHER TREATMENTS IN THE SAME SYMPTOM AREA WITHIN THE PAST 90 DAYS. THE INCIDENT OCCURRED AFTER 230 SHOTS WITH THE HIGHEST ENERGY OF 2.5 MJ BEING USED. ADDITIONALLY, RADIOFREQUENCY POWER ERRORS, UNDER DELIVER ERRORS, AND TIP TOO WARM ERRORS OCCURRED DURING THE TREATMENT. THE USER TRIED TO REAPPLY THE RETURN PAD WHEN THEY FIRST SAW THE ERROR CODE AND REPLACED IT WITH A NE
Patient problems: Blister; Burn(s); Erythema; Skin Inflammation/ Irritation
MAUDE report 19142952 ↗
Other Massager, Therapeutic, Electric devices
- Theragun Theragun PROTherabody, Inc. · Massager, Therapeutic, Electric1 recall51 reports
- Thermage TG-2BSolta Medical · Massager, Therapeutic, Electric1 recall0 reports
- PIEZOWAVE2 100506USRichard Wolf GmbH · Massager, therapeutic, electric1 recall0 reports
- Thermage TTNS3.00E4-1200Solta Medical · Massager, Therapeutic, Electric30 reports
- THERMAGE FLX TT4.00F6-900Solta Medical · Massager, Therapeutic, Electric27 reports
- THERMAGE FLX TT0.25F6-450Solta Medical · Massager, Therapeutic, Electric13 reports
- Thermage TTNS3.00E4-900Solta Medical · Massager, Therapeutic, Electric9 reports
- THERMAGE FLX TT4.00F6-600Solta Medical · Massager, Therapeutic, Electric8 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.